  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

ATN-5 AUTO TUNER NETWORK


ATN-5 ICP Auto Tuner
The ATN-5 ICP Auto Tuner is designed to match the impedance of the load coil while maintaining 50 Ohms of reactive impedance to the RF Generator. During ignition and final tuning the match will vary over
a small impedance range causing ionization of the argon or other gas. 
Upon ionization a more dense plasma will form and the Auto Tuner will adjust to couple the impedance match. Capacitors range until a null is seen in the detector circuit of the auto Tuner.

Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com

The complete unit is separated into two enclosures.
The PT-II-CE is the controller unit and Auto Tuner is the impedance matcher.
Both units are enclosed in aluminum, with the Auto Tuner being yellow irradiated for conductive ground purpose.
The Matching network contains an Air Variable Capacitors for tuning and load.
The unit is considered "L" shape configuration which gives the Auto Tuner direct coupling to Plasma effect. 
This is the most efficient method for power transfer in RF.
The controller in an all self contained power supply and servomotor control unit. 
The front panel has controls for both manual and auto modes. It also provides ignition preset potentiometers for efficient lighting and tuning of the Plasma Process.

SPECIFICATIONS
DESCRIPTION RANGE/CHARACTERISTICS
P&M Detector
Frequency of Operation
13.56 MHz
RF Power Rating Up to 1000 W
Impedance Rating 50 Ohms
Matching Box 9 W x 16 D x 5 H
Input Connector N Type
Cooling
Air
 
PT-II-CE Controller
Electrical Power Required 110-240 VAC
Electrical Power Consumption
45 Watt
AC Power Connector 6DEI EMI Filter
Dimensions

8 W x 9 D x 1.5 H