Toll Free: 866-436-2430                     Email: sales@hotektech.com
 
 
Instruments, calibrators, standards, analyzers, dataloggers. analyzers
 
  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

BLACKBODY FIXED POINT TEMPERATURE CALIBRATOR


Medusa R 999 Blackbody Fixed-Point Temperature Calibrator. Temperature range 30°C to 550°C.
The Medusa R is a blackbody primary source in which radiation pyrometers can be calibrated. Specifically designed to accommodate black body fixed point cells, the Medusa-R is the optimal choice for primary labs needing the highest levels of radiation measurement.

The temperature of the cavity can be set to ±0.1°C over the temperature range 30°C to 550°C. 
The controller reaches 550°C approximately 45 minutes from switch on. 
Also, the cavity temperature, as measured by a radiation thermometer took approximately 45 minutes to reach stability from switch on.

MEDUSA R - Blackbody Primary Source

Blackbody Fixed-Point Temperature Calibrator
Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com

The variation seen on the controller’s temperature indication over a 5 minute period was ±0.2°C. 
Similar variations were detected by a radiation thermometer looking into the cavity.
Using a portable radiation thermometer having a target diameter of 13mm, the 48mm target was surveyed. 
Maximum temperature differences of ±1°C were found for set temperatures in the range 100°C to 500°C.
Temperatures along the inner 200mm of the 285mm long cavity were measured at 400°C and 500°C, using a hand held fiber-optic radiation probe. Maximum temperature differences of ±4°C were found.

Large Fixed Point Cells can be fitted to the Medusa R giving small uncertainties suitable for Primary Standards.

Model 999
Temperature Range

30ºC to 550ºC

Emissivity greater than 0.995
Stability ±0.05°C
Display Resolution

0.01°C to 99.99; 0.1°C from 100.0 to 550.0

Cavity Size

45mm diameter
285mm deep

Time to Temperature 45 minutes
Power

110 VAC ±10% (50/60Hz)
1000 Watts typical

Dimensions

480mm x 425mm x 260mm (HxWxD)

Weight 37.5 lbs (17kg)

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.

[Home] [Top] [Distribution] [Contact Us]