F100 HANDHELD PRECISION THERMOMETER |
|
|
|
F100 Handheld
Precision Digital Thermometer |
|
|
F100 High performance
handheld 2-channel thermometer
High performance handheld 2-channel thermometer for Pt100 and Thermistor
probes.
If you're looking for a step up in measurement accuracy with a battery
powered, handheld thermometer, F100 is for you.
|
 |
-
Accuracy
±0.02°C over full
range
-
Resolution 0.001 °C
-
Range: -200°C to +850°C
-
Stability: <0.005°C per year
-
Common inputs for “SMART” &
normal
PT100 & Thermistor probes
-
USB communications port as
standard
-
Display in °C, °F, K
(Kelvin) or ohms
-
Statistical functions:
Average, Min/Max
and Standard Deviation
|
|
|
|
|
|
|
F100 can handle all your
needs, with accuracies and resolutions normally associated with a bench
top thermometer. Two inputs give you direct temperatures from Pt100 or
Thermistor probes, or you can display the temperature difference between
them. You can log measurements directly to memory or,
if you prefer, use
the USB port to control and data log with your PC.
F100 will display for you a statistical analysis of your measurements,
with average value, min and max values and standard deviation. Because
F100 will display in ohms as well as temperature, you can even use it to
calibrate one probe against another.
Most instruments require you to enter calibrated probe data into memory
to optimize measurement accuracy, which can mean a lot of data entering
if you change probes regularly. With ASL's SMART connector on your
probes, you only need to store the data once - in the connector! The
calibration data stays with the probe - permanently. You can even use it
on another F100 without any further action.
The SMART connector saves
you time and reduces error. If you have existing calibrated or
uncalibrated probes, no problem, F100 automatically registers if a probe
is SMART or normal.
|
|
Specification |
|
|
| Range |
-200°C to +850°C |
| Accuracy |
F100 only: ±0.02°C (±20mK |
| Resolution |
0.001°C |
| Repeatability |
±2 least significant digits
(±0.002°C), included in accuracy |
| Temperature coefficient |
<0.00005°C per °C ambient change |
| Data entry format |
ITS90 and CvD for calibrated probes;
or EN60751 for un-calibrated probes |
| Probe Current |
1mA switched d.c. (Pt100);
Auto-select 1mA, 25µA & 5µA for thermistors |
| Probe Types |
Pt100 and Pt25 resistance thermometers plus NTC thermistors. Cable
length typically 2M |
| Maximum Cable Length |
100 meters for Pt100; 30 meters for
SMART Pt100 |
| Thermometer Input
connectors |
2 x 5 pin industrial DIN socket with screw lock to connecting plug,
accepting both normal and SMART probes. |
| Display |
Backlit LCD |
| Communications |
USB Interface |
| User Functions |
Display hold, Display zero, One-shot measurement, Measurement log,
Statistical analysis - min/max, average, standard deviation. |
| Operating Conditions |
Full specification range: +15°C to +25°C.
Full operating range: 0°C to +40°C <80%RH non-condensing |
| Power |
Nickel-Metal Hydride (Ni-MH) rechargeable battery. Charging socket
built-in. Low battery indicator |
| Dimensions |
232mm(L) x 97mm(D) x 53mm(W) nominal |
| Weight |
500 g(1.1 lbs) |
|
|
|