CT10R LCR Remote
Controlled
Bridge
|
| Parameters
|
C
(Capacitance)
L (Inductance)
R (Resistance)
D (Dissipation Factor)
Q (1/D)
G (Conductance)
D (Deviation) on C, L or R |
|
| Parameter
Combinations
|
Automatically
selected:
Cs
with D
Ls
with Q
R with Cp
or Ls |
Selected
from controller
Cs
with D, Q or Rs
Cp
with D, Q, Rp or G
Ls with D, Q
or Rs
Lp
with D, Q, Rp
or G
D in
absolute Value
D in
% |
| Limits
|
Absolute
Limits,
Deviation Limits in:
Absolute Value, Percentage Value,
Bin Number.
Separate Limit for D or Q. |
|
| Measuring
Frequencies |
100/120
Hz and 1 kHz crystal controlled |
|
| Automatic
Control
|
Range
setting fully automatic 5 ms
per step
Auto-zeroing from the bus interface |
|
| Input
Protection |
Safe
Limits.
2 Joule up to 600 V |
|
| Measuring
Voltages |
1
V rms down to 40 Ohm
0.1 V from 40 Ω to 4
Ω, linear
reduction at lower impedance values |
|
| Input/Output
|
Front
panel: 4 BNC sockets and
RS232C
socket.
Rear Panel: 3 D-connectors
SC1 - SC2 - SC3. |
|
| |
SC1:
Limit outputs
|
Open
collector each rated at
60 V/0.4 A max 0.5 W. Built-in
diode protection or relay coils
(Common + terminal). Bin 0-9
and High/Low (for secondary
parameter).
Bias: External DC Supply, max
3 V. |
| |
SC2 |
IEEE-488
GPIB for IEEE bus interface |
| |
SC3 |
RS232C
input and output. Baud
rate max 9600, selectable.
Trigger |
| Built-in
Battery Back-up |
Memory
Retention min. 5 years |
|
| Temperature
Range |
Operating
15°C to 45°C
Storage -40°C to 75°C |
|