Specifications
CLT10 Component Tester
|
| Measuring
Unit |
| Main
Specification |
|
| Generator
Frequency: |
10kHz
± 2 Hz |
| Voltmeter
Frequency: |
30
kHz |
| Voltmeter
Bandwidth: |
400
or 75 Hz selectable |
| Measuring
Speed: |
Measuring
cycle down to 10 ms excl.
handling. Specifications are valid
for ³14ms cycle with broad
voltmeter bandwidth. Test rate typically
more than 30 components per second*) |
| Accuracy |
± 1 dB or 5% + 1 digit
|
| Component
Range |
|
|
General
|
All passive
impedances. Primarily impedances
within100Ω
to 3 MΩ .
Restrictions for other ranges exist
|
| Range
switching time |
<500
ms |
| Range
switching life |
106
operations |
|
<300W
|
|
| Transformation
ratio |
1:1 |
| Input
impedance |
100Ω
± 2%
|
| Max.
power |
0.2 VA
@ ZX ³ 10
Ω
1
VA @ ZX ³ 50
Ω
4
VA @ ZX ³ 200
Ω
|
| Max.
10 kHz voltage |
36 Vrms
|
| RNL
@ 0.25 VA |
-150 dB,
typical -160 dB
(30 to 300Ω load)
|
| 300
W
- 3 kW |
|
|
Transformation
ratio
|
1:1
|
| Input
impedance |
1 kΩ
± 2%
|
| Max.
power |
3 VA, 4 VA @
ZX £ 2.5 kW
|
| Max.
10 kHz voltage |
100 Vrms
|
| RNL
@ 0.25 VA |
-160dB
|
|
3 kW
- 30 kW
|
|
|
Transformation
ratio
|
1:10
|
|
Input
impedance
|
10 kΩ
± 2%
|
|
Max. power
|
1 VA @
ZX ³ 5 kΩ
4
VA @ ZX ³ 25 kΩ
|
|
Max. 10 kHz
voltage
|
360 Vrms
|
| RNL
@ 0.25 VA |
-1 50 dB,
typical -160 dB
|
|
30 kW
- 3 MW
|
|
|
Transformation
ratio
|
1:10
|
|
Input
impedance
|
100 kW
± 2%
|
|
Max. power
|
4 VA @
ZX £ 250 kΩ
1 VA @ ZX £ 1
MΩ
0.25VA
@ ZX < 3 MΩ
|
|
Max. 10 kHz
voltage
|
1000 Vrms
|
|
RNL @ 0.25 VA
|
-140dB, typical -150 dB
@ ZX £ 300 kΩ
-130dB @ ZX £
3 MΩ
|
| Inputs/Outputs
(rear) |
|
Type of
connector
|
25-pole,
sub-D, female |
|
Inputs
Min.10
kHz level setting
|
0 to 10 V for
0 to 100% voltage output within range
|
|
10 kHz
voltage off
External trigger
|
Contact
closure
Contact closure
|
|
Outputs
|
|
10 kHz
voltage
|
0 to 10 V for
0 to 100%
within actual range
|
|
10 kHz
current
|
'0
to 10 V for 0 to 200 mA |
|
30 kHz
voltage
|
0 to 10 V for
0 to 100%
within actual range
|
|
Control
Outputs
|
|
Impedance
range
|
Open
collector
|
|
Accept, go
|
Open
collector
|
|
Reject, high
& low
|
Open
collector
|
|
10kHz voltage
on
|
Open
collector
|
|
Data Ready (DRDY)
|
TTL
compatible
|
|
Measurement
End (ME)
|
TTL
compatible
|
| Communication
with Control Unit |
|
Type of
connector
|
Fiber Optic
Link
|
|
Type of
interface
|
2 Mbit/s
serial, bi-phase modulation
|
|
Safety Link
|
Contact
closure
|
| Terminals
(front) |
|
Measuring
terminals
|
Two binding
posts accept stand- ard-size 4 mm
banana plugs
|
| General |
|
Temperature
|
|
Operating
temperature
|
5°
to 45° C (41°
to 113° F)
|
|
Storage
temperature
|
- 40°
to 70° C ( - 40°
to 158° F)
|
|
Relative
humidity
|
20 to 80%,
non-condensing
|
|
Line voltage
|
90 to 111 V
AC,
105 to 130 V AC,
180 to 222 V AC,
210 to 260 V AC.
|
| Frequency |
47.5 to 63 Hz
|
|
Power
consumption
|
100 VA
|
|
Dimensions
and Weight
|
|
Height
|
178 mm
(6.9")
|
|
Width
|
483 mm
(19.0")
|
|
Depth
|
442 mm
(17.0")
|
|
Net weight
|
18 kg (40
lbs)
|
|
Shipping
weight
|
25 kg (55
lbs)
|
|
Ordering
Information
|
|
Code
|
Description
|
|
391-081
|
CLT 10
Measuring Unit
|
|
Accessories
|
|
983-436
|
Service
Manual
|
|
983-437
|
Operator
Manual
|
|
900-215
|
Measuring
cable
|
|
906-362
|
Control Unit
connector cable, 1m
|
|
618-101
|
Control Unit
connector cable, 5m
|
| *)
Guideline only; depends on handling |