PRECISION LOOP CALIBRATOR
|
|
LC100 Precision Loop
Calibrator.
LC100 Precision Loop Calibrator provides significantly extended performance
when compared to any competitive calibrator. With an accuracy of 0.015%
of reading, with 0.001 mA resolution,
the LC100 Loop Calibrator has the highest
accuracy in its class.
Features such as a "% Error" function, which eliminates manual
error calculation and allows the display of the actual versus ideal
error at any calibration point, put the LC100 Loop Calibrator way beyond similar
instruments.
The LC100 Loop Calibrator can simulate, power, and measure
two-wire transmitters.
With automatic Step and Ramp functions, the LC100 Loop Calibrator enables remote
calibration of 4-20 mA devices.
|

Precision Loop Calibrator |
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
-
Loop
power measurement capability
-
Simulates,
powers, and measures two-wire transmitters
-
Five
preset outputs (4, 8, 12, 16, 20 mA)
-
Fuseless
input protection up to 250 VAC
|
|
Loop Calibrator
Specifications
(25°C
unless otherwise noted) |
|
|
Input |
|
Current Range
Voltage (Reading)
Input Protection |
0.000 to 24.000 mA
-25.00 to +125.00 %
0 to 28 VDC
fuseless, up to 250 VAC |
|
Output |
|
Source/Simulate Range
Drive Capability
Loop Supply Voltage
Range Select |
0.000 to 24.000 mA
-25.00 to +125.00 %
1200 Ohm without Hart
950 Ohm with Hart
24 VDC
Decade
Incremental to 0.001 mA steps |
|
Resolution |
1
µA |
|
Accuracy |
±0.015%
of reading ± 2
µA |
|
Temperature
Range |
| Operating |
-10
to +55 °C |
| Storage |
-20
to +70 °C |
|
Power
Requirements |
| Battery |
9
V, one required; included |
| AC
Adapter |
optional |
|
Mechanical |
| Dimensions
without boot |
5.7" H x 3.15" W x 1.43" D
(144.7 x 80.0 x 3.63 mm) |
| Weight |
12
ounces (340 gms) |
|
|