 |
|
HIGH
RESISTANCE FAULT LOCATOR MODEL 5762 |
|
|
|
High Resistance Fault Locator
|
|
|
5762 High
Resistance Fault Locator.
The High Resistance Fault Locator, Type 5762, utilizes a
conventional Wheatstone Bridge circuit in which the two sections of
the faulted conductor, one on each side of the fault, comprise the two external
arms of the bridge.
|

|
Introduction
The other two arms of the bridge are
contained in the instrument.
By use of a detector circuit, of
extremely
high input resistance it is possible to locate
high
resistance faults without loss of sensitivity.
|
|
|
|
|
|
|
With this bridge arrangement,
faults having resistance from zero up to 200 MΩ
in dielectrics such as rubber and polythene, can be located with an
accuracy to well within ±0.5%, a typical error being 6” in 500 feet
(150mm in 150m) or ±0.1%. The limitation is determined by the
uniformity of the conductor.
Accuracy
Well within ±0.5%, a typical error being 6” in 500 feet (150mm
in 150m) or ±0.1%.
Connection Methods
a. WHEN BOTH ENDS OF THE FAULTED CONDUCTOR ARE TOGETHER, as in
cable manufacture, or when the cable is on a cable reel. (Fig I and
Fig.2)
|
|
|
b. WHEN THE ENDS OF THE FAULTED CONDUCTOR ARE NOT AVAILABLE AT THE SAME
POINT, as in the case of a cable in service.
Two methods may be used. The preferred method is to run out two
insulated wires to the distant end to serve as current and potential
leads. The resistance of these leads should be as low as possible, e.g.
less than I Ohm. Alternatively, a good conductor in the cable may be used. Connect the good
conductor and the faulted conductor solidly together at
the distant end and adequately insulate the joint from earth. Make
connections at the near end as shown in Fig.3.
|
|
|
|
This method effectively doubles the length of the faulted conductor and
reduces the accuracy to half
Fault Location
i) With all the connections made as in (2) switch on the fault locator
and allow the instrument to settle for one minute.
ii) By means of the zero control adjust the instrument until the
microammeter reads zero.
iii) Connect the second current lead to the battery.
iv) Adjust the Fault Locator main dial for zero reading on the
microammeter.
The reading of the main dial then indicates the location of the fault
as a percentage of the total length of cable loop from the end connected
to the “+“ terminal on the instrument.
v) Reverse the connections to the conductor terminals and rebalance.
The sum of the readings obtained in (iv) and (v) should add to 100%.
Note:
The total length of the cable loop means the total length of cable
between the potential points, that is the points at which the leads to the
terminals of the instrument are connected to the cable.
This is equal to
the length of the cable in the case shown in Figs I and 2 but is equal to
twice the length of cable in the case shown in Fig. 3.
|