MC1000
Process Calibrator
MC1000
Multifunction Process Calibrator provides the combination of a high accuracy
calibrator and common Multimeter DMM functions. This Process
Calibrator will
calibrate virtually any process device — read and source
thermocouples, RTDs, Ohms, Voltage, current, frequency, pressure and
test continuity as well.
The MC1000 Process Calibrator is compatible with BETA pressure
modules with the addition of a simple connector/adapter.
The MC1000 Calibrator features a simple user interface, large display, and an RS232
interface.
|
|
Pressure
Ranges |
Various
ranges in 29 different pressure modules – see separate Pressure
Module data sheet |
| T/C
Ranges |
J,
K, T, E, R, S, N, B, L, U including – 10 to 75 mV |
| RTD Ranges |
Pt
385 (100, 200, 500, 1000 Ohms) Pt 392, JIS, Ni120, Cu10, YSI400 |
| Ohms
Ranges |
0
to 400.00 and 400 to 3,200 |
| mA Range |
0
to 24.000 mA |
| Voltage |
(Low)
Read 0-20.000 VDC
Source 0-20.000 VDC
(High) Read only 0-250 VDC or VAC |
| Frequency |
|
Source |
Measure |
| CPM |
2.0 -
1,000.0 |
0.0 -
1,000.0 |
| Hz |
0 -
1,000 |
0.0
- 1,000.0 |
| kHz |
1.0
-10.0 |
0.00
-10.00 |
| Continuity |
0
to 400.0 Ohms, audible tone when under 10 Ohms |
| Accuracy |
T/C
Type J: ±0.3
°C
±10
mV
(±0.4
°C
total error)
RTD Pt 100:
±0.1
°C
±.075
Ohm (±0.3
°C
total error)
Ohms (400):
±0.1Ohm
Ohms (3,200): ±1.0
Ohm |
| mA
Compliance |
Up
to 1,000 load |
| Voltage
Drive Capability |
1
mA |
| RTD IEX-Range |
0.01
mA to 3 mA |
| RTD
Frequency Response |
10
ms – works with all pulsed transmitters |
| Environmental |
Operating
Temperature -10 °C
to 50 °C
Storage Temperature -20 °C
to 70 °C |
| Power
Requirements |
6
VDC Batteries 4 AA alkaline batteries, 25 hours typical operating
time |
| Mechanical |
Dimensions
8.7" H x 4.2" W x 2.3" D (220.9 x 106.6 x 58.4 mm)
Weight 30.5 ounces (863 g) |
| Accessories
Included |
Test
leads, 4 AA alkaline batteries,
NIST Certificate, and instruction manual |
| Optional
Accessories |
Carrying
Case, Model CC572 |
|