MANOMETER LOW PRESSURE
DIGITAL METER |
|
|
|
HD2114P Manometer Low Pressure Digital Meter |
|
|
HD2114P.0K
Micro manometer-Thermometer for air speed and flow rate measurement by means
of Pitot tubes, full scale 20mbar for speed measurements from 2...55m/s.
Temperature measurement by K type thermocouple. Storage of maximum, minimum,
average value. Functions: REL, HOLD and auto power off which can be
disabled, IP 67 protection degree.
HD2114P.2K Micro manometer-Thermometer for air speed and flow rate
measurement by means of Pitot tubes, full scale 20mbar for speed
measurements from 2...55m/s. Temperature measurement by K type thermocouple.
Datalogger which stores the maximum, minimum, average value and can store up
to 36.000 couples of samples. RS232C/ USB output for data transfer in real
time to a PC or printer. Functions: REL, HOLD and auto power off which can
be disabled, IP 67 protection degree.
Measuring units °C - °F - Pa - mbar - mmH2O - PSI - m/s - km/h - ft/m - mph
- knot - l/s - m3/h - cfm
|
 |
The HD2114P.0
and HD2114P.2 are portable
micro-manometers using Pitot tubes and a large
LCD display.
They are used to perform measurements in the
fields of air conditioning, heating and ventilation.
They measure the differential pressure detected by a
Pitot tube connected to the instrument inputs
acquiring the wind speed and flow rate inside
pipelines and vents.
They also measure the temperature using a type K thermocouple sensor.
The instruments can be used as thermometers,
and can be employed with any kind of thermocouple
K sensor if a standard miniature connector is used.
|
|
|
|
|
Measurement range
Differential pressure: ±20mbar Speed (*): 2 … 55m/s Temperature using thermocouple K: -200…+1370°C Temperature using Pitot tube: -200…+600°C Maximum overpressure: ±300mbar
Resolution Differential pressure: 0.005mbar - 0.5Pa Speed: 0.1 m/s - 1 km/h - 1 ft/min - 1 mph - 1 knots Flow rate: 1l/s - 0.01·103m3/h - 0.01·103cfm Temperature: 0.1°C
Accuracy Differential pressure: ±0.4%f.s. Speed: ±(2% reading+0.1m/s) Temperature: (**) ±0.1°C Minimum speed: 2 m/s
(*) At 20°C,
1013mbar and Ps negligible. (**) The accuracy only refers to the instrument. The error due
to the thermocouple or to the cold junction reference sensor is
not included.
|
ORDER CODES
HD2114P.0 K: The kit is composed of the HD2114P.0 with 20mbar
full scale and thermocouple K input, 4 1.5V alkaline batteries,
operating manual, case. The Pitot tubes have to be ordered separately.
HD2114P.2 K: The kit is composed of the HD2114P.2 datalogger with
20mbar full scale and thermocouple K input, connection cable HD2101/USB,
4 1.5V alkaline batteries, operating manual, case and DeltaLog9
software. The Pitot tubes have to be ordered separately.
HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female
for RS232C.
HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole
MiniDin.
DeltaLog9: Software for download and management of the data on PC
using Windows 98 to XP operating systems.
PW: Extension with male-female standard miniature connectors to
connect the Pitot tube’s thermocouple K to the instrument, length 2m.
AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains
voltage.
S’print-BT: On request, portable, serial input, 24 column thermal
printer, 58mm paper width.
|
|
|