Instruments, calibrators, standards, analyzers, dataloggers. analyzers
 
  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  

STANDARD CAPACITOR CAPACITANCE STANDARD


1409 Capacitance Standard. Standard Capacitor.
Highly stable cost-effective capacitance standards with low temperature coefficient, low losses and a wide range of values.

Features

  • 0.001 µF to 1000 µF

  • ±0.01 %/year stability

  • Verify meter and instrumentation calibration

  • Two-to-five terminal configuration, depending on model

  • Calibration Certificate Traceable to NIST 

Adjustment Accuracy: within ±0.05% of nominal value.
Calibration Accuracy: 0.01% for 1 µF; 0.012% for 5 µF; 0.02% for 10 µF,
0.035% for 100μF; and 0.4% for 1000 µF
Stability: <0.01% per year.
Temperature Coefficient: <+35 ±10ppm for capacitance ≤1μF;
<-60±50 ppm/
°C for capacitance of 10 µF;
<±110 ppm/
°C for capacitance over 10 µF.
Operating Temperature: 10
°C to 50°C.
Dissipation Factor: 0.01 µF - 1 µF: 0.0003 @ 1 kHz;
10 µF: 0.0005; 100μF: 0.001;
1000 µF: 0.002 @ 100 Hz and 120 Hz;
0.02 @ 1 kHz.
Series Inductance: Typically < 0.06 μH, 0.01 µF - 1μF.
Series Resistance @1 MHz: 0.02 Ω, 0.01 µF - 0.1 µF; 0.03 Ω, 1 µF.
Leakage Resistance: 5000 ohm-Farads or 100 GΩ, whichever is less.

Model Value Accuracy Dissipation
Factor
(typical)
Maximum
Voltage**
(V)
1409-F
1409-L
1409-T
1409-Y
1409-10μF
1409-100μF
1409-1000μF
1409-X
 
1 nF
10 nF
100 nF
1 µF
10μF
100 µF
1000 µF
Custom
±0.05%
±0.05%
±0.05%
±0.05%
±0.02%
±0.04%
±0.4%
*
0.0003
0.0003
0.0003
0.0003
0.0005
0.001
0.001
*
500
500
500
500
44 Vrms+
22 Vrms+
22 Vrms+
*

+ Maximum allowable Vrms; subject to maximum Vdc = 50 V and max
Vrms = (39000/f) for C = 10 μF; (26000/f) for C = 19 μF; (13000/f) for
C ≥ 100 μF; (9500/f) for C ≥ 1000 μF, where f = frequency (in Hz).
* Depends on Custom value
** Peak up to 10 kHz.

Catalog Number                       Item
1409-9706                              1409-F,    0.001μF
1409-9712                              1409-L,    0.01μF
1409-9720                              1409-T,    0.1μF
1409-9725                              1409-Y,   1.0μF
1409-9730                              1049,       10μF
1409-9735                              1409,       100μF
1409-9740                              1409,       1000μF
1409-X                                   1409,       Custom Value

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.

[Home] [Top] [Distribution] [Contact Us]