WIRELESS
DATA LOGGING SYSTEM |
|
|
A wireless data logging system for remotely
monitoring up to 200 data loggers |
|
|
MicroLog Plus wireless
logger system enables
wireless communication with up to 200
MicroLog
devices at the same time, to automatically deliver real-time
temperature and humidity measurements directly to your PC. Data from
external sensors can also be transmitted in the same manner.
The
complete MicroLog Plus wireless logger system is made up of the MicroLog data logger
device, the cradle that the device is mounted on, a small receiver
that is placed on the PC, and the user-friendly software that allows
for convenient central management. Enabling wireless communication
in this manner between multiple loggers and a single computer,
eliminates the need to download data from each unit manually, making
the MicroLog Plus wireless logger system is ideal for the ongoing data logging needs of
supermarkets, food transport, storage, air conditioning and
ventilation, clean rooms, warehouses and galleries.
|

Wireless Data Logger system
|
-
Capable of receiving data from loggers located up to 100m (US
Version) away from the computer
-
License-free wireless communication
-
Handles data from up to 200 loggers
-
Easy-to-use, color-coded user interface reports each logger
status
-
Programmable threshold for each logger that will activate an
alarm when crossed
-
Cost
effective, convenient and advanced solution for multiple data
logging
|
|
|
|
|
|
|
The
MicroLog
Logger comes with built in temperature and humidity sensors and an
analog port that is compatible with any type of external sensor.
The MicroLog
Cradle serves as a mount for the logger, and is the wireless RF
transmitter that enables transmitting readings at user-set intervals
to a remote PC. This item can be purchased separately.
The MicroLog Receiver is a small RF receiver that connects to the
monitoring station computer in order to receive wireless input from
the MicroLog Logger.
The MicroLog Plus
Software provides centralized control and management for
reliably monitoring up to 200 MicroLog devices from a remote
computer.
|
MicroLog
Models
EC600 temperature and external
sensor data logger
EC650 temperature, relative
humidity and external sensor data
logger
Inputs (Two built-in sensors)
Temperature -30ēC 50ēC
(resolution 0.5ēC, accuracy ą0.6ēC)
Relative Humidity 0 90 %
(resolution 0.5%, accuracy ą3%)
External Sensors
Voltage 0 10 V
Current 0 20 mA
Temperature -50 100 ēC
PT100 Temperature -10 10° C
-100 120 °C
pH 0 14 pH
Contact Open/Closed
Outputs
Three digit 7 segment LCD
IRDA interface to HP-printer and
host computer
RS232 serial communication at
19,200 bps
Memory Capacity 16,000 recording samples
Power Supply
Internal Battery Lithium - 3.6V 1.2AH 1/2AA
Battery Life Approximately two years, replaceable
(May vary when connecting external
sensor)
Sampling Rate
User Defined From every 10 second to every 2 hours
Dimensions
Round 72mm diameter
Thickness 22.9mm
Weight 55gr
CE and FCC standard compliance
|
North America RF
Transmission
EMC conformant to EN 301 489-3,
FCC PART 15.249
Usable range to 120m (30m indoors)
1mW at 914.5MHz
Harmonics/spurious emissions
-55dBc
8cm length antenna
CE and FCC standard compliance
MicroLog Plus Receiver
European Version
Usable range to 300m (75m indoors)
One RS232 communication port to
the computer
Power Supply
Internal Lithium battery 3.6V, 1.2AH, 1/2AA
External AC/DC 6V adapter
RF Receiver Red LED indicating RF signal
Green LED indicating valid data
being received
Type approved to ETS 300-220
CE and FCC standard compliance
North America Version
Usable range to 120m (30m indoors)
RS232 communication port to
the PC
Power Supply
Internal Lithium battery 3.6V, 1.2AH, 1/2AA
External AC/DC 6V adapter
RF Receiver Red LED indicating Ext power
connected
Green LED indicating valid data
being received
CE and FCC standard compliance |
MicroLog Cradle
Cradle Alarms
Audible Alarm
Visual Alarm Led
Serial Communication Channels
RS232 at 19.2Kbps
USB at 1.5Mbps
Cradle Memory
2000 samples holding the sensor
samples
Connectors
4-pin flat connection to the MicroLog
4-pin flat connection to any
MicroLog external sensor
Screw terminal for External DC
supply
Screw Terminal Board connections:
- Power supply: DC 6-30 V
- External sensor
- External contact sensor
- High alarm open collector 30V/2A
- Low alarm relay open collector
30V/2A
Power Supply
Internal Lithium battery, 3.6V
External 6 30 V, minimum 300mA
European RF Transmission
EMC conformant to EN 301 489-3
Type approved to ETS 300-220
Usable range to 300m (75m indoors)
418 (UK) & 433.92MHz
1mW on 418MHz, 10mW on
433.92MHz
2nd harmonic < -60dBc
16cm length antenna
|
MicroLab Plus Software
Data Displaying (from up to 200 MicroLogs)
Real-time temperature and humidity readings
Visual alarm when the logger crosses an upper or lower
alarm threshold for temperature or humidity
Battery level
An Excel file containing all of the measured data received
from the device
email/SMS messaging
Setting up the MicroLog
The ID of each of the monitors
The alarm levels of each monitor
Comment
The sampling rate from every minute to every hour
Minimum PC requirements
Windows 95/98/2000/XP/NT
6MB available disk space
CD ROM drive for software installation
Available communication port
Ordering Information
Part No. Description
EC600 Temperature data logger
EC650 Temperature and relative humidity data logger
DT132 External temperature sensor
DT252 External PT100 temperature sensor -10 10 °C
DT253 External PT100 temperature sensor -100 120 °C
DT140 External 0 10 V adapter
DT139 External 0 20 mA adapter
DT141 External contact sensor
DT175 Transmitter and alarm cradle
DT152 MicroLog Plus receiver
SFTMCL007 MicroLab Plus software
SFTMCL012 MicroLog Delux
DT047 Serial communication cable
DT180 USB communication cable
DT242 220/6V AC/DC receiver adapter
DT243 110/6V AC/DC receiver adapter
DT245 220/6V AC/DC transmitter adapter
DT246 110/6V AC/DC transmitter adapter
DT086 3.6V Lithium battery
10940 External antenna |
|