MC-1200 MULTIFUNCTION CALIBRATOR
|
|
MC1200 Multifunction
Calibrator
MC1200 Multifunction
Calibrator provides a feature set unmatched in high accuracy, hand-held
calibrators in its price range. The MC 1200 Multi-Function Calibrator provides the
functions and accuracy associated with fixed installation, laboratory
instruments, and has everything needed for virtually any calibration
task.
Model MC1200 Multi-Function Calibrator Measure and source thermocouples, RTDs, current, voltage, and
frequency, and source pulse trains.
|

Multifunction
Calibrator |
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
-
Measure
and source T/Cs (13 types), RTDs (13 types), Ohms, current, voltage,
frequency; source pulse
trains
-
Isolated
mA/V read-back circuit for complete transmitter calibration
-
Pressure
module communication port compatible with Fluke 700 Series, Mensor,
and Martel pressure modules
-
Built-in
24 V supply can drive 4 to 20 mA loops up to 1000 Ohms
-
Direct
entry of custom RTD coefficients (RO, A, B, C)
|
-
All
source modes can be programmed with dedicated setpoints to speed
calibration and
linearity tests
-
Highest
accuracy in class to 0.015% of reading
-
Meets CE
requirements and is designed to IEC 1010 safety standards
-
Fluke,
Mensor, and Martel Pressure Modules
|
A communications port compatible with pressure modules is provided,
as is an isolated mA/V read-back circuit. Arrow keys, direct numeric
keypad entry, and three software-driven function buttons, plus a large
backlit, menu-driven graphics display combine to provide a highly
intuitive, simple yet powerful operator interface.
Built-in 250 Ohm resistor for Hart™ compatibility, compatibility with
smart transmitters and PLCs,
full fuseless protection, and a serial
communications port for full control with ASCII commands, are just some
of the additional features that make the MC 1200 Multifunction Calibrator the single, most
indispensable tool available for virtually any calibration task.
The MC1200 Multifunction Calibrator is supplied in a tough, rubber boot; a carrying case is also
available as an option.
|
|
MC1200 Multi-Function Calibrator
Specifications
(23 °C
±5
°C
unless otherwise noted) |
|
| Voltage
Read and Source |
| Source |
0.000
to 20.000 VDC |
| Read |
|
|
Isolated |
0.000
to 30.000 VDC |
|
Non-isolated |
0.000
to 20.000 VDC |
| Thermocouple
mV |
| Read
and Source |
-10.000
to +75.000 mV |
| Current
(mA) |
| Source |
0.000
to 24.000 mA |
|
Read |
| Isolated |
0.000
to 24.000 mA |
| Non-isolated |
0.000
to 24.000 mA |
| Frequency
(1 to 20 V selectable amplitude) |
| CPM
Source and Read |
2.0
to 600.0 CPM |
| Hz
Source and Read |
1.0
to 1000.0 Hz |
| kHz
Source and Read |
1.0
to 10.00 kHz |
| Pulse
(Source only; 1 to 20 V selectable
amplitude) |
| Pulses |
1
to 30,000.0
2 CPM to 10 kHz |
| Ohms |
| Source |
5.0
to 4000 Ohms |
| Read |
0.00
to 4000.0 Ohms |
| Thermocouple
Read and Source |
| J
Thermocouple |
-200.0
to +1200.0 °C |
| K
Thermocouple |
-200.0
to +1370.0 °C |
| T
Thermocouple |
-200.0
to +400.0 °C |
| E
Thermocouple |
-200.0
to +950.0 °C |
| R
Thermocouple |
-20.0
to +1750.0 °C |
| S
Thermocouple |
-20.0
to +1750.0 °C |
| B
Thermocouple |
+600.0
to +1800.0 °C |
| C
Thermocouple |
0
to +2316.0 °C |
| XK
Thermocouple |
-200.0
to +800.0 °C |
| BP
Thermocouple |
0
to +2500.0 °C |
| L
Thermocouple |
-200.0
to +900.0 °C |
| U
Thermocouple |
-200.0
to +400.0 °C |
| N
Thermocouple |
-200.0
to +1300.0 °C |
| RTD
Read and Source |
| Ni120
(672) |
-80.0
to +260.0 °C |
| Pt100
(385) |
-200.0
to +800.0 °C |
| Pt100
(3926) |
-200.0
to +630.0 °C |
| Pt100
(3916) |
-200.0
to +630.0 °C |
| Pt200
(385) |
-200.0
to +630.0 °C |
| Pt500
(385) |
-200.0
to +630.0 °C |
| Pt1000
(385) |
-200.0
to +630.0 °C |
| Cu10 |
-100.0
to +260.0 °C |
| YSI400 |
+15.00
to +50.00 °C |
| Cu50 |
-180
to +200 °C |
| Cu100 |
-180
to +200 °C |
| Pt385-10 |
-200
to +800.0 °C |
| Pt385-50 |
-200
to +800.0 °C |
|
| Environmental |
| Operating
Temperature |
-10
°C to +50 °C |
| Storage
Temperature |
-20
°C to +70 °C |
| Stability |
±0.005%
of reading/ °C
outside of 23 °C
±5
°C |
| Power
Requirements |
6
VDC |
| Batteries |
4
AA; alkaline or optional rechargeable |
| Mechanical |
| Dimensions |
8.7"
H x 4.2" W x 2.3" D (220.9 x 106.6 x 58.4 mm) |
| Weight |
30.5
ounces (863 gms) |
| Accessories
Included |
Test
leads, 4 AA alkaline batteries, NIST Certificate, instruction
manual |
| Optional
Accessories |
Carrying
Case, Model CC572 |
| Accuracy |
| Voltage |
±0.015%
of reading, ±2
mV |
| Thermocouple
mV |
±0.02%
of reading, ±10
µV |
Thermocouple
Errors
(in °C;
add 0.2 for Cold Junction
Compensation error) |
Type
J
Type K
Type T
Type E
Type R
Type S
Type B
Type C
Type XK
Type BP
Type L
Type U
Type N |
0.2 °C
0.3 °C
0.2 °C
0.2 °C
1.2 °C
1.2 °C
1.2 °C
0.6 °C
0.2 °C
0.9 °C
0.2 °C
0.25 °C
0.4 °C |
| RTD Read and
Source |
Ni120
(672)
Pt100 (385)
Pt100 (3926)
Pt100 (3916)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Cu10
YSI400
Cu50
Cu100
Pt385-10
Pt385-50 |
0.2 °C
0.2 °C
0.2 °C
0.2 °C
0.8 °C
0.4 °C
0.2 °C
1.4 °C
0.1 °C
0.4 °C
0.3 °C
1.4 °C
0.4 °C |
| Current
(mA) |
| Read |
±0.015%
of reading, ±2
µA |
| Source |
±0.015%
of reading, ±}2
µA |
| CPM |
| Read |
±0.05%
of reading, ±}1
LSD |
| Source |
±0.05%
of reading |
| Hz |
| Read |
±0.05%
of reading, ±1
LSD |
| Source |
±0.05%
of reading |
| kHz |
| Read |
±0.05%
of reading, ±1
LSD |
| Source |
±0.125%
of reading |
| Ohms
works with all pulsed transmitters down to 5 ms |
| 400
Ohm Range |
| Read |
±0.025%
of reading, ±0.05Ω. |
| Source |
±0.025%
of reading, ±0.05Ω |
| 4000
Ohm Range |
| Read |
±0.025%
of reading, ±0.5Ω |
| Source |
±0.025%
of reading, ±0.5Ω |
|
|