101A
ONE OR THREE PHASE POWER ANALYZER
|
|
|
101A One or three
Phase Power Meter
Analyzer.
101A Power Meter Analyzer
is suitable for
measurements on electronically switched loads.
Because of their DC-coupling capability correct RMS and power values
result at all times.
Take the case of line voltage multiplied by half wave rectified
current.
AC-coupling yields wrong RMS current and power. DC-coupling yields
correct values.
|
Power Meter Analyzer |
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
|
The 101A Power Meter
Analyzer is simple to use.
In the side menu you select
current- and voltage ranges (automatic or manual),
synchronization, measurement time, special function
"Peak-Hold", and display modes L1, L2, L3, SUM, and ALL (3-phase display).
In the display number fields you can place any measured quantity from
the measurement function selection table including harmonics of
current and voltage.
In the graphics field waveforms and bar graphs of current and voltage can be displayed.
|
|
|
A load connection box is available which connects your single- or three
phase load to the Power Analyzer and to the wall power outlet. Optional battery back-up is
available for 6 hours autonomous operation.
It prevents data loss, such as energy,
during power failure.
|
|
101A Power Analyzer
|
| Product |
Description |
| 101A-1 |
Low Cost Single Phase
Power / Energy Analyzer,
0.15% accuracy, peak hold function, tilt stand. |
| |
|
| 101A-3 |
Low Cost Three Phase Power / Energy Analyzer,
0.15% accuracy, peak hold function, tilt stand. |
|
|
101A Specification
|
|
Voltage |
Ranges, 4 ranges 1-3-10-sequence;
30V, 100V, 300V, 1000V, max. 600V |
- |
| Frequency range |
DC-50kHz |
Accuracy 23°K ±3°K;
10Hz-2kHz = ±(0.1% rdg + 0.1% range)
2kHz-10kHz = ±(0.8% rdg + 0.3% range) |
|
|
Current |
Ranges, 5 ranges 1-3-10-sequence;
600mA,2A,6A,20A,60A |
- |
Accuracy 23°K ±3°K;
10Hz-400Hz = ±(0.15% rdg + 0.15% range)*
400Hz-2kHz = ±(1.5% rdg + 1.0% range)
*Lowest range multiply percentage figures by 2.
|
|
|
Power |
Accuracy: Add accuracy percentage
figures of current and voltage, PF = 0 to ±1. |
- |
|
Measured
Quantities |
Vrms, V max, V peak hold, Arms, Amax,
A peak hold, power, apparent power, power factor, energy, apparent
energy, elapsed time, voltage- and current harmonics n=1 to 20. |
- |
|
Display |
Blue liquid crystal graphic display
with EL backlight. Displays values of each phase, total values, or
values of all three phases. Displays wave forms and bar graphs. |
- |
|
Dielectric |
Inputs to case, Line input to case: 2
kV/2 kV. |
- |
|
Dimension |
H x W x D = 132 x 185 x 267 mm.
Weight 2.5 kg. |
- |
|
Options |
RS-232 Interface and Windows
Operating Software
(95, 98, NT, 2000, ME, XP).
Battery back up for 6 hours operation with built in charger. |
- |
|
|
|