MULTIFUNCTION CALIBRATOR |
|
|
MC1200 Multifunction Calibrator Temperature , Current, Voltage, Frequency, RTD. |
|
MC1200 Multi-Function
Calibrator.
MC1200 Multi-Function
Calibrator provides a feature set unmatched in high accuracy, hand-held
calibrators in its price range. The MC 1200 Multi-Function Calibrator provides the
functions and accuracy associated with fixed installation, laboratory
instruments, and has everything needed for virtually any calibration
task.
Model MC1200 Multi-Function Calibrator Measure and source thermocouples, RTDs, current, voltage, and
frequency, and source pulse trains.
|

Multi-Function Calibrator |
-
Measure
and source T/Cs (13 types),
RTDs (13 types), Ohms, current, voltage,
frequency; source pulse
trains
-
Isolated
mA/V read-back circuit for complete transmitter calibration
-
Pressure
module communication port compatible with Fluke 700 Series, Mensor,
and
Martel pressure modules
-
Built-in
24 V supply can drive 4 to 20 mA loops
up to 1000 Ohms
-
Direct
entry of custom RTD coefficients
(RO, A, B, C)
|
|
|
|
|
|
-
All
source modes can be programmed with dedicated setpoints to speed
calibration and
linearity tests
-
Highest
accuracy in class to 0.015% of reading
-
Meets CE
requirements and is designed to
IEC 1010 safety standards
-
Fluke,
Mensor, and Martel Pressure Modules
|
A communications port compatible with BETA pressure modules is provided,
as is an isolated mA/V read-back circuit. Arrow keys, direct numeric
keypad entry, and three software-driven function buttons, plus a large
backlit, menu-driven graphics display combine to provide a highly
intuitive, simple yet powerful operator interface.
Built-in 250 Ohm resistor for Hart™ compatibility, compatibility with
smart transmitters and PLCs,
full fuseless protection, and a serial
communications port for full control with ASCII commands, are just some
of the additional features that make the MC 1200 Multi-Function Calibrator the single, most
indispensable tool available for virtually any calibration task.
The MC1200 Multi-Function Calibrator is supplied in a tough, rubber boot; a carrying case is also
available as an option.
|
|
MC1200 Multi-Function Calibrator
Specifications
(23 °C
±5
°C
unless otherwise noted) |
|
| Voltage
Read and Source |
| Source |
0.000
to 20.000 VDC |
| Read |
|
|
Isolated |
0.000
to 30.000 VDC |
|
Non-isolated |
0.000
to 20.000 VDC |
| Thermocouple
mV |
| Read
and Source |
-10.000
to +75.000 mV |
| Current
(mA) |
| Source |
0.000
to 24.000 mA |
|
Read |
| Isolated |
0.000
to 24.000 mA |
| Non-isolated |
0.000
to 24.000 mA |
| Frequency
(1 to 20 V selectable amplitude) |
| CPM
Source and Read |
2.0
to 600.0 CPM |
| Hz
Source and Read |
1.0
to 1000.0 Hz |
| kHz
Source and Read |
1.0
to 10.00 kHz |
| Pulse
(Source only; 1 to 20 V selectable
amplitude) |
| Pulses |
1
to 30,000.0
2 CPM to 10 kHz |
| Ohms |
| Source |
5.0
to 4000 Ohms |
| Read |
0.00
to 4000.0 Ohms |
| Thermocouple
Read and Source |
| J
Thermocouple |
-200.0
to +1200.0 °C |
| K
Thermocouple |
-200.0
to +1370.0 °C |
| T
Thermocouple |
-200.0
to +400.0 °C |
| E
Thermocouple |
-200.0
to +950.0 °C |
| R
Thermocouple |
-20.0
to +1750.0 °C |
| S
Thermocouple |
-20.0
to +1750.0 °C |
| B
Thermocouple |
+600.0
to +1800.0 °C |
| C
Thermocouple |
0
to +2316.0 °C |
| XK
Thermocouple |
-200.0
to +800.0 °C |
| BP
Thermocouple |
0
to +2500.0 °C |
| L
Thermocouple |
-200.0
to +900.0 °C |
| U
Thermocouple |
-200.0
to +400.0 °C |
| N
Thermocouple |
-200.0
to +1300.0 °C |
| RTD
Read and Source |
| Ni120
(672) |
-80.0
to +260.0 °C |
| Pt100
(385) |
-200.0
to +800.0 °C |
| Pt100
(3926) |
-200.0
to +630.0 °C |
| Pt100
(3916) |
-200.0
to +630.0 °C |
| Pt200
(385) |
-200.0
to +630.0 °C |
| Pt500
(385) |
-200.0
to +630.0 °C |
| Pt1000
(385) |
-200.0
to +630.0 °C |
| Cu10 |
-100.0
to +260.0 °C |
| YSI400 |
+15.00
to +50.00 °C |
| Cu50 |
-180
to +200 °C |
| Cu100 |
-180
to +200 °C |
| Pt385-10 |
-200
to +800.0 °C |
| Pt385-50 |
-200
to +800.0 °C |
|
| Environmental |
| Operating
Temperature |
-10
°C to +50 °C |
| Storage
Temperature |
-20
°C to +70 °C |
| Stability |
±0.005%
of reading/ °C
outside of 23 °C
±5
°C |
| Power
Requirements |
6
VDC |
| Batteries |
4
AA; alkaline or optional rechargeable |
| Mechanical |
| Dimensions |
8.7"
H x 4.2" W x 2.3" D (220.9 x 106.6 x 58.4 mm) |
| Weight |
30.5
ounces (863 gms) |
| Accessories
Included |
Test
leads, 4 AA alkaline batteries, NIST Certificate, instruction
manual |
| Optional
Accessories |
Carrying
Case, Model CC572 |
| Accuracy |
| Voltage |
±0.015%
of reading, ±2
mV |
| Thermocouple
mV |
±0.02%
of reading, ±10
µV |
Thermocouple
Errors
(in °C;
add 0.2 for Cold Junction
Compensation error) |
Type
J
Type K
Type T
Type E
Type R
Type S
Type B
Type C
Type XK
Type BP
Type L
Type U
Type N |
0.2 °C
0.3 °C
0.2 °C
0.2 °C
1.2 °C
1.2 °C
1.2 °C
0.6 °C
0.2 °C
0.9 °C
0.2 °C
0.25 °C
0.4 °C |
| RTD Read and
Source |
Ni120
(672)
Pt100 (385)
Pt100 (3926)
Pt100 (3916)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Cu10
YSI400
Cu50
Cu100
Pt385-10
Pt385-50 |
0.2 °C
0.2 °C
0.2 °C
0.2 °C
0.8 °C
0.4 °C
0.2 °C
1.4 °C
0.1 °C
0.4 °C
0.3 °C
1.4 °C
0.4 °C |
| Current
(mA) |
| Read |
±0.015%
of reading, ±2
µA |
| Source |
±0.015%
of reading, ±}2
µA |
| CPM |
| Read |
±0.05%
of reading, ±}1
LSD |
| Source |
±0.05%
of reading |
| Hz |
| Read |
±0.05%
of reading, ±1
LSD |
| Source |
±0.05%
of reading |
| kHz |
| Read |
±0.05%
of reading, ±1
LSD |
| Source |
±0.125%
of reading |
| Ohms
works with all pulsed transmitters down to 5 ms |
| 400
Ohm Range |
| Read |
±0.025%
of reading, ±0.05Ω. |
| Source |
±0.025%
of reading, ±0.05Ω |
| 4000
Ohm Range |
| Read |
±0.025%
of reading, ±0.5Ω |
| Source |
±0.025%
of reading, ±0.5Ω |
|
|