| Model |
TTI-8 |
| Temperature
Range |
-200°C to 962°C |
| Accuracy |
±0.01°C over full
range |
| Set
Point Resolution |
0.001°C over range |
| Repeatability |
±2 least
significant digits (±0.002°C), included in accuracy |
| Stability |
±0.005°C per year |
|
Temperature
Co-Efficient |
<0.00005°C per
°C ambient change |
| Data
Entry Format |
ITS90
coefficients, Calendar van Dusen coefficients, or EN60751 for
un-calibrated probes. |
| Probe
Current |
1mA constant current
source |
| Probe
Types |
Pt100 to
EN60751 with nominal Ro=100 and 25 high-alpha
probes up to 0.00392 |
|
Maximum
Cable
Length |
100 meters of 4
core 19/0.15 SPC/PTFE screened cable. |
|
Thermometer
Input
Connectors |
(2 ch.) or
(8 ch.) 5 pin industrial DIN socket can be used with passive or
proprietary SMART
connector. |
| Communications |
Galvanically
isolated RS232C: 4KV isolation. Baud rate 9600. |
| Operating
Conditions |
0°C to 50°C,
10-90% RH non-condensing. |
| Power |
90 to 264VAC
universal IEC input, (50/60 Hz) 30VA max. |
| Overall
Dimensions |
101.6mm x 152.4mm x
203.2mm (HxWxD) |
| Weight |
2.2 lbs (1kg) |