The Pneumatic Deadweight Tester, DWT can be leveled by means of four adjustable feet and a spirit
level.
The set of non-magnetic weights is conveniently stored on the case
and the tester comes complete
with a set of "quick-fit" adaptors and seals for easy mounting of
the items under test to the
Deadweight Tester.
|
|
Available Models |
|
|
|
S3000P Single Pressure
Models |
|
|
|
Air Operated - Single PCU |
Air Operated - Single PCU |
| Model# |
Pressure Range |
Model# |
Pressure Range |
| S3012-4-P |
1.5 to 100 kPa |
S3015-1-P* |
0.2 to 35 bar |
| S3012-6-P |
15 to 1,000 mbar |
S3015-2-P* |
0.2 to 35 kgf/cm2 |
| S3012-7-P |
5 to 400 inH2O |
S3015-3-P* |
3 to 500 psi |
| |
|
S3015-4-P* |
20 to 3,500 kPA |
| S3013-4-P |
3 to 200 kPa |
S3015-5-P* |
0.02 to 3.5 MPa |
| S3013-6-P |
30 to 2,000 mbar |
|
|
| S3013-7-P |
12 to 800 inH2O |
S3016-1-P |
1 to 70 bar |
| |
|
S3016-2-P |
1 to 70 kgf/cm2 |
| S3014-1-P |
0.2 to 10 bar |
S3016-3-P |
10 to 1000 psi |
| S3014-2-P |
0.2 to 10 kgf/cm2 |
S3016-4-P |
100 to 7000 kPa |
| S3014-3-P |
3 to 150 psi |
S3016-5-P |
0.1 to 7 MPa |
| S3014-4-P |
20 to 1,000 kPA |
|
|
| S3014-5-P |
0.02 to 1 MPa |
|
|
| * -P indicates
built-in pump optional |
|
|
|
|
S3000 Vacuum and Dual
Vacuum / Pressure Models |
|
|
| Vacuum Operated -
Single PCU |
| Model# |
Vacuum Range |
| S3011-4-P* |
3 to 100 kPa |
| S3011-6-P* |
30 to 1,000 mbar |
| S3011-8-P* |
1 to 30 inHg |
| S3011-9-P* |
30 to 760 mmHg |
| * -P indicates
built-in pump optional |
|
|
|
| Air Operated -
Dual PCU |
| Model# |
Pressure Range |
Vacuum Range |
| S3022-4-P* |
1.5 to 100 kPa |
3 to 100 kPa |
| S3022-6-P* |
15 to 1,000 mbar |
30 to 1,000 mbar |
| S3022-7-P* |
5 to 400 inH2O |
1 to 30 inHg |
| |
|
|
| S3023-4-P* |
3 to 200 kPa |
3 to 100 kPa |
| S3023-6-P* |
30 to 2,000 mbar |
30 to 1,000 mbar |
| S3023-7-P* |
12 to 800 inH2O |
1 to 30 inHg |
| |
|
|
| S3025-1-P* |
0.2 to 35 bar |
30 to 1,000 mbar |
| S3025-2-P* |
0.2 to 35 kgf/cm2 |
30 to 760 mmHg |
| S3025-3-P* |
3 to 500 psi |
1 to 30 inHg |
| S3025-4-P* |
20 to 3,500 kPa |
3 to 100 kPa |
| S3025-5-P* |
0.02 to 3.5 MPa |
3 to 100 kPa |
| * -P indicates
built-in pump optional |
|
|
|
|
S3000 Liquid Lubricated
Gas Operated Models |
|
|
| Model# |
Pressure Range |
| S3031-1 |
1 to 70 bar |
| S3031-2 |
1 to 70 kgf/cm2 |
| S3031-3 |
10 to 1,000 psi |
| S3031-4 |
100 to 7,000 kPa |
| S3031-5 |
0,1 to 7 MPa |
| |
|
| S3032-1 |
4 to 140 bar |
| S3032-2 |
4 to 140 kgf/cm2 |
| S3032-3 |
40 to 2,000 psi |
| S3032-4 |
400 to 14,000 kPa |
| S3032-5 |
0.4 to 14 MPa |
|
|
|
|
Specifications
|
|
|
Accuracy*
±0.015% of reading (±0.008% optional)
*Accuracy based on % of reading from 10% to 100%
of the piston range when used in accordance with the corrections
found on the calibration certificate. Below 10% ± (accuracy class) x
10% of the piston range.
|
| Mass and PCU
Materials of Construction |
| Standard weight material |
Series 3 non-magnetic austenitic
stainless steel |
| Weight density |
7.8 g/cm3 |
| Optional fractional weights |
Solution heat treated aluminum |
| Weight density |
2.7 g/cm3 |
| |
| Model |
Piston Material |
Cylinder Material |
Coefficient of
Expansion |
S3011, S3012, S3013
S3022, S3023, S3025(V) |
Ceramic |
Martensitic steel |
11 ppm/degC |
S3014, S3015, S3025 (P)
S3031, S3032 (P & V) |
Tungsten carbide |
Martensitic steel |
16.5 ppm/degC |
| S3016 |
Tungsten carbide |
Tungsten carbide |
11 ppm/degC |
|
| |
| General |
| Test Port Adaptors |
1/8, 1/4, 3/8 and 1/2 NPT or BSP |
| Instrument weight |
24 lbs |
| Instrument size |
17.5W x 12 D x 8.5 H |
| Mass set weight (typical) |
65 lbs |
| Weight Increments |
| Minimum Standard
Weight Increments |
S3012, S3022 Pressure
S3013, S3023 Pressure
S3014, S3015, S3025 Pressure
S3016, S3031, S3032
S3011, S3022, S3023
S3025 Vacuum |
1 inH2O, (5 mbar)
2 inH2O, (10 mbar)
1 psi, (0.1 bar)
1 psi, (0.1 bar)
0.2 inHg, (10 mbar)
0.2 inHg, (10 mbar) |
| Optional Fractional
Weights |
| S3014, S3015, S3025 Pressure |
0.1 psi (0.01 bar) |
|
|
|
Built-in Hand Pumps
Pressure mode: max pressure 300 psi (20 bar)
Vacuum mode: to 90% vacuum
Operating Fluid for Liquid Lubricated Models
55-655 mineral oil (standard)
Krytox, for oxygen-safe applications (optional)
|
|
|