DB210 LCR Component Tester
|
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
|
Test Fixture
A 4-terminal (Kelvin) test fixture JIG32 can be supplied in case
that the DB210 should be used as a bench instrument.
The 4-terminal test fixture will effectively cancel all errors due
to contact resistance and connecting cable impedance.
The JIG32 has gold plated contacts, and up to 1meter long cables can
be connected between the test fixture and the DB210 without any
deterioration of the accuracy.
|
| Measured Parameters: |
C, L, R, (serial or parallel) Tan d,
ESR, Rs, Rp, L/Q, |
| Measuring Frequencies: |
100kHz, 10kHz, 1kHz, 100Hz and120Hz selectable
from the keyboard or by data-link |
| Measuring Voltages: |
1 V RMS down to 40 Ohm
0,3 V RMS from 40 Ohm to 4 Ohm
Linear reduction at lower impedance values |
| Measuring Speed: |
100Hz |
120Hz |
1kHz |
10kHz |
100kHz |
From trig to end of measurement
|
180ms |
180ms |
38ms |
38ms |
38ms |
| From trig to data ready: |
190ms |
190ms |
46ms |
46ms |
46ms |
| Add. time per meas. by average |
160ms |
160ms |
34ms |
34ms |
34ms |
| Measuring Cables: |
1m (39.3 inch) from
bridge to fixture |
| Bias Voltage
External: |
Up to ±48V DC |
| Accuracy
C & tan δ: |
Frequency |
100Hz
&120Hz |
1kHz |
Accuracy ±1 digit |
| Capacitance |
Tan d |
| |
100pF -
300µF |
1pF-99pF |
0,5pF* |
± .0010 |
| 100pF-389pF |
0,1% |
± .0005 |
| 390pF -
3µF |
0.05% |
± .0005 |
| 3mF - 30µF |
0.1% |
± .0010 |
| |
>3mF C: (C measured
/ 0.3mF) * 0.1% |
Tan d: (C measured /
0.3mF) * 0.005 |
| |
10kHz |
100kHz |
*) Accuracy ± 0,2pF |
| |
39pF - 3µF |
39pF - .3µF |
0,05% |
± .0005 |
|
|
|
| Bin Sorting: |
Up to 12 limits for 1st
parameter and 4 limit for 2nd parameter by opto-couplers |
| Interfaces |
Rear panel: IEEE 488 (GPIB),
RS232C and Ethernet connector |
| Control: |
Measure end, data ready,
trig ready, fault and status |
| Trig input |
DC, AC and contact closure |
| Keyboard & Display |
Front panel: For manual
settings, etc |
| Environment: |
Ambient temp.: 10-30 degrees
Celsius |
| Warm-up time: |
Minimum 30 minutes |
| Power: |
90-130 and 200-260 V AC,
50-60 Hz, |
| Calibration Interval: |
Minimum: Every 12 months |
|
|
|