RIPPLE NOISE METER |
|
|
RM-103 Ripple Noise Meter |
|
RM-103
AC
Ripple Noise Meter
RM-103
Ripple Noise Meter which can measure p-p Ripple Noise, p-p Noise
of Switching Power Supply by digital Measurement. (up to 100MHz).
Ripple Noise or Noise
Measurement has been though as trouble making, because Oscilloscope
won't detect peak of Ripple Noise or Measured data may be changed if
an Operator read such data with an Oscilloscope. Actually some noise
measurement tools are available but these never covers faster Ripple
and Noise more than 50MHz. Using narrow band tool makes no sense for
high speed noise which can gave undesired effect for equipments.
|
 |
-
AC
Ripple Noise and Switching Ripple
can be measured independently.
-
Human
reading errors can be completely eliminated.
-
Only
one Ripple Noise Meter which
conforms to the Standard of EIAJ ,
Electronic Industries Association of Japan
-
Ripple Noise Meter can measure faster
noise up to 100MHz
-
Automatic
Ripple Ratio* Selection Function
*:Ripple Ratio is Parameter to
measure
p-p Ripple Noise accurately
|
|
|
|
|
New Automated Measurement
method
Now you
can terminate such trouble completely. RM-103 can measure p-p
Ripple Noise and p-p Noise accurately and automatically by GPIB and
manually.
With SC-82 Scanner, Multiple Input up to 16
channel can be applied .
You can select Input channel of SC-82
via RM-103 by GPIB for p-p Ripple Noise and p-p Noise
measurement.
Also you can add DP-100 Differential Probe to eliminate
Common Mode Noise.
Of course all these family covers wide band range up to 100MHz
|
Specifications
| DC Volt.
Measurement |
| Range |
+-6.0000V |
+-60.000V |
+-500.00V |
| Resolution |
0.1mV |
1.0mV |
10.0mV |
| Accuracy(*5) |
+-0.025% of rdg.
+-0.025% of f.s. |
| Max. Applied Volt. |
+-500V |
| Measurement Time |
less than 90ms(Fast
Mode)/less than250ms(Slow Mode) |
| Ripple Noise
Measurement |
| Ranges |
300.0mVp-p |
3000mVp-p |
| Resolution |
0.1mV |
1.0mV |
| Accuracy(*1, *2, *3 ) |
+-2% of rdg. +-1% of
f.s. |
| Nominal Bandwidth |
100MHz |
| LF Bandwidth |
50Hz - 2KHz |
| HF Bandwidth |
2KHz- |
| THRU Bandwidth |
50Hz-100MHz |
| Band Pass Filter |
20MHz |
| Ripple Ratio (*3) |
0.0% - 50% (Resolution
0.5%) |
Measurement Time
(*3, *4 ) |
approx.170ms (TYP) |
| Allowed Input |
4Vp-p |
|
Input
Condition |
| Impedance |
DC:1MOhm,High
Frequency :50Ohm |
| Input Cable |
50Ohm
Signal Cable RG-58/Differential Probe DP-100 |
|
Interface |
| GP-IB |
Conform
to IEEE488-1987 |
| PASS/FAIL Judgment |
Photo
Coupler Output same Common (14 pin Connector) |
| Start Trigger |
Photo
Coupler Input same Common (14 pin Connector) |
| 4CH Output |
Photo
Coupler Output forSC-81 Control |
| Input Connector |
BNC
Connector |
| Other Function |
Automatic
Ripple Noise Ratio Selection Function |
|
General |
| Input Voltage |
AC90-264V(50Hz/60Hz) |
| Consumption Power
Weight |
20VA
approx.1.7kg |
| Dimensions |
180(W)x85(H)x300(D) |
|
*1
Effective when Ripple Ratio is setup between 0% and 10%.
*2 Effective when Frequency Range is between 10kHz and 10MHz.
*3 Operates when Frequency Range is between 10kHz and 100MHz.
*4 Measurement Time in the same Measurement Range.
*5 Guaranteed for 6 months when Operational Temp. is 23+-5degree C