| Model |
QuickCal-560 |
| Temperature
Range |
-10°C to +140°C in
an ambient of 20°C |
| Set
Point Resolution |
0.1°C over range |
| Accuracy |
±0.2 to ±0.4°C
using Comparison
Techniques
±0.4°C against display temperature with a single probe in the
4.5mm pocket |
| Stability |
±0.1°C |
| Time to
Set Point |
9 minutes from 0°C
to 100°C |
|
Additional
Time
for Best Stability |
Typically 5 minutes |
| Calibration
Volume |
Two 13mm diameter
pockets 120mm deep
Supplied with two sleeves, one blank and one drilled 6.8mm
diameter 120mm deep |
| Power |
15V DC or
100-240V, Switch mode power supply (50 / 60 Hz) |
| Overall
Dimensions |
65mm x 152mm x 175mm
(HxWxD) |
| Weight |
3.3 lbs (1.5kg) |
|
| Model |
QuickCal-550 |
| Temperature
Range |
+30°C to +350°C in
an ambient of 20°C |
| Set
Point Resolution |
0.1°C over range |
| Accuracy |
±0.2 to ±0.4°C
using Comparison
Techniques
<100°C ±0.4°C against display temperature with a single
probe in the 4.5mm pocket
>100°C ±1.5°C against display temperature with a single
probe in the 4.5mm pocket |
| Stability |
±0.1°C |
| Time to
Set Point |
9 minutes from
Ambient to 350°C |
|
Additional
Time
for Best Stability |
Typically 5 minutes |
| Calibration
Volume |
25mm diameter x
120mm deep
Supplied with insert, 5 pockets, 8mm,
6.5mm, 6.5mm, 4.5mm, 4.5mm, all
115mm deep |
| Power |
110 VAC ±10% (50 /
60 Hz) |
| Overall
Dimensions |
65mm x 152mm x 175mm
(HxWxD) |
| Weight |
3.3 lbs (1.5kg) |
|