Specifications
CLT20 Component Linearity Harmonics Test System
|
| All
impedances are measured by 4-terminals |
| Input
impedance approximately 10Ω |
|
Output
impedance approximately 1Ω |
|
For normal resistance range >10Ω
and up to 3MΩ the instrument will
measure with 2 terminals (as CLT10) this means that the CLT20
will be measure exact the same as a standard CLT10.
|
| Measuring
voltage for the low impedance unit is 1/30 of the voltage
programmed by the control unit (CU) |
| Maximum
measuring voltage of the low Impedance unit is 1,2V in the range
<300 Ohm (36V/30V=1,2V) |
| In
the range 300 Ohm - 3 kOhm is 1,43V the maximum measuring
voltage (43V/30=1,43V) |
|
Specifications:
|
|
Parameters:
|
10kHz level
|
Measured 30kHz
|
|
ZX < 300Ω
|
Volt
|
DB
|
|
300Ω
|
8,66V
|
-174 dB
|
|
100Ω
|
5,0V
|
-160 dB
|
|
30Ω
|
2,74V
|
-157 dB
|
|
10Ω
|
1,58V
|
-156 dB
|
|
3Ω
|
0,16V
|
-143 dB
|
|
ZX Low Impedance
|
|
|
|
30Ω
|
0,5V
|
-161 dB
|
|
3Ω
|
0,5V
|
-161 dB
|
|
3Ω
|
0,167V
|
-172 dB
|
|
3Ω
|
0,01V
|
-150 dB
|
|
1Ω
|
0,5V
|
-163 dB
|
|
1Ω
|
0,167V
|
-170 dB
|
|
1Ω
|
0,01V
|
-149 dB
|
|
0,1Ω
|
0,334V (3,34A)
|
-151 dB
|
|
0,1Ω
|
0,1V
|
-167 dB
|
|
0,1Ω
|
0,01V
|
-149 dB
|
|
0,01Ω
|
0,1V (10A)
|
-158 dB
|
|
0,01Ω
|
0,05V
|
-159 dB
|
|
0.01Ω
|
0,01V
|
-149 dB
|
|
0.01Ω
|
0,00334V
|
-140 dB (0,334nV)
|
|
Measurements in
the specification scheme were measured with components
connected
to the 4-terminal fixture. The component used is very low
distortion 4-terminal units
|
|