BENCH CALIBRATOR LAB STANDARD
M-2000 Bench Calibrator Laboratory Standard

M2000 Bench Calibrator Laboratory Standard.
The M2000 has user friendly control functions with direct keyboard and scroll capability for data entry. 
It can also be operated through an RS232, or optional IEEE interface for automated testing. 
The M2000 is compatible and works with Fluke Met/Cal automated calibration software.
The M2000 was designed to calibrate and test a very wide range of electrical devices


  • Handheld and bench multimeters

  • Analog voltage and current instruments

  • Data loggers

  • Strip chart recorders

  • Other related instruments

  • Engineers involved in product development where sensors, A/D converters, data acquisition and process control cir­cuitry are being developed.

  • Production testing of devices that need analog signals as a stimulus.

  • Process control calibrator reference instrument standards (used in plant I.E. shops).

For price and info please email sales@hotektech.com or call 253-927-6186 ext. 1

Related Instruments Calibrators
Digital Multimeters
Analog Multimeters
Clamp-On Meters
Ohmmeters
Bridges Wheatstone, Kelvin

The M2000 has the ability to store up to 9 set points for each output range. The set points can be selected manually or automatically stepped at timed intervals.

The M2000 contains a standby/operated safety feature typically found in calibrators selling for considerably more than the M2000. This feature protects both the operator and the calibrator from potential overload conditions.

For superior calibration accuracy the M2000 features Tellurium copper binding posts reducing thermal emf's to less than 1 microvolt.

Specifications Specifications (1 year at 23°C ±5°C unless otherwise noted)
 

Output Voltage
Range 0 to 100.000 mV
0 to 1.00000 V
0 to 10.0000 V
0 to 100.000 V
Resolution
0 to 100 mV Range 1microV
0 to 1 V Range 10 microV
0 to 10 V Range 100 microV
0 to 100 V Range 1 mV
Accuracy (% of reading)
0 to 100 mV Range ±0.003% (30ppm) ± 3.0 microV
0 to 1 V Range ±0.003% (30ppm) ± 20.0 microV
0 to 10 V Range ±0.003% (30ppm) ± 200.0 microV
0 to 100 V Range ±0.003% (30ppm) ± 2.0 mV
Maximum Burden (≤ 1 Ohm output impedance)
0 to 100 mV Range 10 mA
0 to 1 V Range 10 mA
0 to 10 V Range 10 mA
0 to 100 V Range 1 mA

 

Output Current
Range 0 to 100.000 mA
Resolution 1 microA
Accuracy (% of reading) ±0.01% ± 2 microA
Maximum Burden 10 V
Stability
Warm-up Time 30 minutes to rated accuracy
Temp Co. (≤18°C/>28°C) 10% of accuracy spec/°C

Temperature Range
Operating 0°C to +50°C
Storage -20°C to +70°C

Power Requirements
Voltage range 90 to 240 VAC <15 VA

Mechanical
Dimensions 11.5" x 4.7" x 8.75" (29.21 cm x 11.93 cm x 22.00cm)
Weight Approximately 5 pounds (2.27 kg)
Display 16 x 2 large character alphanumeric, backlit LCD

General Specifications
Power Requirements: 100/120/240 VAC
Remote Interface RS-232 and GPIB (488 optional)
Circuit Protection Self resetting along with conventional fuses
Warm-up Time 30 min. to best accuracy
Temperature Range Operating 0°C to 40°C
Storage -20°C to 60°C
Stability For temps less than -18°C or greater than 28°C, 10% of accuracy spec/°C
Display 16 x 2 large character alphanumeric with LED backlight
Size Approx. 11.5" x 4.7" x 12."
Weight Approx. 5 pounds
 