All ProductsAnalyticalAnalyzersCalibratorsHumidityData LoggersLight MetersMeasurement

 

MultimetersOhmmetersPressureStandardsTemperatureTransmitters

AG0313 RF PLASMA AMPLIFIER GENERATOR
RF Plasma Generator Amplifier

AG 0313 RF Plasma Amplifier Generator.
AG 0313 RF Plasma Amplifier Generator has 300 Watts RF Power at 13.56 MHz for Industrial and Laboratory Applications.
AGC Power Leveling: Output Power Control to better than ± 3 W of set value.

Burst & Sweeping with custom defined settings.

  • 13.3 MHz to 13.8 MHz up to 
    greater than 350 Watts
  • Low harmonic level at 300W h2£ -40 dBc, 
    h3 and higher < -55 dBc
  • Digital Meter, measures forward, reflected, 
    and load power simultaneously.
  • Front Panel Control of Generator and 
    Amplifier functions.
  • Data acquisition: Status Monitoring & Power Measurement at Analog Port
  • Optional RS232 communication: 
    Full Control Of Generator/Amplifier Functions
SPECIFICATIONS
DESCRIPTION RANGE/CHARACTERISTICS
Class of Operation Class B
Frequency of Operation / Frequency Stability
13.56 MHz center. Adjustable within 13.3 MHz to 13.8 MHz / 0.005%
RF Power Output / RF Power Margin 350 Watts into 50 Ohm / 28%
Gain (as amplifier) 54 dB @ 100W ±1 dB
RF Input Drive
(as amplifier)
Recommended +1 to +3 dBm for the best operation
Input Drive Source (amplifier) Signal or function generator, analog computer input capable of up to 2 Vp-p @ 50 Ohm
Input and Output Impedance
50 Ohm
Input VSWR 1.2:1 max
Output VSWR 3:1 max
Output VSWR Protection 75 W max reflected power limit. Automatic,
limits typically within 0.5 ms after reverse power reaches 75W or power amplifier current preset limit.
Harmonic level @ 300W
Better then - 40 dBc for 2-nd harmonic, any other > -55 dBc
Spurious Output - 76 dBm equivalent noise level generated by internal circuits
Mismatch Power Output

Forward power reference to rating:
350W nominal into 1.1:1 VSWR
350W FWD, 185W Load into 2:1,
180W FWD, 115W Load into 3:1
115W FWD, 43W Load into 9:1

Dynamic Power Range 0 to 350W, settings within +/- 2W
RF Input / Output connectors

Input BNC Female
Output N Female
Rear Panel

Power Source 90 - 240 VAC, 47 - 63 Hz, broad input voltage with no adjustment required.
Power connection IEC Standard Power Entry
Cooling Forced air, temperature controllable, heatsink temperature monitored via RS232 GUI Interface
Dimensions (with hardware) H135mm x W112mm x L445mm ( 5.25” x 8.3” x 17.5” )
Weight 10.5 kg, 23.5 lbs.
Mounting Half Rack, 2U high. Optional: Rack Mount Kit, Adapter Kit, Coupling Screws.
BURST Operation Min burst ON time 1 us, min burst repetition time 1ms. User settings via GUI and RS232 port. Activation from GUI only.
SWEEP Operation 13.3 to 13.8 MHz. Min time 10 ms, max 10s. Settings and activation from GUI only.
RF Output Settings Front Panel EDITOR,
RS 232 port from GUI,
hardware port 10V scale
Output Blanking For pulsed applications, amplifiers and generators offer blanking of the output signal for minimum noise RF spectrum
Operating Temperature 10° to 40° C ambient air