The DB233 LCR Component Bridge Tester is specially designed for manual as well as
automatic high-speed high accuracy testing of capacitors or other LCR
applications. The instrument is reliable, user-friendly and easy to
set up to any test application on production lines, in quality control
departments or in laboratories.
|
DB233 LCR Component Bridge Tester
|
The DB233
LCR Component Bridge is well suited for mounting on sorting machines or other automatic test
applications
where the distance between the front
panel of the DB233 and the Jig is less than 50cm.
When the distance is longer, the DB232
should be preferred to provide maximum accuracy.
The DB233
LCR Component Bridge performs
capacitance and loss factor tests at any of the
4 standard frequencies.
Dual, triple and quadro frequency tests are
popular to give an immediate presentation of
capacitance and loss factor measurements
over a range of frequencies.
|
|
|
|
|
|
|
As standard the
instrument has a built-in comparator for deviation measurements,
IEEE488 (GPIB) and RS232C data interfaces as well as handler interface
(opto-coupler type) with 12+4 bins for production sorting. The
high-speed data interfaces may be used for an external computer in
order to control the system, or for collection of data for statistics
and analysis.
Bin sorting with up to
12 bins for capacitance for 1st frequency and up to 4 bins
for tan d using 2nd frequency. Or tan d may be measured at
several frequencies using the 4 bins for different levels of the
dissipation factor.
|
The
standard fitted PCMCIA card is the smart way of storing set-ups.
Fail-safe loading of set-ups to several instruments will be done fast
and efficient.
The
test cables are as standard connected to the front panel of the
instrument.
Another possibility is to order the DB233 in the version MCR in order to have the test cables connected to the rear panel only.
Optional
protection box PB10 protecting the instrument against charged
capacitors is available.
|
| Measured
Parameters: |
C,
L, R, Z (serial or parallel) tan¶,
ESR, Rs, Rp, L/Q, R-X, Z-q
(deg or rad) |
| Measuring
Frequencies: |
100k,
10k, 1kHz and 100Hz with multiple frequency facility |
| Measuring
Voltages:
|
1
V RMS up to 100mF at 100Hz |
| 1
V RMS up to 10mF
at 1kHz |
| 1
V RMS up to 1MF at 10kHz |
| 1
V RMS up to 0.1MF at 100kHz |
|
Above:
(linearly decreasing with the impedance).
Programmable in 0.1V steps (maximum
1,5V RMS)
|
| |
|
100Hz |
1kHz |
10kHz |
100kHz |
| Measuring
Speed
|
From
trig to end of measurement*: |
180ms
|
20ms |
20ms |
20ms |
| From
trig to data ready *: |
190ms
|
28ms |
28ms |
28ms |
| Additional
time per meas. by average |
160ms
|
16ms |
16ms |
16ms |
|
*)
allowing 3ms contact bouncing or 1 range change
Multiple measurements (average): The sum of each
measurement (from trig to end of measurement) + 8ms for
calculation time
|
| Measuring
Cables: |
1m
(39,3 inch) from front panel to fixture (Cables supplied)
|
| Input
Protection: |
2
Joule up to 1kV or 4mF
charged 1000V
|
| Bias
Voltage internal: |
Up
to ±3,0VDC
on generator terminal, set in 0,1V steps (internally generated)
|
| Frequency |
Accuracy
±1
digit |
Average ³2 |
| Capacitance
|
100Hz |
1kHz
|
Capacitance
|
Tan
d
|
|
300pF-3nF
|
1pF
- 39pF |
0.5%* |
±,0010
|
| |
40pF
- 3,9mF |
0,05%*
|
±,0002
|
| 3nF-30mF |
4mF
- 399mF |
0,1% |
±
,0007
|
| 30mF-300mF |
|
0,1% |
±
,0010
|
| 300mF-3mF |
400mF
- 1mF |
1% |
±
.0020 |
|
10kHz
|
100kHz
|
|
|
| ,1pF-3,9pF |
,03pF
- ,9pF |
0.1% |
±
,0010 |
| 4pF-3,9mF |
1pF-,9mF |
0.05%'' |
±
,0002 |
| 4mF-39mF |
- |
0.1% |
±
,0007 |
| - |
1mF
- 9mF |
0.2% |
±
.0010 |
| 40mF-400mF |
10mF
- 40mF |
1% |
±
.0020 |
*Accuracy
± 0,2pF
**Accuracy ± 2pF.
The above specifications require a stable jig with capacitance lower
than 30pF
|
| Inductance |
100Hz |
1kHz |
10kHz |
100kHz |
Accuracy |
| |
10mH-100H |
1mH-10H |
0,1mH-1H |
0,1mH
-1H |
1
parameter
0.1%
2 parameter
±(0.2%+005xQ) |
|
Resistance
|
0,4Ω-40Ω |
0.4Ω-40Ω |
0.4Ω-40Ω |
0.4Ω-40Ω |
0.1% |
| |
40Ω-4MΩ |
40Ω-4MΩ |
40Ω-4MΩ |
0.4Ω-1MΩ |
0.05% |
The above
specifications are valid for measurements with constant voltage
|
| Bin
Sorting
Up to 12 limits for 1st parameter
and 4 limits for 2nd parameter
by opto-couplers |
| Interfaces:
|
Rear
panel: |
IEEE
488 (GPIB) and RS232C
|
| Control: |
Measure end, data ready, trig ready, fault and status
|
| Trig
input: |
DC,
AC and contact closure
|
| Front
panel: |
PC
card for set-ups, save and loading
|
| Environment:
|
Ambient
temperature
|
10-30
degrees Celsius
|
|
Warm-up
time:
|
Minimum
30 minutes |
| Power: |
90-130 and 200-260 V AC, 50-60 Hz,
|
| Calibration
interval: |
Minimum: |
Every
12 months
|
|
|
Dimensions:
| |
Mainframe: |
Export
Packing Europe |
Export
Packing Overseas |
| Height |
140mm
or 5,5 inch |
30cm
or 11.7 inch |
32cm or 12.8 inch |
| Width |
438mm
or 17,2 inch |
51cm
or 20 inch |
52cm or 20.4 inch |
| Depth |
360mm
or 14,2 inch |
56cm
or 22 inch |
55cm or 21.6 inch |
| Weight |
Total
16kg or 36 lb. |
20kg
or 45 lb |
22kg or 49,5 lb |
|
|
|