| Measured values
|
pH - mV - χ - Ω - TDS - NaCl - mg/l O2 - %O2 - °C - °F |
Working temperature
Storing temperature
Working relative humidity
Protection degree |
-5
50°C
-25
65°C
0
90% RH without condensate
IP66 |
|
RS232Cserial interface
Type
Baud rate
Data bit
Parity
Stop bit
Flow control
Length of serial cable
USB interface
Type |
RS232C electrically isolated
Can be set from 1200 to 38400 baud
8
None
1
Xon/Xoff
Max 15m
1.1 - 2.0 electrically isolated |
|
Bluetooth interface
|
Optional for PCs fitted with Bluetooth input or HD
USB.KL1 Bluetooth / RS232 adapter. The interface can be installed in
Delta Ohm only. |
| Measurement of pH by
instrument |
|
Measuring range |
-9.999
+19.999pH |
|
Resolution |
0.01 o 0.001pH selectable from menu |
|
Accuracy |
±0.001pH ±1digit |
|
Calibration error @25°C |
|Offset| > 20mV
Slope > 63mV/pH or Slope < 50mV/pH
Sensitivity > 106.5% or Sensitivity < 85% |
|
Measurement of mV by instrument |
|
Measuring range |
-1999.9
+1999.9mV |
|
Resolution |
0.1mV |
|
Accuracy |
±0.1mV ±1digit |
|
Measurement of saturation index of dissolved oxygen |
| Measurement range
|
0.0
600.0% |
| Resolution |
0.1% |
| Accuracy instrument
|
±0.3% ±1digit (in the range 0.0
199.9%)
±1% ±1digit (in the range 200.0
600.0%) |
| Salinity
setting |
| Setting |
directly from menu or automatically by conductivity measurement
|
| Setting range
|
0.0
70.0g/l |
| Resolution |
0.1g/l |
|
Temperature measurement with the sensor inside the O2 probe
|
| Measurement range
|
0.0
50.0°C |
| Resolution |
0.1°C |
| Accuracy instrument
|
±0.1°C |
|
Measurement of temperature by instrument |
| Pt100 Measurement range
|
-50
+150°C |
| Resolution |
0.1°C |
| Accuracy instrument
|
±0.1°C ±1digit |
|
Measurement of conductivity by instrument
|
| |
Range |
Resolution |
|
Measurement range (K cell=0.01) |
0.000
1.999μS/cm |
0.001μS/cm |
| Measurement range (K cell=0.1)
|
0.00
19.99μS/cm |
0.01μS/cm |
| Measurement range (K cell=1)
|
0.0
199.9μS/cm
200
1999mS/cm
2.0
19.99mS/cm
20
199.9mS/cm |
0.1μS/cm
1μS/cm
0.01mS/cm
0.1mS/cm |
| Measurement range (K cell=10)
|
200
1999mS/cm |
1mS/cm |
|
Accuracy (conductivity) instrument |
±0.5% ±1digit |
|
Measurement of resistivity by instrument
|
| Measurement range (K cell=0.01)
|
Up to 1GΩ cm |
|
| Measurement range (K cell=0.1)
|
Up to 100MΩ cm |
|
| Measurement range (K cell=1)
|
5.0
199.9Ωcm
200
999Ωcm
1.00k
19.99kΩcm
20.0k
99.9kΩcm
100k
999kΩcm
1
10MΩcm |
0.1Ωcm
1Ωcm
0.01kΩcm
0.1kΩcm
1kΩcm
1kΩcm |
| Measurement range (K cell=10)
|
0.5
5.0Ωcm |
0.1Ωcm |
|
Accuracy (resistivity)
instrument |
±0.5% ±1digit |
|
Measurement of total dissolved solids
|
| Measurement range (K cell=0.01)
|
0.00
1.999 mg/l |
0.005 mg/l |
| Measurement range (K cell=0.1)
|
0.00
19.99 mg/l |
0.05 mg/l |
| Measurement range (K cell=1)
|
0.0
199.9 mg/l
200
1999 mg/l
2.00
19.99 g/l
20.0
199.9 g/l |
0.5 mg/l
1 mg/l
0.01 mg/l
0.1 g/l |
| Measurement range (K cell=10)
|
100
999 g/l
|
1 g/l |
| Accuracy (TDS) instrument |
±0.5% ±1digit |
| Measurement of salinity |
| Measurement range |
0.000
1.999 g/l
2.00
19.99 g/l
20.0
199.9 g/l |
1 mg/l
10 mg/l
0.1 g/l |
| Accuracy (salinity) instrument |