All ProductsAnalyticalAnalyzersCalibratorsHumidityData LoggersLight MetersMeasurement

 

MultimetersOhmmetersPressureStandardsTemperatureTransmitters

DOCUMENTING MULTIFUNCTION CALIBRATOR HART
DMC-1400 Documenting Calibrator Laboratory Standard

DMC-1400 Documenting Multifunction Calibrator
The DMC-1400 Calibrator fully documents any calibration easily while performing the normal job. No special software is required! It features the ability to test for pass/fail criteria, measure and simulate thermocouples, RTDs, current, voltage, and frequency, and source pulse trains.
In addition, the built-in 250 Ohm resistor for HART compatibility with smart transmitters and PLCs, full fuseless protection, and a serial communications port for full control with ASCII commands


 
  • Maintain calibration data for 50 instruments
    in non-volatile memory

  • Print calibration certificates without PC

  • Communications port compatible with
    BetaPort-P pressure modules included

  • All source modes can be programmed with dedicated setpoints

  • Built-in 250 Ohm resistor for HART compatibility
    with smart transmitters and PLCs

  • Direct entry of custom RTD coefficients(R0,A,B,C)

  • Meets CE Requirements and is designed to
    IEC 1010 safety standards

For price and info please email sales@hotektech.com or call 253-927-6186

Related Instruments Calibrators
Digital Multimeters
Analog Multimeters
Clamp-On Meters
Ohmmeters
Bridges Wheatstone, Kelvin

Capture calibration data for 50 instruments with both AS FOUND & AS LEFT data
Measure and source T/Cs (13 types), RTDs (13 types), Ohms, current, voltage,
frequency; source pulse trains
IsolatedmA/V read-back circuit for complete transmitter calibration.
Pressure module communication port compatible w ith BetaPort-P digital pressure
modules
using optional BPPA-100 adapter.
Built- in 24 VDC loop power supply can drive 4 – 20 mA loops up to 1000 Ohms.
Direct entry of custom RTD coefficients (R0, A, B, C).
All source modes can be programmed with dedicated setpoints.
Highest accuracy
in class to ±0.015% of reading.
Meets CE requirements and is designed to IEC 1010 safety standards.

DMC-1400 Specifications (23 °C ±5 °C unless otherwise noted
 

Voltage Read and Source
Source 0.000 to 20.000 VDC
Read
Isolated 0.000 to 30.000 VDC
Non-Isolated 0.000 to 20.000 VDC

Thermocouple mV
Read/Source -10.000 to +100.000 mV

Current (mA)
Source 0.000 to 24.000 mA
Read
Isolated 0.000 to 24.000 mA
Non-Isolated 0.000 to 24.000 mA

Frequency (1 to 20 V selectable
amplitude)
CPM Source and Read 2.0 to 600.0 CPM
Hz Source and Read 1.0 to 1000.0 Hz
kHz Source and Read 1.0 to 10.00 kHz

Pulse (Source only; 1 to 20 V
selectable amplitude)
Pulses 1 to 30,000.0
2 CPM to 10 kHz

Ohms
Source 5.0 to 4000 Ohms
Read 0.00 to 4000.0 Ohms

Thermocouple Read and Source
J Thermocouple -200.0 to +1200.0 °C
K Thermocouple -200.0 to +1370.0 °C
T Thermocouple -200.0 to +400.0 °C
E Thermocouple -200.0 to +950.0 °C
R Thermocouple -20.0 to +1750.0 °C
S Thermocouple -20.0 to +1750.0 °C
B Thermocouple +600.0 to +1800.0 °C
C Thermocouple 0 to +2316.0 °C
XK Thermocouple -200.0 to +800.0 °C
BP Thermocouple 0 to +2500.0 °C
L Thermocouple -200.0 to +900.0 °C
U Thermocouple -200.0 to +400.0 °C
N Thermocouple -200.0 to +1300.0 °C

RTD Read and Source
Ni120 (672) -80.0 to +260.0 °C
Pt100 (385) -200.0 to +800.0 °C
Pt100 (3926) -200.0 to +630.0 °C
Pt100 (3916) -200.0 to +630.0 °C
Pt200 (385) -200.0 to +630.0 °C
Pt500 (385) -200.0 to +630.0 °C
Pt1000 (385) -200.0 to +630.0 °C
Cu10 -100.0 to +260.0 °C
YSI400 +15.00 to +50.00 °C
Cu50 -180.0 to +200.0 °C
Cu100 -180.0 to +200.0 °C
Pt385-10 -200.0 to +800.0 °C
Pt385-50 -200.0 to +800.0 °C
Environmental
Operating Temperature -10 °C to +50 °C
Storage Temperature -20 °C to +70 °C
Stability ±0.005% of reading/°C outside of 23
°C ± 5°C
 

Power Requirements
6 VDC; Batteries 4 AA; alkaline or optional rechargeable

Mechanical
Dimensions 8.7” H x 4.2”W x 2.3” D
(220.9 x 106.6 x 58.4 mm)
Weight 30.5 ounces (863 gms)

Accessories Included
Test leads, 4 AA alkaline batteries, NIST Certificate, instruction
manual

Optional Accessories
Carrying Case, Model CC572

Accuracy (read and source unless otherwise noted)
Voltage ±0.015% of reading, ±2 mV

Thermocouple mV ±0.02% of reading, ±10 μV

Thermocouple RTD
(add 0.2°C for Cold Junction Compensation error)

Type J 0.2 °C Ni120 (672) 0.2 °C
Type K 0.3 °C Pt100 (385) 0.2 °C
Type T 0.2 °C Pt100 (3926) 0.2 °C
Type E 0.2 °C Pt100 (3916) 0.2 °C
Type R 1.2 °C Pt200 (385) 0.8 °C
Type S 1.2 °C Pt500 (385) 0.4 °C
Type B 1.2 °C Pt1000 (385) 0.2 °C
Type C 0.6 °C Cu10 1.4 °C
Type XK 0.2 °C YSI400 0.1 °C
Type BP 0.9 °C Cu50 0.4 °C
Type L 0.2 °C Cu100 0.3 °C
Type U 0.25 °C Pt385-10 1.4 °C
Type N 0.4 °C Pt385-50 0.4 °C

Current (mA)
READ                                 SOURCE
±0.015% of rdg ±2 μA         ±0.015% of rdg ±2 μA
CPM
READ                                 SOURCE
 ±0.05% of rdg±1 LSD         ±0.05% of rdg
Hz
READ                                 SOURCE
 ±0.05% of rdg ±1 LSD        ±0.05% of rdg
kHz
READ                                 SOURCE
 ±0.05% of rdg ±1 LSD         ±0.125% of rdg
400ohmRange
READ                                 SOURCE
 ±0.025% of rdg ±0.05ohm    ±0.025% of rdg ±0.05ohm
4000ohmRange
READ                                  SOURCE
±0.025%of rdg ±0.5ohm       ±0.025% of rdg ±0.5ohm
Ohms works with all pulsed transmitters down to 5 ms

Measuring Pressure with the DMC-1400
The DMC-1400 is compatible with BetaPort-P digital pressure modules using the BPPA100 adapter.
Please refer to the separate
Pressure Modules data sheet for module ranges and model
numbers DMC-1400 Preliminary Data Sheet 060803

Information and specifications are subject to change without notice.

 

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or