| INFRATEK
AG
|
|
HAND HELD SINGLE & THREE PHASE POWER ANALYZER |
|
|
|
Hand Held Single and Three Phase Power Analyzer |
|
|
MODEL 31 Hand Held
Battery Operated Single and Three Phase Power Analyzers are
valuable engineering tools designed to offer precision measurement on
all types of signals including electronic drives.
The Model 31 hand-held Power
Analyzer are sophisticated measuring instruments, single phase and
three phase, designed for laboratory and field use. Unlike other
instruments in this price range the
Model 31 Power Analyzers are
suitable for measurements on frequency inverter drives, light ballasts,
and other electronically controlled loads. The wide frequency range, the
high common mode rejection, and the fact that you can measure currents
of up to 50A directly without having to use clamps guarantees excellent
accuracy for all types of measurements.
|
Handheld Power Analyzer |
Main
Features
-
100 mA-50
A, DC to 300 kHz
-
1 V-1000
V, DC to 300 kHz
-
0.3 W-150
kW, no CT's required
-
Suitable
for frequency inverter drives
-
Measures
RMS, power-, and harmonic values
-
Line-to-line
voltage, torque, efficiency
-
Harmonics
1-63, IEC1000-3-2
-
Data
logging for dynamic processes
-
Accuracy
grades 0.1%, 0.2% and 0.4% (low cost)
|
|
|
|
|
|
|
Processing
Power
-
Simultaneous
measurements in 3 phase system
-
All data,
including harmonics in real time
-
Efficient
data transfer to computer
|
Interface -
Input/Output
Display
|
The Model
31 Power Analyzer are controlled by six menu soft keys M1 through M6,
five cursor soft keys, and one power-on/off soft key. The operating
procedure, to configure the display, to configure the interface, to
set scaling factors, and to select the input, the ranges, and many
more functions are self-explanatory. If you desire you can store your
personal instrument configuration in nonvolatile memory and have the
unit start-up with your personal configuration at power-on.
Efficient data processing makes simultaneous measurements in a three
phase system possible.
Three True RMS line voltages, True RMS line-to-line voltages, True RMS currents,
power, power factor, minimum, maximum, harmonics of 3 currents and 3
voltages, and all derived quantities are determined from the same
measurement interval (minimum 250ms). You have the choice to display
phase values including wave forms or bar graphs, or you can display
total and average values in the 3 phase system.
You can also display the values of all three phases including their
sums, and averages.
|
The Model 31 Power Analyzer includes a data logging mode to transfer
data from a dynamic process to a personal computer for further analysis.
Also, current harmonics according to IEC1000-3-2 in a 3 phase system are
determined. By means of the operating software simple pass-fail criteria
are generated.
|
|
Model 31 High Speed Power Analyzers
|
| Product |
Description |
| Model 31-1/0.3L |
Economy Single Phase High Speed Power Analyzer,
0.3% accuracy, saves 10 instrument settings, no data
logging, no IEC1000-3-2 |
| Model 31-1/0.3 |
Single Phase High Speed Power Analyzer, 0.3%
accuracy, saves 10 instrument settings, includes data logging, no
IEC1000-3-2 |
| Model 31-1/0.1 |
Single Phase High Speed Power Analyzer, 0.1%
accuracy, saves 10 instrument settings, includes data logging
and IEC1000-3-2 |
| Model 31-1/0.05 |
Single Phase High Speed Power Analyzer, 0.05%
accuracy, saves 10 instrument settings, includes data logging
and IEC1000-3-2 |
| Model 31-3/0.3L |
Economy Three Phase High Speed Power Analyzer,
0.3% accuracy, saves 10 instrument settings, no
line-to-line measurement, no data logging, no
IEC1000-3-2 |
| Model 31-3/0.3 |
Three Phase High Speed Power Analyzer, 0.3%
accuracy, saves 10 instrument settings, includes line-to-line
measurement and data logging, no IEC1000-3-2 |
| Model 31-3/0.1 |
Three Phase High Speed Power Analyzer, 0.1%
accuracy, saves 10 instrument settings, includes line-to-line
measurement, data logging, and IEC1000-3-2 |
| Model 31-3/0.05 |
Three Phase High Speed Power Analyzer, 0.05%
accuracy, saves 10 instrument settings, includes line-to-line
measurement, data logging, and IEC1000-3-2 |
| Model 31-3/0.05M |
Three Phase High Speed Power Analyzer for
dynamic motor testing (torque transducer required), 0.05%
accuracy, saves 10 instrument settings, includes line-to-line
measurement, dynamic torque measurement, IEC1000-3-2,RS-232
interface, frequency/analog in-out, data logging and operating
software for motor testing. |
|
|
Accessories
|
| Accessory |
Description |
| ACS1 |
Current clamp with connector to Model 31 clamp
input; 0-400A, DC-1kHz, 2% |
| ACS2 |
Current clamp with connector to Model 31 clamp
input; 0-200A, DC-20kHz, 1% |
| ACS3 |
Soft carrying case for Model 31 |
| ACS4 |
Set of test leads, max. 32A, 1.5m (2 red; 2
black) |
| ACS5 |
Clamp input connector |
| ACS6 |
Service Manual Model 31 |
| ACS7 |
Rack Mounting Kit |
| ACS8 |
Official Calibration Certificate from SCS
(Swiss Calibration Service) |
|
|
|
|
Options
|
| Option |
Description |
| 01 |
High Speed RS-232 Interface, external
synchronization input, Windows Operating Software (95,
98, NT) without data logging capabilities, without
IEC1000-3-2 |
| 02 |
High Speed RS-232 Interface, external
synchronization input, trigger input, Windows Operating
Software (95, 98, NT) including software for high speed
data logging and IEC1000-3-2 testing |
| 03 |
One frequency input with selectable
frequency range, nine analog inputs (±5V / ±10V), ten
analog outputs(±5V) |
|
|
|
|
|
Power Analyzer Specification
|
|
|
| Voltage |
7
ranges: 1V, 3V, 10V, 30V, 100V, 300V, 1000V |
- |
| Frequency
range: DC-100kHz ¦ Coupling: AC/AC+DC |
1Hz-100kHz
/ DC-100kHz |
| Crest
Factor 3:1 at 50% full scale ¦ Common Mode 50Hz / 100kHz |
130dB/70dB |
| Built-in
star point network 500kΩ ¦ Maximum Input: Hi to Lo / Lo to
case |
1000V /
600V |
Accuracy
23°K ±3°K; rms, rdg= reading
1Hz-1kHz
1kHz-10kHz
10kHz-100kHz |
±k(0.1% rdg + 0.07% range)
±k(0.3% rdg + 0.2% range)
±k(0.05%/kHz rdg + 0.3% range) |
|
Accuracy Grades:
k=1, k=2, k=4 |
| Current |
10
ranges: 100mA,300mA,1A,3A,10A; 1,3,10,30,100A |
Clamp:
1A - 1000A |
| Frequency
range: DC-100kHz ¦ Coupling: AC/AC+DC |
1Hz-100kHz
/ DC-100kHz |
| Crest
Factor 3:1 at 50% full scale ¦ Common Mode 50Hz / 100kHz |
150dB/100dB |
| 3A
input: 3A cont./10A 5s; 50A input: 40A cont./50A 20s. |
Ri=30mΩ
/ 3mΩ |
Accuracy
23°K ±3°K;
rdg= reading, rng= range
1Hz-500Hz
500Hz-2kHz
2kHz-10kHz
10kHz-100kHz |
50A/Clamp input
±k(0.1% rdg + 0.1% rng)
±k(0.6% rdg + 0.6% rng)
±k(0.8% rdg + 0.8% rng)
±k(0.1%/kHz rdg + 0.8% rng)* |
3A input
±k(0.1% rdg + 0.1% rng)
±k(0.8% rdg + 0.8% rng)
±k(2% rdg + 1% rng)*
±k(0.1%/kHz rdg + 2% rng)* |
Lowest range multiply percentage
figures by 2. ¦ *typical |
Accuracy Grades:
k=1, k=2, k=4 |
| Power |
70
Ranges corresponding to the products of voltage ranges time
current ranges. |
DC-100kHz |
Accuracy
23°K ±3°K; 50A/Clamp input, 3A input
1Hz-1kHz Add accuracy percentage figures
of current and voltage input
1kHz-100kHz Add accuracy percentage figures of current and
voltage input,
add ±2% (1-PF)/kHz of
range |
PF = 0 to ±1
|
| Computed
Values |
Reactive
Power, Var = ±(VA² - W²)½;
Apparent Power: VA = Arms Vrms;
Power Factor: PF = W/VA;
Crest Factor: CF = Ap/Arms, Vp/Vrms;
Maximum = Ap, Vp;
Minimum: -Ap, -Vp; PtP:
Maximum - Minimum;
Impedance: Z = Vrms/Arms;
Total Harm Dist: THD1 = (Irms² - IFund²)½/Irms
THD2 = (I2_rms² + I3_rms² + In_rms²)½/Irms |
Add
accuracy percentage
figures of values involved
in computation. |
| Mechanical
Values |
Total
input power, Nm/s; Output power to load, Nm/s; Torque at axis of
rotating
machine, Nm; Speed, rpm; Efficiency: Eta = Output power / Input
power
|
Synchronous
machines only.
|
| Harmonic
Analysis |
Frequency
range of fundamental ¦ 4Hz-50kHz |
Harmonic
1-63 |
| Accuracy:
harmonic current and voltage, same as RMS current and True RMS voltage |
- |
Computed
values: harmonic power, harmonic phase angle (power factor)
harmonic impedance. |
Add
accuracy percentage
figures of values involved
in computation. |
| Frequency |
2Hz-100kHz;
A; V: ±0.1% |
- |
| Integrator |
Energy,
Accuracy: Wh, VAh; Basic accuracy of integrated quantity. |
- |
| Data
Logging |
Output
values, speed and duration is programmable; maximum speed:
10 values from a 3-phase system in 150ms. |
Range
of signal frequency 5Hz
to 1kHz |
| Current
harmonics |
Current
harmonics in a 3-phase system are determined according to IEC1000-3-2 |
|
Display
Power/Battery
Dielectric
Strength
Dimensions |
Display:
Blue liquid crystal graphic display with FL backlight; - 64 x
120mm; - 128 x 240 pixels
Charger: AC, 50-60Hz, 115/230V, 0.2AF, 10VA; sealed Lead gel
battery, 6 hours
Dielectric Strength: Current inputs to case; Voltage inputs to
case; Line input to case
Dimensions: H x W x D |
58x108mm;
128x240 pixels
3kV; 2kV; 3kV 50Hz/1 min
75x160x245mm; 1.5kg |
| Options |
RS232
interface
Windows operating software, 95, 98, NT
Current clamps 0-200A / 0-400A
|
- |
|
|
|