PC6-PRO PRESSURE CALIBRATOR |
|
PC6-Pro Pressure Calibrator
The
PC6-Pro Pressure Calibrator combines
the latest developments in pressure instrumentation with advanced sensor
and component technology to offer a versatile calibration instrument.
The PC6 can simultaneously display pressure versus voltage,
current, date and time, percentage of pressure, pressure switch status,
temperature, and maximum or minimum values. Data can be logged and
stored with time and date stamping for future reference. Windows™-based
SICAL software can be linked to the PC6 pressure calibrator for remote
operation and data transfer.
|
 |
-
Pressure measurement range -1 to 1000bar / 15,000psi
-
Voltage and current measurement and sourcing
-
Pressure to Current or Voltage conversion
-
Current or Voltage to Pressure conversion
-
Two wire Transmitter Simulation
-
Simultaneous display of pressure and selected function
-
Timed / Event data storage with date & time stamping
-
Up and download Cal procedures via RS232 Interface
-
Alternative languages available e.g. French, German,
Italian, Spanish,
etc.
-
Pressure switch test
-
Leak rate measurement 1 second to 999 days
-
Maximum and minimum pressure values
-
Display hold facility
-
Supply output for transducer / transmitter
|
|
|
|
|
It
is designed for accurately testing and calibrating the following types of
instruments:
-
Pressure
Transducers / Transmitters
-
Pressure
Switches / Relief Valves
-
Pressure
Indicators (analogue / digital)
-
Process
Indicators
-
Current /
Voltage to Pressure Converters
|
The compact
size of the PC6-Pro does not compromise functionality. It can simultaneously
show pressure versus voltage, current, date & time, percentage of
pressure, pressure switch status, maximum or minimum pressure values .
Data can be logged and stored with time and date stamping for future
reference. PC6 can source current and voltage using various options e.g.
continuous, step or square root functions.
|
The
PC6-Pro is supplied with our hand held test systems:
-
LTP1 Low
Pressure pneumatic test system -1 to 2 bar (30 psi)
-
TP1
Pneumatic pressure test system -1 to 25 bar (360 psi)
-
HTP1
Hydraulic pressure test system to 1000 bar (15,000 psi)
|
Pressure units
bar, mbar, Mpa, kPa,
lb/in², kg/cm², atm, inH2O, mH2O, mmH2O, inHg, mmHg, unit1, unit2 (user
defined) ²² ²
Pressure Display
Resolution 6 Digits
Overload Pressure 10%
of FS
Temperature Display Temperature
of pressure sensor module ( Accuracy: ±0.5°C )
Environmental
Conditions Non-dusty,
non-corrosive
Operating
Temperature 0 to
40ºC
Calibration
Temperature 20ºC
± 2°C
Calibrated Temp.
Range 10 to 40°C
Storage Temperature -20
to 70ºC
Temperature Effects 0.001%
of reading / °C between 10 to 40°C
Humidity 5
to 95% relative humidity non-condensing
Pressure Switch Test
Status OPEN /
CLOSED
RS232 Parameters Baud rate 110, 300, 1200, 2400, 4800, 9600, stop
bits 1 or 2, status ON or OFF
Zero Reset Manual - keyboard operation
Data Capacity 719 records into a maximum of 20 files, 32k memory
(additional 744 records,
32k memory option available)
Display LCD, 16 characters x 2 line alpha-numeric
Pressure Media Water, Oil, Non-corrosive gases ( for other media
Contact
Us )
Resolution Select Increase or decrease by a factor of 10
Power Supply 7.2V NIMH rechargeable battery pack / mains operation
via charger (supplied)
( To be charged in Non-Hazardous areas only )
Battery Life 10 Hours fully charged
Low Battery Continuous check, audio and visual warning
Dimensions 92 x 110 x 60mm
Weight approx. 1.2 kg
Electrical Connections Ground Standard 2mm socket
Voltage / Current Input Standard 2mm sockets
Pressure Switch Input Standard 2mm sockets
Battery Charger 2-pole miniature round connector
RS232 8-pole miniature round connector
EPM 14-pole miniature round connector
Pressure Connections 3/8" BSP male 'quick-fit' / 1/4"BSP
or 1/4"NPT male (Pneumatic)
3/8" BSP female 'quick-fit' / 1/4"BSP or 1/4"NPT male
(Hydraulic)
Software (optional) ‘SiCal’ will allow users to control the PC6
Calibrator remotely through a virtual instrument panel on the computer screen. The calibration data can be logged, viewed and
printed in graphical format or on a calibration certificate. ( To be used in Non-Hazardous
areas only )
System Requirements: 486 with math coprocessor. Windows 3.1 or higher.
|
|