Accuracy |
<±1 mK (±1 ppm of full scale ratio) at
unity ratio (0.01ºC) |
|
<±2 mK (±2 ppm of full scale ratio) up
to 250ºC |
|
<±5 mK (±5 ppm of full scale ratio) up
to full scale |
|
<±10 mK with 2.5 ohm and 0.25ohm high
temperature thermometers. |
|
Resolution |
0.1mK (25 ohm and 100ohm PRT), 1mK
(0.25ohm and 2.5ohm PRT) |
|
Linearity |
<±1 least significant digit (ratio) |
|
Measurement time |
10 second for full balance |
|
Warm up time |
<30 seconds with external reference
resistor |
|
Thermometers |
R0.01 of 1000ohm, 100ohm,25ohm, 10ohm,
1ohm, 0.25ohm and any intermediate value |
|
External reference resistor range |
1ohm to 1000ohm AC/DC resistance |
|
Internal Standard Resistor |
100 ohm ±50ppm, with t/c of <0.3ppm/°C.
A trim control allows calibration against an external standard. |
|
Measurement Range |
0-3.999999 ratio of two resistors;
0-3.999999 ohms where Rs = 1 ohm |
|
0-399.9999 ohms where Rs = 100 ohms;
0-1500.000 ohms where Rs = 1000 ohms. |
|
Sensor current |
0.1, 1 , 5, or 10mA constant current,
stabilized to ±1% of nominal value. |
|
Communications |
RS232 |
baud rate can be set between 110 and
19200 |
|
IEEE-488 |
Address selectable from 0 to 31 |
|
Analogue outputs |
±10V null balance signal (1 Hz
bandwidth) |
|
Operating conditions |
0°C to 40°C with external reference,, 10
to 90%RH |
|
Power requirements |
240,220,120,100 VAC (±10%) selectable on
rear panel, 50 or 60 Hz, 45VA max. |
|
Dimensions / weight |
442mm (W) x 104mm (H) x 320mm (D). 2U x
19" front panel. 8.5Kg |