  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

1413 CAPACITANCE DECADE BOX


1413 Series Decade Capacitor are made to the same exacting specs as GenRad. 
The 1413 Capacitance Decade is not only a capacitance precision standard, it is a systems component as well - connections are made at the rear for this purpose.

Six precision decades are employed to provide a range of 0 to 1.11111µF in increments as small as 
1 pF and with an accuracy of 0.05% +0.5pF. Air capacitors are used for the two lower decades and precision silvered-mica capacitors are used for the remainder. The lower four decades contain adjustments that are factory set but accessible for readjustment later if desired.
Laboratory Standard, Working Standard, Calibrating Working Standards, Standard for Dissipation Factor,
Verification & Calibration of LCR Meters, Capacitance Measurement Functions, Verification of Calibration of Multimeters, For Calibrating Instrumentation

Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com
  • 0 to 1.11111 µF

  • 0.05% + 0.5 pF basic accuracy

  • Step 1 pF

  • 6-digit resolution

  • 3-terminal connections

  • Optional rear-mounted GR-874 locking connectors.

The shielding is divided into two parts, arranged to provide low terminal-to-guard capacitances. 
When the two shields are connected together, the 1413 becomes a well-shielded three-terminal capacitor with an extremely low zero-capacitance, suitable for a variety of applications.

Range: 0 to 1.11111µF, controlled by six in-line-readout dials.
Accuracy: ±(0.05% +0.5 pF) at 1 kHz.
Stability: ±(0.05% +0.5 pF) at 1 kHz 
Temperature Coefficient: 20ppm/ °C from 10 to 50°C.
Zero Capacitance: 0.1 pF.
Voltage Rating: 500 Vpk max up to 10 kHz.

 

1 pF to
100 pF

101 pF to
1000 pF

1001 pF to
2000 pF

2001 pF to
0.1 µF

0.1 uF
1.11111 µF

Dissipation Factor,
max at 1 kHz

0.002 0.001 0.0005 0.0003 0.0004

Insulation Resistance, 3
terminal after 2 minutes
at 500V dc

5 x l010 5 x l09

Terminal
Capacitance, max
high to case
high to guard
low to guard



4 pF
85 pF
45 pF



8 pF
110 pF
70 pF



10 pF
125 pF
80 pF



30 pF
165 pF
110 pF



60 pF
200 pF
120 pF