1404 CAPACITANCE STANDARD CAPACITOR
|
|
|
1404 Standard
Capacitor. Standard Capacitance.
These capacitors have been
designed as primary reference standards of capacitance with which
working standards can be compared.
The 1404 Series are the
standard of choice in metrology labs, and still used today by
standards bodies around the world.
|
|

|
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
|
-
Available in 10, 100, and
1000pF
-
20 ppm per year stability
-
Coaxial connections
-
Hermetically sealed in dry
nitrogen
-
1404-9701
1404-A, 1000pF
-
1404-9702
1404-B,100pF
-
1404-9703
1404-C, 10pF
|
In combination with an
accurately known external resistor, this capacitor also becomes a
standard of dissipation factor.
All critical parts of the plate assembly are made of invar for
stability and low temperature coefficient.
After heat cycling and adjustment the assembly is mounted in a heavy
brass container, which, after evacuation, is filled with dry nitrogen
under pressure slightly above atmospheric and sealed.
The container is mounted on an aluminum panel and protected by an
outer aluminum case.
Each capacitor is subjected to a series of temperature cycles to
determine hysteresis
and temperature coefficients and to stabilize the capacitance.
Two locking G874 coaxial connectors are used as terminals.
The outer shell of one is connected to the case, but the outer shell
of the other is left unconnected to permit the capacitor to be used
with an external resistor as a dissipation -factor standard.
|
Calibration: A
certificate of calibration is supplied with each capacitor, giving the
measured direct capacitance at l kHz and at 23°±1°C. The measured
value is obtained by a comparison to a precision better than ±1ppm
with working standards whose absolute values are known to an accuracy
of ±5ppm, determined and maintained in terms of reference standards
periodically measured by the National Institute of Standards and
Technology.
Adjustment Accuracy: The capacitance is adjusted before
calibration with an accuracy of ±5ppm to a capacitance about 5ppm
above the nominal value relative to the capacitance unit maintained by
the reference standards.
Stability: Long term drift is less than 20 ppm per year.
Maximum change with orientation is 10ppm and is completely reversible.
Temperature Coefficient of Capacitance: 2 ±2ppm/ °C for
1404-A and -B, 5 ±2ppm/ °C for 1404-C, from -20°C to +65°C. A
measured value with and accuracy of ±1ppm/ °C is given on the
certificate.
|
|
|