Available Ranges |
| Gauge |
Vacuum |
Absolute |
Differential* |
|
0-0.03 psig
0-1 psig
0-5 psig
0-7.2 psig
0-15 psig
0-30 psig |
0-100 psig
0-300 psig
0-1000 psig
0-1500 psig
0-3000 psig
0-5000 psig |
0 to –1 psig
0 to –5 psig
0to –15 psig |
0-15 psia
0-30 psia
0-50 psia |
0-1 psid
0-5 psid
0-30 psid
0-50 psid |
|
|
All ranges come standard with eight engineering
units, readout
selection, keypad selectable by the operator (in H2O, inHg, PSI,
kPA,
mmHg, mBar, cmH2O, Bars, plus a user defined scale).
*Maximum static pressure, 200 psig |
|
|
|
Special ranges
available upon request. |
| Accuracy: |
|
Range (PSI)
0.3
1
5-100
300-1500
Above 1500 |
0.07% of reading + 0.03% of scale + 1 digit
0.07% of reading + 0.03% of scale + 1 digit
0.05% of reading + 0.02% of scale + 1 digit
0.06% of reading + 0.03% of scale + 1 digit
0.1% of full scale + 1 digit
|
|
Approval (optional): classified by GSA as
intrinsically safe for hazardous locations Class I, Div. 1, Groups
A, B, C, and D; test report LR92907-3 |
|
|
Electrical:
0-10 VDC 0.025% of range + 1 digit
0-30 VDC 0.1% of range + 1 digit
0-20/50 mA 0.025% of range + 1 digit
Sensor Configuration (specify):
a. Single—psig/ABS/vacuum
b. Dual—psig/ABS/vacuum (any combination)
c. Single—differential
Resolution: 50,000 counts maximum (depends
on
range and engineering units)
Media Compatibility: clean, dry,
non-corrosive air
or gases for low pressure ranges through 100 psi; for
300 psi and higher, any media that is compatible with
316 SS.
Temperature:
a. Operating: 0 to 50 degC
b. Effect on accuracy: none, over operating
range
c. Storage: -30 degC to 5000, with batteries; -
30 degC to 7000 without batteries |
Overpressure:
Operating, without damage or calibration shift:
0-15* psi 400% of sensor range
30-250 psig 300% of sensor range
300-5000 psi 150% of sensor range
*200" W.C. is 300% of sensor range
Pressure Fittings: 1/8" FNPT 316 SS
Electrical Connections: miniature recessed
banana
jacks
Display: simultaneous pressure and
electrical on a
two-line alphanumeric LCD display
Battery: NiCd (four AA cells); 50 hours
nominal at
2500, rechargeable
Battery Charger: 115 VAC, 50/60 Hz, wall
mount;
12-14 hours for full charge; optional 12 VDC charger
available
|