DB234 LCR Component Tester is
especially designed for high accuracy testing of capacitors on
production lines, not least for integration with sorting machines in a
production environment.
The instrument is reliable, user-friendly and
easy to set up to any test.
The DB234 LCR Component
Tester
utilizes an
external bridge module allowing the user to install the measuring bridge very close to the measuring Jig. This ensures high measuring accuracy.
Especially when measuring at 100kHz or
200kHz cables are main causes to noise.
When installing an LCR bridge on a production line, some distance between the instrument and
the Jig is unavoidable.
|
|
DB234 LCR Component Tester
|
-
With the DB234, total cable length of
up to 3m (118 inches) is supplied.
-
The DB234 utilizes a well-proven input
protection system to protect the bridge
module from damages owing to
exposure to charged capacitors.
-
This secures that the DB234 does not
break down as easily as other LCR
bridges,
when exposed to charged capacitors.
|
|
|
|
|
The DB234 LCR Component
Tester can perform
dual frequency tests at any combination of frequencies.
A popular configuration is to test capacitance at 1kHz and loss factor
at 100kHz or 200kHz.
As standard, it can sort capacitors into bins according to the
measured parameters at two frequencies simultaneously.
Bin sorting with up to
12 bins for capacitance for 1st frequency and up to 4 bins
for tan d using 2nd frequency. Or tan d may be measured at
several frequencies using the 4 bins for different levels of the
dissipation factor.
|
As
standard the instrument has a built-in comparator for deviation
measurements, IEEE488 (GPIB) and RS232C data interfaces as well as
handler interface (opto-coupler type).
All measured data are collectable from the data interfaces.
Via the
PCMCIA slot it possible to easily store set-ups to distribute to other
instruments quickly, without operator mistakes.
|
| Measured
Parameters: |
C,
L, R, Z (serial or parallel) tan¶,
ESR, Rs, Rp, L/Q, R-X, Z-q
(deg or rad) |
| Measuring
Frequencies: |
200k,
100k, 10k and 1kHz with multiple frequency facility |
| Measuring
Voltages:
|
1
V RMS up to 10mF at 1kHz |
| 1
V RMS up to 1mF
at 10kHz |
| 1
V RMS up to 10nF at 200kHz |
| 1
V RMS up to 10nF at 1MHz |
|
Above:
(linearly decreasing with the impedance).
Programmable in 0.1V steps (maximum
1,5V RMS)
|
| |
1kHz |
10kHz |
100kHz |
200Hz |
| Measuring
Speed
|
From
trig to end of measurement*: |
20ms
|
20ms |
6ms |
6ms |
| From
trig to data ready *: |
28ms
|
28ms |
14ms |
14ms |
| Additional
time per meas. by average |
16ms
|
16ms |
2ms |
2ms |
|
*)
allowing 3ms contact bouncing or 1 range change
Multiple measurements (average): The sum of each measurement
(from trig to end of measurement) + 8ms for calculation time
|
| Measuring
Cables: |
1m
(39,3 inch) from bridge module to fixture (Cables supplied)
|
| Input
Protection: |
2
Joule up to 1kV or 4mF
charged 1000V
|
| Bias
Voltage internal: |
Up
to ±3,0VDC
on generator terminal, set in 0,1V steps (internally generated)
|
| Bias
Voltage external: |
Up
to ±48V DC
|
| Frequency |
Accuracy
±1
digit |
Average ³ |
| Capacitance
|
1kHz |
10kHz
|
Capacitance
|
Tan
d
|
|
1pF-39pF
|
0,1pF
- 3,9pF |
0.5%* |
±
,0010
|
| 40pF-3,9mF |
4pF
- 3,9mF |
0,05%**
|
±
,0002
|
| 4mF-399mF |
4mF
- 39mF |
0,1% |
±
,0007
|
| 400mF-1mF |
40mF
- 400mF |
1% |
±
,0020
|
|
100kHz
|
200kHz
|
|
|
| ,03pF-,9pF |
,01pF
- 3,9pF |
0,1pF |
±
,0010 |
| 1pF-,9mF |
|
0,05%** |
±
,0002 |
| |
4nF
- 9,9nF |
0,1% |
±
,0007 |
|
1mF-9mF
|
10nF-
99nF |
0,2% |
±
,0010 |
| 10mF-40mF |
100nF
- 999nF |
1% |
±
,0020 |
*Accuracy
± 0,2pF
**Accuracy ± 2pF.
The above specifications require a stable jig with capacitance lower
than 30pF
|
| Inductance |
1
kHz |
10kHz |
100kHz |
200kHz |
Accuracy |
| |
10mH-100H |
1mH-10H |
0,1mH-1H |
0,1mH-0,1H |
1
parameter
0.1%
2 parameter
±(0.1%+005xQ) |
|
Resistance
|
0,4Ω-40Ω |
0.4Ω-40Ω |
0.4Ω-40Ω |
0.4Ω-40Ω |
0.1% |
| |
40Ω-4MΩ |
40Ω-4MΩ |
40Ω-1MΩ |
40Ω-1MΩ |
0.05% |
The above
specifications are valid for measurements with constant voltage
|
Bin
Sorting
Up to 12 limits for 1st parameter
and 4 limits for 2nd parameter
by
opto-couplers |
| Interfaces:
|
Rear
panel: |
IEEE
488 (GPIB) and RS232C
|
| Control: |
Measure end, data ready, trig ready, fault and status
|
| Trig
input: |
DC,
AC and contact closure
|
| Front
panel: |
PC
card for set-ups, save and loading
|
| Environment:
|
Ambient
temperature
|
10-30
degrees Celsius
|
|
Warm-up
time:
|
Minimum
30 minutes |
| Power: |
90-130 and 200-260 V AC, 50-60 Hz,
|
| Calibration
interval: |
Minimum: |
Every
12 months
|
|
|
Dimensions:
| |
Mainframe: |
Bridge
module |
Export
Packing Europe |
Export
Packing Overseas |
| Height |
140mm
or 5,5 inch |
35mm
or 1,4 inch |
30cm
or 11.7 inch |
32cm or 12.8 inch |
| Width |
438mm
or 17,2 inch |
192mm
or 7,5 inch |
51cm
or 20 inch |
52cm or 20.4 inch |
| Depth |
360mm
or 14,2 inch |
205mm
or 8,1 inch |
56cm
or 22 inch |
55cm or 21.6 inch |
| Weight |
Total
16kg or 36 lb. |
1 kg
or 2.3 lb
|
21kg
or 47.3 lb |
23kg or 51.8 lb |
|
|
|