All ProductsAnalyticalAnalyzersCalibratorsHumidityData LoggersLight MetersMeasurement

 

MultimetersOhmmetersPressureStandardsTemperatureTransmitters

THERMOCOUPLE DIGITAL THERMOMETER 
HD2108 Digital Thermometer Logger Probes K, J, T, N, R, S, B, E

HD2108.1K Thermocouple thermometer for probes type K, J, T, R, N, S, B, E. Storage of maximum, minimum, average value, RS232 C output for data transfer in real time to a PC or printer.
Functions: REL, HOLD and auto power off which can be disabled, IP 67 protection.
HD2108.2K Thermocouple thermometer for probes type K, J, T, R, N, S, B, E. Datalogger which stores maximum, minimum, average value, and can store up to 76.000 samples. RS232 C/USB output for data transfer in real time to a PC or printer. Functions: REL, HOLD and auto power off.
Measuring unit °C - °F - °K - mV - mV*C

The HD2108.1 and HD2108.2 with one input portable instruments with a large LCD display. They measure
the temperature using immersion, penetration air or contact probes. The sensor may be a thermocouple
of type K, J, T, N, R, S, B or E.
The HD2108.2 datalogger memorizes up to 76,000 samples. These data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0.
The storing interval, printing, and baud rate can be
configured using the menu.
All the models are fitted with an RS232C serial
port and can transfer the acquired measurements
in real time to a PC or to a portable printer.

Measurement of temperature by Instrument
TC measurement range: K -200…+1370°C
TC measurement range: J -100…+750°C
TC measurement range: T -200…+400°C
TC measurement range: N -200…+1300°C
TC measurement range: R +200…+1480°C
TC measurement range: S +200…+1480°C
TC measurement range: B +200…+1800°C
TC measurement range: E -200…+750°C
Resolution                     0.05°C up to 199.95°C
                                    0.1°C from 200.0°C to the full scale
Instrument precision
Thermocouple K            ±0.1°C up to 600°C
                                   ±0.2°C over 600°C
Thermocouple J            ±0.05°C up to 400°C
                                   ±0.1°C over 400°C
Thermocouple T            ±0.1°C
Thermocouple N           ±0.1°C up to 600°C
                                  ±0.2°C over 600°C
Thermocouple R           ±0.25°C
Thermocouple S           ±0.3°C
Thermocouple B           ±0.35°C
Thermocouple E           ±0.1°C up to 300°C
                                   ±0.15°C over 300°C

The accuracy only refers to the instrument. Error due to the thermocouple
or to the cold junction reference sensor is not included.

Temperature drift @ 20°C 0.02%/°C
Drift after 1 year 0.1°C/year

ORDER CODES
HD2108.1K: The kit is composed of the instrument HD2108.1 with one input, connection cable for serial output HD2110CSNM, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software.
The probes must be ordered separately.
HD2108.2K: The kit is composed of the HD2108.2 with one input, datalogger, connection cable HD2101/USB, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software.
The probes must be ordered separately.
HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C.
HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin.
DeltaLog9: Software for download and management of the data on PC using Windows 98 to XP operating systems.
AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains voltage.
S’print-BT: On request, portable, serial input, 24 column thermal printer, 58mm paper width.

 

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.

[Home] [Top] [Distribution] [Contact Us]