1863 1864 HIGH RESISTANCE MEGOHMMETER |
|
|
|
1863 1864 High Resistance Megohmmeter |
|
|
1863 High
Resistance Megohmmeter. Test Voltage up to 500 VDC.
1864 High Resistance Megohmmeter. Test Voltage from 10 VDC to
109 VDC and to 1090 VDC.
The GenRad 1863 1864 High
Resistance Megohmmeter is the choice for production and inspection
tests.
The 1863 Megohmmeter
is the choice for
production and inspection tests. The 1863 will measure resistance at
any of five common test voltages up to 500VDC, has fewer controls,
and is the lower priced model. It is the best selection for those
applications requiring a lesser voltage and resistance range.
The GenRad 1864
Megohmmeter is the choice for more demanding applications. It is the
more flexible of the two instruments. The test voltage can be set to
any value from 10VDC to 109VDC in 1-volt steps and to 1090VDC in 10
volt steps. Thus the GenRad 1864 can be set to any common, or
uncommon test voltage for ceramic, mica or paper capacitors, or
other devices. The reverse resistance of rectifiers can be readily
measured; the low test voltages are especially useful in measuring
solid state diodes.
An additional range on the 1864 Megohmmeter
permits measurements up to 200 TΩ (2 x 1014Ω)
|
|

|
-
Rugged and Portable
Analog Megohmmeter
-
3% Basic Accuracy
-
Up to 200 Test
Voltages from 10 V to 1090 VDC
-
50 kΩ
to 200 TΩ
Measurement Range
-
Maximum of 2 mA
Charge Current
-
Analog Output
-
Large, Direct
Reading Front Panel Meter
|
|
|
|
|
|
|
|
Both megohmmeters are easy to
use with direct-reading meter indication and lighted range
switch that shows the multiplier for each range and voltage.
The maximum current possible at the terminals is limited to
a 5mA and a panel light near the terminals warns when
voltage is present..
|
|
Stable power supplies
and feedback voltmeter circuit minimize drift and
time wasting adjustments.
Guard and ground terminals permit measurement of
grounded or ungrounded two- or three-terminal
resistors.
The instruments are supplied in a
convenient, portable, Flip-Tilt case that is a stand
for the meter in use and protects
it in transit and storage.
|
|
Resistance Accuracy:
1863:(min reading 0.5)
Range 1-5: ± 2(meter reading +1)%
Range 6: ± 2% to accuracy above
Range 7: ± 4% to accuracy above
1864:(min reading 0.5)
Range 1-5: ± 2(meter reading +1)%
Range 6: ± 2% to accuracy above
Range 7: ± 3% to accuracy above
Range 8: ±5% to accuracy above
Voltage Accuracy:
(across unknown): +/-2%
Short-Circuit Current:
5mA approximately
Input Terminals: Front Panel Mounted:
4 Binding Posts
(+) unknown (red)
(-) unknown (red)
guard (red)
ground (silver)
|
Display: Analog meter
Caution High Voltage warning indicator
Dimensions: Flip Tilt Case
(w x h x d) : (9.63 x 10 x 6.75 inches)
(245 x 254 x 172 mm)
Weight: Approx. 4.4 kg (9.5 lbs) - Net
Approx. 7.0 kg (14 lbs) - Shipping
Power: 100 to 125V or 200 to 250V,
50-400Hz, 13W
Environmental: Operating: 0°C to +40°C, stated
accuracy <70% RH
Storage: -20°C to +60°C, <80% RH
|
|
|
|
1863 Voltage and Resistance Ranges |
|
|
| Voltage |
Rmin
(Full Scale) |
Rmax |
Useful
Ranges |
| 50VDC, 100VDC |
50 kohm |
2 Tohm |
7 |
| 200, 250 & 500 VDC |
500 kohm |
20 Tohm |
7 |
|
|
|
|
1864 Voltage and Resistance Ranges |
|
|
|
Voltage |
Rmin
(Full Scale) |
Rmax |
Useful
Ranges |
| 10VDC to 50VDC |
50 kohm |
2 Tohm |
7 |
| 50VDC to 100VDC |
200 kohm |
20 Tohm |
8 |
| 100VDC to 500VDC |
500 kohm |
20 Tohm |
7 |
| 500VDC to 1090VDC |
5 Mohm |
200 Tohm |
8 |
|
|
|