pH mV DIGITAL METER DATALOGGER
|
|
|
HD2105.1 pH meter
Thermometer, Storage of maximum, minimum, average value, RS232 C output for
data transfer in real time to a PC or printer. Functions: REL, HOLD and auto
power off which can be disabled, IP 67 protection degree.
HD2105.2 pH meter Thermometer Datalogger which stores the
maximum, minimum, average value and can store up to 34.000 samples. RS232 C/
USB output for data transfer in real time to a PC or printer. Functions: REL,
HOLD and auto power off which can be disabled, IP 67 protection degree.
|
 |
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
HD2105.1 and
HD2105.2 are portable instruments
with a large LCD display. They measure the pH and the redox potential
(ORP)
in mV. Temperature measurement using Pt100 or Pt1000 immersion, penetration or
contact probes. The electrode calibration can be carried out on one,
two or three points and the calibration sequence can be chosen from a list
of 13 buffers. The temperature probes are fitted with an automatic detection module, with
the factory calibration settings already being memorized inside. HD2105.1 and HD2105.2 models are fitted with an
RS232C serial port and can transfer the acquired measurements in real time
to a PC or to a portable
printer.
|
INSTRUMENT TECHNICAL
CHARACTERISTICS
Instrument
Dimensions
(Length x Width x Height) 185x90x40mm
Weight 470g (complete with batteries)
Materials ABS, rubber
Display 2x41⁄2 digits plus symbols
Visible area: 52x42mm
Operating conditions
Operating temperature -5…50°C
Warehouse temperature -25…65°C
Working relative humidity 0…90%RH without condensation
Protection degree IP67
Power
Batteries 4 1.5V type AA batteries
Autonomy 200 hours with 1800mAh alkaline batteries
Power absorbed with instrument off 20μA
Mains Output mains adapter 9Vdc / 250mA
Security of memorized data Unlimited, independent of battery charge
conditions
Measurement of pH by Instrument
Measurement range -2.000…+19.999pH
Resolution 0.01 or 0.001pH selectable from menu
Accuracy ±0.001pH
Input impedance >1012Ω
Calibration error @25°C Offset>20mV
Slope<50mV/pH or Slope>63mV/pH
Sensitivity < 85% or Sensitivity > 106.5%
Measurement of mV by Instrument
Measurement range -1999.9…+1,999.9mV
Resolution 0.1mV
Accuracy ±0.1mV
Drift after 1 year 0.5mV/year
Measurement of temperature by Instrument
Pt100 measurement range -200…+650°C
Pt1000 measurement range -200…+650°C
Ni1000 measurement range -50…+250°C
Resolution 0.1°C
Accuracy ±0.1°C
Drift after 1 year 0.1°C/year
|
ORDER CODES HD2105.1KE: The kit is composed of: instrument HD2105.1, KP30
electrode, TP87 temperature probe, 4.01pH and 6.86pH buffer
solutions, connection cable for serial output HD2110CSNM, 4 1.5V
alkaline batteries, operating manual, case and DeltaLog9 software.
HD2105.1K: The kit is composed of: instrument HD2105.1, TP87
temperature probe, connection cable for serial output HD2110CSNM, 4
1.5V alkaline batteries, operating manual, case and DeltaLog9
software. The electrodes must be ordered separately. HD2105.2KE: The kit is composed of: instrument HD2105.2
datalogger, KP30 electrode, TP87 temperature probe, 4.01pH and
6.86pH buffer solutions, connection cable HD2101/USB, 4 1.5V
alkaline batteries, operating manual, case and DeltaLog9 software.
HD2105.2K: The kit is composed of: instrument HD2105.2
datalogger, TP87 temperature probe, connection cable HD2101/USB, 4
1.5V alkaline batteries, operating manual, case and DeltaLog9
software. The electrodes must be ordered separately. HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole
female for RS232C. HD2101/USB: Connection cable USB 2.0 connector type A -
8-pole MiniDin. DeltaLog9: Software for download and management of the data on PC
using Windows 98 to XP operating systems. AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains
voltage. S’print-BT: On request, portable, serial input, 24 column thermal
printer, 58mm paper width.
|
pH Electrodes
KP20: Combined pH electrode, gel-filled, with screw connector
S7, body in Epoxy, Ag/AgCl sat. KCl. KP30: Combined pH electrode, cable 1m, gel-filled, body in
Epoxy, Ag/AgCl sat. KCl. KP60: Combined pH electrode, 1 diaphragm, gel-filled, with
screw connector S7, body in glass, Ag/AgCl sat. KCl. KP61: Combined pH electrode, 3 diaphragms for milk, cream,
etc. gel-filled, with screw connector S7, body in glass, Ag/AgCl
sat. KCl. KP62: Combined pH electrode, 1 diaphragm for pure water,
paints, etc. gel-filled, with screw connector S7, body in glass, Ag/AgCl
sat. KCl. KP70: Combined pH electrode, micro diam. 6 x L=70mm,
gel-filled, with screw connector S7, body in glass, Ag/AgCl sat. KCl.
KP80: Combined pointed pH electrode, gel-filled, with screw
connector S7, body in glass, Ag/AgCl sat. KCl. CP: Extension cable 1.5m with BNC connectors on one side and
S7 on the other side for electrode without cable. CE: Screw connector S7 for pH electrode. BNC: Female BNC for
electrode extension.
|
|
|