SCANNER FOR MEGOHMMETER

 DB640 Scanner for Insulation Measurements

DB640 Scanner for Insulation Test has been designed to be used together with the DB601, DB602 or DB604 Megohmmeters for applications such as sample testing of capacitors in QC departments or automatic testing of leads in a multiwire-cable. 
The outstanding specifications of the DB601/2/4 are maintained when measuring through the DB640 Scanner. It is built of modules each having four inputs, and can be configured as 4, 8,12 up to 40 inputs.



DB640 Scanner for Insulation Test
  • It is built of modules each having four inputs, 
    and can be configured as 4, 8,12 up to 40 inputs.
  • A complete scan of forty components can be 
    completed in less than 2,5 seconds, 
    which makes the DB640 will suited for use in 
    automatic capacitor testing machines, 
    using the scanner principle. 
  • The DB640 Scanner is standard fitted with 
    forty handler outputs for GO/NO GO testing 
    against a minimum limit programmed into the DB6012/2/4.
  • The handler outputs are electrically insulated
    with optocouplers.

For price and info please email sales@hotektech.com or call 253-927-6186 ext. 1

Related Instruments LCR Meters
Bridges Wheatstone, Kelvin
Component Testers
Ohmmeters
Insulation Testers
Digital Multimeters

All inputs feature a contact test function for cable/capacitors. 
Also each of the forty items under test can be tested for a short circuit in order, for example,
to switch it out, preventing overloading of the power supply.

The operation of the DB640 Scanner is done via a high speed serial link between the DB640 and a DB601/2/4 Megohmmeter. Quick, logical and easy manual operation takes place from the DB601/2/4 keyboard with soft key menus and the results are presented for example in tabular form on the large LCD display. Remote computer control is done from the IEEE or RS232 interfaces of the DB601/2/4.

Specifications Megohmmeter DB640

RESISTANCE MEASUREMENTS:

 

Range:

10 kOhm to 1 POhm (dependent on test Voltage)
(Peta = 1.000.000 Giga)

Accuracy:

Rx< 20 GOhm x Vtest: ± 2% of value

Rx> 20 GOHM x Vtest: ± Vtest/1pA in Ohm,
max reading Vtest/0.1pA

DC CURRENT MEASUREMENTS:

 

Range:

1 pA to 1 mA

Accuracy:

±2% of value ±1pA

DISPLAY READING

 

Direct :

3,5 digits up to 40 channels R or I

Single Channel Bar Graph:

8 ranges with 2 decades per graph with fixed- or autoscaling

Single Channel x – y Graph:

Current or resistance scale versus time scale

INPUT RESISTANCE:

10 kΩ for I-test> 10nA  
Programmable   10 kOhmmor 
1 MOhm for I-test< 10nA

TEST VOLTAGE

 

DB601 Range:

10 V to 1000kV in 1 V steps

DB602 Range 10V to 500V in 1V steps
DB604 Range 50V to 5kV in 1V steps

Accuracy:

±2% of value or ±1 Volt in un-calibrated mode

Stability: 

±10 ppm / degree C
±10ppm for 10% change in line voltage

DB604: 50V to 5000V in 1V steps

Source Resistance:

100 Ω

DB604: 3,3kΩ

Max currents:

Measurement: 2 mA
Charging: 25 mA (pass word protected)

DB602: precharge: 100mA
DB604: 4mA

Switching:

Manually ON/OFF from front panel or controlled by a built-in timer, or by remote control.

TIMING:

Programmable Fast Charging: 
0 to 9999 msec (charge resistance 200 Ohm)

Programmable Measurement delay:

0 to 9999 msec

Discharge Time:

t = 0.1 x Cx (in µF) with Vtest decreasing to 1 % of test level

DB604 Discharge Resistance: 100kΩ

MEASURING SPEED:

 

Trig Mode:

One Measurement: <52 msec (excl. charging)
Average up to n = 100measurements: <52 + (N-1) x 40 msec (excl. charging)

Continuous Mode:

Direct Reading: 90 to 4000 msec depending on average
Bar Graph: Display update every 52 msec

LIMITS:

5 built-in programmable limits on resistance or current

STANDARD FITTED INTERFACES:

 

IEEE 488: (IEEE 488-1 and 2)

"Talker Only" and "Talker/Listener" Modes. True sub-set of Standard protocol

RS232C:

Baud rate up to 9600 Baud. Full two-way control/output

Control I/O:

Optocoupler input / output 25 V/10mA. Trig, Measure END Signal, Trig Ready Signal, Data Ready Signal, Fault Signal, Limit Outputs

Ambient Temperature:

10 to 40 degrees Celsius

Power:

90 – 130 and 200 – 260 V AC 50 – 60 Hz

DIMENSIONS:

 

Height:

140 mm/ 5.5 inch

Width:

438 mm/ 17.2 inch

Depth:

360 mm/ 14.2 inch

Weight:

14.9 kg/ 32.8 lbs.

ACCESSORIES SUPPLIED:

Line power connector, Two 1.5 m coax cables with H.T. BNC connectors
Brackets for 19" rack-mounting, Manual in English

Options:

Item number

 

Higher measure current and / or charge

   

Fixture for axial and radial components:

   

Fixture for SMD components requires Jig 600

   

Test probe with trig contact:

   

© December 2001, Danbridge A/S ALL SPECIFICATIONS SUBJECT TO CHANGES WITHOUT PRIOR NOTICE

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.