Toll Free: 866-436-2430                     Email: sales@hotektech.com

 
 
Instruments, calibrators, standards, analyzers, dataloggers. analyzers
 
  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

HD98569 pH CONDUCTIVITY DISSOLVED OXYGEN TEMPERATURE


HD 98569 Portable multi-parameter data logger for electrochemical parameters measurement and logging.
The HD 98569 is a portable multi-parameter data logger for electrochemical measures: pH, conductivity, dissolved oxygen and temperature. It is fitted with a large back-lighted LCD display.
The instrument measures:
• pH, mV, redox potential (ORP) with pH, redox or combined pH/temperature electrodes complete with SICRAM module;
• conductivity, resistivity in liquids, total dissolved solids (TDS), and salinity with combined 4-ring and
2-ring conductivity and temperature probes with SICRAM module.
• Concentration of dissolved oxygen in liquids (in mg/l), saturation index (in %) using SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor.

Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com

The instruments equipped with HD22BT Bluetooth option can transfer the data without any
connection to a PC fitted with USB/Bluetooth converter HD USBKL1, or to the printer S’print-BT with Bluetooth interface or to a PC with Bluetooth input.
The serial connection RS232C can be used for direct printing of labels with a 24 column printer
(S’print-BT).

The instrument ifs fitted with input for the measurement of temperature with Pt100 immersion, penetration or contact probes with SICRAM module.
• The pH electrode calibration can be carried out on one or five points and the calibration sequence can be chosen from a list of 8 buffers. Temperature compensation can be automatic or manual.
• The conductivity probe calibration can be performed with automatically detected conductivity calibration solutions: 147μS/cm, 1413μS/cm, 12880μS/ cm, 111800μS/cm or manually with calibration solutions having different values.
• The dissolved oxygen probe’s quick calibration function guarantees long term correctness of the performed measurements.
• pH, conductivity dissolved oxygen and temperature probes fitted with SICRAM module can store factory and calibration data inside.

Measured values pH - mV -  χ - Ω - TDS - NaCl - mg/l O2 - %O2 - °C - °F
Working temperature
Storing temperature
Working relative humidity
Protection degree
-5 … 50°C
-25 … 65°C
0 … 90% RH without condensate
IP66

RS232Cserial interface
Type
Baud rate
Data bit
Parity
Stop bit
Flow control
Length of serial cable
USB interface
Type


RS232C electrically isolated
Can be set from 1200 to 38400 baud
8
None
1
Xon/Xoff
Max 15m

1.1 - 2.0 electrically isolated

Bluetooth interface

 

Optional for PCs fitted with Bluetooth input or HD USB.KL1 Bluetooth / RS232 adapter. The interface can be installed in Delta Ohm only.

Measurement of pH by instrument

Measuring range

-9.999…+19.999pH

Resolution

0.01 o 0.001pH selectable from menu

Accuracy

±0.001pH ±1digit

Calibration error @25°C

|Offset| > 20mV
Slope > 63mV/pH or Slope < 50mV/pH
Sensitivity > 106.5% or Sensitivity < 85%

Measurement of mV by instrument

Measuring range

-1999.9…+1999.9mV

Resolution

0.1mV

Accuracy

±0.1mV ±1digit

Measurement of saturation index of dissolved oxygen
Measurement range 0.0…600.0%
Resolution 0.1%
Accuracy instrument ±0.3% ±1digit (in the range 0.0…199.9%)
±1% ±1digit (in the range 200.0…600.0%)
Salinity setting
Setting directly from menu or automatically by conductivity measurement
Setting range 0.0…70.0g/l
Resolution 0.1g/l
Temperature measurement with the sensor inside the O2 probe
Measurement range 0.0…50.0°C
Resolution 0.1°C
Accuracy instrument ±0.1°C
Measurement of temperature by instrument
Pt100 Measurement range -50…+150°C
Resolution 0.1°C
Accuracy instrument ±0.1°C ±1digit

Measurement of conductivity by instrument

  Range Resolution

Measurement range (K cell=0.01)

0.000…1.999μS/cm

0.001μS/cm

Measurement range (K cell=0.1) 0.00…19.99μS/cm 0.01μS/cm
Measurement range (K cell=1) 0.0…199.9μS/cm
200…1999mS/cm
2.0…19.99mS/cm
20…199.9mS/cm
0.1μS/cm
1μS/cm
0.01mS/cm
0.1mS/cm 
Measurement range (K cell=10) 200…1999mS/cm 1mS/cm

Accuracy (conductivity) instrument

±0.5% ±1digit

Measurement of resistivity by instrument

Measurement range (K cell=0.01)

Up to 1GΩ cm

 
Measurement range (K cell=0.1) Up to 100MΩ cm  
Measurement range (K cell=1)

5.0…199.9Ωcm
200…999Ωcm
1.00k…19.99kΩcm
20.0k…99.9kΩcm
100k…999kΩcm
1…10MΩcm

0.1Ωcm
1Ωcm
0.01kΩcm
0.1kΩcm
1kΩcm
1kΩcm
Measurement range (K cell=10)

0.5…5.0Ωcm

0.1Ωcm

Accuracy (resistivity) instrument

±0.5% ±1digit

Measurement of total dissolved solids

Measurement range (K cell=0.01)

0.00…1.999 mg/l

0.005 mg/l
Measurement range (K cell=0.1) 0.00…19.99 mg/l 0.05 mg/l
Measurement range (K cell=1) 0.0…199.9 mg/l
200…1999 mg/l
2.00…19.99 g/l
20.0…199.9 g/l   
0.5 mg/l
1 mg/l
0.01 mg/l
0.1 g/l
Measurement range (K cell=10) 100…999 g/l    1 g/l
Accuracy (TDS) instrument ±0.5% ±1digit
Measurement of salinity
Measurement range 0.000…1.999 g/l
2.00…19.99 g/l
20.0…199.9 g/l
1 mg/l
10 mg/l
0.1 g/l
Accuracy (salinity) instrument

±0.5% ±1digit

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.

[Home] [Top] [Distribution] [Contact Us]