  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

HIGH RESISTANCE RESISTOR DECADE BOX


4721 High Resistance Resistor Decade. High Voltage Resistor Decade Box.
The Model 4721 high resistance resistor decade box is designed for the checking and calibration of insulation testers at voltages of up to 5000V DC.

The total range covered is from 1 kΩ to 1TeraΩ using specially selected resistors and a high quality switch mounted on an easy to read panel. 
This resistance box is housed in a strong aluminum case with lid and carrying handle to provide an ideal solution to on-site and laboratory calibration of high resistance insulation testers. 

High resistance decade

High Resistance Decade

Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com
Calibration of insulation tester
Range 1kΩ to 1TΩ
Voltage rating up to 5,000V DC
Portable field or laboratory use 

The instrument can be supplied complete with a calibration certificate provided by our own in-house calibration laboratory. 

Switch Settings Max d.c. Volts Accuracy  Temperature Coefficient
1kΩ 10V 0.1% 0.015%/°C
10kΩ 50V 0.1% 0.015%/°C
100kΩ 150V 0.1% 0.015%/°C
1MΩ 300V 0.1% 0.025%/°C
10MΩ 500V 0.1% 0.05%/°C
100MΩ 1000V 1% 0.1%/°C
1GΩ 5000V 1% 0.1%/°C
10GΩ 5000V 1% 0.1%/°C
100GΩ 5000V 1% 2%/°C
1TΩ 5000V 1% 2%/°C
Overall Range 1 kohm to 1 Tohm in 10 steps with open circuit on eleventh position. 
Accuracy Stated at 23°C±1°C 10 to 50%RH 
Earthing  Metal earth terminal on switch panel.
Connections Plug-in shielded plugs with 1 meter shielded cable (2 off High and Low) 
Dimensions approximately  220 x 110 x 90 mm
Net Weight approximately 0.8kgm.