CDN 31 and CDN 33 Coupling Networks

Description
The coupling networks CDN 31 and CDN 33 are used in conjunction with the EFTC 30 burst generator.
The networks are fully relay controlled in order to enable complete test sequences to be produced.
A choice of either 1 phase (3 lines) or 3 phases (5 lines) are available.
Low weight and a carrying handle ensures complete freedom of use - also in the field.
The specifications meets or surpasses the requirements of the IEC 1000-4-4 standard.

CDN 30 series coupling network

For price and info please email sales@hotektech.com or call 253-927-6186 ext. 1

Related Instruments ESD Discharge Simulator
EFTC Burst Generator
Coupling Networks, Clamps
Nondestructive Test
Technical specifications of the CDN 31 and CDN 33 networks
Conforms to IEC 1000-4-4 and IEC 801-4
   CDN 31
   Lines L1, N, PE (1 phase network)
   Max. current 16A
   CDN 33
   Lines L1, L2, L3, N, PE (3 phase network)
   Max. current 16A
   Coupling control Relays
   Frequency range 1 to 100 MHz
   Coupling capacitor 33 nF
   Line to line cross talk > 30 dB
   Decoupling attenuation > 20 dB
   Coupling attenuation < 1 dB