Specifications
JP15U High Voltage Insulation Resistance Tester
|
| JP15U |
15
kV insulation tester. 110-240 V AC. 37546. |
| Stability |
Output
voltage varies less than ± 1% for supply voltage variations of ±10% |
| Ionization
Indication
|
An
active L.P. filter in series with an integrated audio amplifier
feeds a loudspeaker providing an audible ionization indication.
Amplifier gain is adjusted with a potentiometer. An output panel
connector provides external indication on headphones or on an
oscilloscope.
Min. load 100Ω. |
| Meter
Outputs |
5V
full-scale meter outputs are provided on a panel connector.
Min.
load 10kΩ. |
| Guard
Terminal |
For
connection to guard electrode. Current to this terminal bypasses
the current meter. |
| Test
Voltage
|
0-7,5
kV and 0-15 kV DC; adjusted by means of a 10-turn potentiometer.
Test voltage is applied to an insulated test probe via a shielded
coaxial cable. The probe is equipped with a high tension switch
and is terminated with a 4 mm straight tip. |
| Test
Voltage Switch
|
The
high tension is switched on with the H.T. switch located on the
test probe. A panel connector for external switching is located on
the front panel. |
| Voltmeter |
Three
ranges, 7.5 kV and 15kV full-scale.
Accuracy ±5%. |
| Max
Output Current |
Approximately
300µA from 0 to 15 kV, and when shortened. |
| Current
Meter |
Three
ranges 1 µA - 10 µA - 100 µA full scale.
Accuracy: ±5%. |
| Resolution |
20
nA. |
|
Output
Resistance
|
Approximately
200 kΩ at 15kV decreasing to
120kΩ below
1kV,measured by I out = 100 µA |
| Power
Supply |
100-130
V and 200-260 V AC 45-400 Hz,15-30 VA, depending on output voltage
and current. 12-13 V DC max 2 A from external source or from
optional battery. |
| Dimensions
|
Height:
148 mm
Width: 438 mm
Depth: 300 mm
Weight: 3,2 kg Shipping Weight 6 kg
Weight: 6.2 kg. (incl. of Battery and AC power supply)
Shipping Weight: 9 kg |
Options:
12 V 2Ah rechargeable, maintenance free-battery
Leather carrying case. Battery module for 9 D-cells or 10 Ni-Cd
cells |