Toll Free: 866-436-2430                     Email: sales@hotektech.com

 
 
Instruments, calibrators, standards, analyzers, dataloggers. analyzers
 
  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

STANDARD REFERENCE RESISTOR


RR Standard Reference Resistor. Resistance Standard Accuracy +/- 0.01%
RR Standard Resistors series of 4 wire reference resistors offer high accuracy and stability at an economic cost.
Designed specifically for use as a comparison standard for resistance thermometry on the ASL range of AC resistance bridges, the RR series may also be used in electrical standards laboratories.
In addition to standard fixed values, resistors can be supplied in any value between 1 and 1000 Ohms. Typical applications for the RR series are as transfer standards or reference source in temperature calibration or electrical standards laboratories

Standard Reference Resistor

Standard Reference Resistor
Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com
Standard Reference Resistor Specification
Tolerance 1 to 10 Ohms                     ±0.005% (50ppm)
>10 to 100 Ohms                ±0.0025% (25ppm)
>100 to 1000 Ohms             ±0.001% (10ppm)
Test certificate ±5ppm
Optional calibration Up to ±0.1ppm
Stability <±5ppm per year
Temperature coefficient <1ppm per °C
Frequency dependence Typically <1ppm (<0.25mK), up to 400Hz
Working temperature 20°C, but within the range of 15°C to 40°C
Connection type Full 4 wire to 4mm insulated terminals, plus earth connection to case.
Dimensions / weight Diameter: 76.2mm, Height: 90mm, Weight: 1kg.

ORDER CODES
RRx  where x is 1, 10, 25, 100, 300 or 1000 ohms. Other values available.
BNC/ADAP  Adaptor for converting 4mm insulated terminals to BNC socket (2 req'd)
CAL-RES-A Optional UKAS calibration d.c. @ 20°C

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.