TEMPERATURE CALIBRATOR DRY BLOCK
Oceanus-6 Dry Block Liquid Bath Blackbody Temperature Calibrator Range -25ºC to 110ºC

Oceanus-6 Temperature Dry Block Calibrator Range -25ºC to 110ºC
The Oceanus-6 Temperature Dry Block Calibrator has all the advantages of the Isocal-6 models but with a substantially larger calibration volume, 2 inch diameter by 11.8 inches deep (52mm Ø x 300mm). 
The Oceanus-6 can be used as a Dry Block, a Liquid Bath, a Blackbody Source for infrared thermometers, a Surface Sensor Calibrator and for performance to a few milliK (0.001ºC) ITS-90 Fixed Points. Like all Isotech calibrators Oceanus-6 is available in two models, the BASIC (B) version with one controller, and the SITE (S) version with onboard secondary display for high accuracy digital probe matching.

OCEANUS 6 - P.O.T.T.S Temperature Calibrator

Temperature Dry Block Calibrator
  • Universal Input Temperature Indicator
    Included in S Model to allow for 
    programming of up to five “correction” points

  • ITS-90 Fixed Points
    Can be used with the Gallium and 
    Triple Point of Water Fixed Point Cells

  • Dry or Liquid Bath

  • Windows Software and PC Interface

  • High Capacity Deep Block for Very Long Thermometers

  • Ideal for Laboratory Comparison and 
    Fixed-Point Calibrations

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.
Increased resolution of ±0.01 available throughout the range via the PC interface and 
from  -19.99 to +99.99 locally on the auto-ranging front display. 
The controller features multi-point block to display correction providing excellent absolute accuracy.
The SITE model has a universal sensor input allowing Platinum Resistance Thermometers, Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs
including 4-20mA current transmitters to be displayed on the indicator. 
The indicator can be programmed with up to five calibration points to provide high accuracy digital
probe matching. 
The indicator and controller are both addressable over the communications link.

Temperature Range 45°C below ambient to +110°C

Absolute Minimum
Temperature

-45°C

Absolute Stabilities
Over 30 Minutes:

Metal Block Bath:
Stirred Liquid Bath:
Ice/Water Bath:
Black Body Source:
Surface Sensor Calibrator:
ITS-90 Fixed Point Apparatus:

±0.03°C
±0.025°C
±0.001°C
±0.3°C
±0.5°C
±0.0002°C

Stabilization Times 10 minutes
Calibration Volume 52 dia x 300mm deep
Uniformity ±0.018°C
Controller Resolution 0.1 to 0.01 (4 digit display)
Indicator Resolution 0.1 to 0.01 (4 digit display)
Indicator Units °C, °F, K
Communications

Supplied as standard with serial interface,
PC adapter cable and Cal Notepad. 

Power 300W, 110 VAC ±10%, 50/60Hz
Overall Dimensions 430mm x 310mm x 300mm (HxWxD)
Weight 37.48 lbs (17kg)