Pt
100 SENSOR DIGITAL THERMOMETER
|
|
HD9214 Thermometer, Pt100
sensor, range –200°C +600°C, HOLD key, °C/°F. TP 9… series probes may be
connected.
HD9215 Thermometer, Pt100 sensor, range -50°C +199°C, HOLD key,
°C/°F. TP 9… series probes may be connected.
These Thermometers are very compact portable instruments. Their high technological features are accompanied by a pleasing design. With
their range of interchangeable probes, they are indispensable instruments
for measuring temperature in the fields of maintenance, heating/air-conditioning,
laboratories, food and agriculture and all other sectors where temperature measurements
must be precise, fast and repeatable. The reading can be in °C or in °F. The calibration is managed by the keyboard.
|
|

Digital Thermometer
|
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
|
Automatic change of scale. Instrument switches off
automatically after 8 minutes
(the auto power off function may be deactivated)
Temperature measurement with a
Pt100 Platinum
sensor (100ohm at 0°C), 4-wire connection.
Probes available for
measurements by immersion,
surface contact and penetration (pointed), all
with 4wire connection and precision in accordance
with standards IEC
751/1983, BS 1904/1984, DIN 43760/1980 in precision classes A, 1/3 DIN,
(on request 1/5 DIN). Low battery charge warning
light Battery
power supply.
|
Calibration with storage of
the calibration data in the memory.
LCD display with 3 /half
digit, height 8 mm.
The total precision of the
instrument plus the chosen measuring probe is given by the sum of the
instrument error plus the error of the chosen probe depending on whether
the probe is class A, 1/3 DIN or 1/5 DIN.
The ohmic there and back
resistance of the cable of the probe is 0,40ohm.
Instrument working range:
-5'C...+50'C
Storage temperature: -20oC...+60'C.
Relative humidity: 0 ... 90% R.H.
Number of conversions: 1 per
second.
Power supply: 9V battery, IEC 6LF22, zinc/carbon battery life approx. 200
hours.
Low battery charge
warning light.
Instrument connector: male 8-pole circular
connector DIN 45326.
Case: ABS Bayer NOVODUR, gray 7553CF
Dimensions:
instrument 42x1
85x23 mm weight 130 grams
|
| Model |
Measuring
range |
Uncertainty |
Resolution |
Temperature
drift
-5...+50°C
|
°C/°F |
Hold |
MIN/
MAX |
| HD 9215 |
-50,0...+199,9°C |
±0,2°C |
0,1°C |
0°C/°C
18...25°C
0,01°C/°C
in the remaining
field |
* |
* |
 |
HD 9214 |
-199,9...+600°C
|
-199,9...-50,0°C ±2°C
-50,0...+199,9°C ±0,2°C
+200...+600°C ±2°C |
0,1°C
0,1°C
1°C
|
* |
*
 |
|
|
|
|