  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

106A ONE AND THREE PHASE POWER ANALYZERS


106A Power Analyzer. One and Three Phase Power Analyzer.
State of the art high performance power analyzers, single and three phase.
Unlike other instruments at this price level the 106A Power Analyzer is designed to cope with the extreme signals generated on frequency inverter drivers. You don't have to worry about the signal wave forms. 
106A Power Analyzer will always provide precise and reliable measurements.
The large and very bright monitor lets you read the display values from a distance up to 4m.

106A Power Analyzer

Power Analyzer.
Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com

From checking power of your coffee machine to determining the pertinent power parameters of a frequency inverter driven system is a simple task. You have all values displayed. In large letters by the way, well readable, even in dark rooms. The user menu makes operation easy. 
The measured values you can either print, send them to a PC via IEEE- or RS232 interface, or to a chart recorder via the analog outputs. You can have all available options installed in your instrument.

The 106A Power Analyzer will give you the highest performance at low costs. 
The analyzer inputs are all galvanically isolated, are broad band DC-300kHz, have a wide input range (0.3V - 1000V, 15mA - 40A), and have an exceptional common mode rejection for use in frequency inverter driven systems. The accuracy is 0.1% (0.05% versions are available). 
The bright LCD monitor displays up to 10 measured values in well legible 9mm high numbers. 
The Three Phase Power Analyzer puts up to 32 measured values on the screen. 
You have the choice to visualize wave forms, bar graphs or trend plots. 
A unique feature of these instruments allows a combination of meter mode and graphic mode.

Manufacture is dedicated to power measurement and will continue to add features to the 106A. 
You can purchase upgrades at modest costs. 
The Power Analyzers are warranted one year and are supplied with a traceable calibration certificate. 

106A Power Analyzer
Product Description
106A-1/0.1 Single Phase Power Analyzer, 0.1% accuracy
106A-1/0.05 Single Phase Power Analyzer, 0.05% accuracy
106A-1/nh Single Phase Power Analyzer, without harmonics without graphics,
 0.1% accuracy
106A-3/0.1 Three Phase Power Analyzer, 0.1% accuracy
106A-3/0.05 Three Phase Power Analyzer, 0.05% accuracy
106A-3/nh Three Phase Power Analyzer, without harmonics without graphics, 
0.1% accuracy
Accessories
Accessory Description
ACS1 Current clamp with connector to 106A shunt input; 0-400A, DC-1kHz, 2%
ACS2 Portable printer (106 x 180 x 88mm) with centronics interface and cable (400gr.)
ACS3 Soft carrying case for 106A
ACS4 Set of test leads, max. 32A, 1.5m (2 red; 2 black)
ACS5 Shunt input connector
ACS6 Service Manual PC10
ACS7 Rack Mounting Kit
ACS8 Official Calibration Certificate
Options
Option Description
01 RS-232 Interface and Centronics printer output
02 RS-232- and IEEE-488 Interface, Centronics printer output
03 RS-232- and IEEE-488 Interface, Centronics printer output, 4 analog outputs,
8 analog inputs
03A RS-232- and IEEE-488 Interface, Centronics printer output, 4 analog outputs
with provision to also output total power, 8 analog inputs
04 3-Phase current sensor module 0-100A (supply by 106A)
05 Operating software under DOS to control 106A, read data and store data via
RS-232 or IEEE-488 (National-, Keithly interface)
06 106A driver for Nat. Instrument LabView
07 Standalone 106A software based on LabView
08 TTL-input for external synchronization
09 Network to form artificial neutral (mainly frequency inverter measurements)
10 Nonvolatile memory to save instrument settings:
Single phase: complete setting
Three phase: IEEE-488 address only
Must be factory installed on Opt. 01,02 or 03
106A Specification
 
Voltage Ranges, 8 ranges 1-3-10-sequence;
0.3V, 1V, 3V, 10V, 30V, 100V, 300V, 1000V
-
Frequency range DC, 0.1Hz-300kHz
Crest Factor 4:1 at 50% full scale (fs)
Input Impedance >1MΩ
Common Mode - 50Hz/100kHz 155dB/95dB
Standard accuracy 23°K ±3°K; rms, mean, rectified mean; for 0.3, 1V range, input >50% fs
1Hz-1kHz =           ±(0.1% rdg + 0.1% range)
DC, 1kHz-10kHz = ±(0.2% rdg + 0.2% range)
10kHz-100kHz =    ±(0.3% range + 0.04%/kHz rdg) *
100kHz-300kHz =  ±(0.3% range + 0.04%/kHz rdg), typical
improved accuracy
±(0.05% rdg + 0.05% range )

* 0.3V range typical
Current Ranges, 10 ranges 1-3-10-sequence; 
15mA,50mA,150mA,500mA,1.5A,5A; 1,3,10,30,100,300A
-
Frequency range DC, 0.1Hz-300kHz
Crest Factor 4:1 at 50% full scale (fs)
Common Mode - 50Hz/100kHz 160dB/120dB
Standard accuracy 23°K ±3°K
1Hz-1kHz
DC, 1kHz-10kHz
10kHz-100kHz
100kHz-300kHz
5A-/Shunt input¹
±(0.1% rdg + 0.1% range)
±(0.2% rdg + 0.2% range)
±(0.3% rdg + 0.04%/kHz rdg)
±(0.3% rdg + 0.04%/kHz rdg)
30A input¹
±(0.1% rdg + 0.1% range)
±(0.9% rdg + 0.2% range)
±(0.3% rdg + 0.5%/kHz rdg)
,typical
improved accuracy 1Hz-400Hz
±(0.05% rdg + 0.05% range)

¹for 2 lowest ranges, input
>50% fs
Power 80 Ranges corresponding to the products V x A -
Frequency range DC, 0.1Hz-300kHz
Accuracy 23°K ±3°K
1Hz-1kHz                 Add accuracy percentage figures of current and voltage
DC, 1kHz-10kHz       input and add 0.04%/kHz of Vrms x Arms / PF
10kHz-100kHz

PF = 0 to ±1
PF = 0 to ±1
PF = 1
Frequency 0.1Hz-300kHz, A or V triggered; Accuracy ±0.1% -
Computed Values Accuracy; Reactive Power, Var = ±(VA² - W²)½; 
Apparent Power: VA = Arms Vrms;
Power Factor: PF = W/VA; 
Crest Factor: CF = Ap/Arms, Vp/Vrms; 
Form Factor: FF = At/Arms, Vt/Vrms; 
Impedance: Z = Vrms/Arms;
Total Harm Dist: THD = (Irms² - IFund²)½/Irms
Add accuracy percentage
figures of values involved
in computation.
Integrator Energy, Charge; Accuracy - Wh, VAh, Varh, Ah; Basic accuracy of integrated quantity. -
Harmonic Analysis Frequency range of fundamental - 2.5Hz-100kHz -
Range of harmonic 1-99
Accuracy, Harmonic current and voltage
2Hz-1kHz =        ±(0.2% rdg + 0.1% range)
1kHz-10kHz =     ±(0.5% rdg + 0.5% range)
10kHz-100kHz = ±(0.7% range + 0.1%/kHz rdg), typical
-
Display Blue liquid crystal graphic display with FL backlight; - 64x120mm; - 128x240 pixels -
Power AC, 50-400Hz; Fuse: Power 85V-240V; 2AF/30VA
Dielectric Strength Inputs to case or power supply
Line input to case
Input to Input
2.5kV/50Hz/1 minute
1.5kV/50Hz/1 minute
4kV/50Hz/1 minute
Dimension H x W x D; Weight 150 x 235 x 320mm; 4kg
Options IEEE-488-2, RS232, Centronics printer output
4 Analog outputs, Output impedance 100Ω
4 Analog inputs, low range, input impedance 200kΩ
4 Analog inputs, high range, input impedance 200kΩ
Rack mounting kit
Humidity: KYG according to DIN 40040, max. 85% RH non-condensing

0 - ±5V
0 - ±1V
0 - ±10V
Shunt Input
HiLo Ranges in mV: 60, 60(10)½, 600, 600(10)½, 6000, 6000(10)½
Accuracy : same as 5A-input
Input impedance: 200kΩ; input of 60mV corresponds to 1.0000A display.
-