Toll Free: 866-318-6186    Email: sales@hotektech.com

 
 
Instruments, calibrators, standards, analyzers, dataloggers. analyzers
 
  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

HIGH RESISTANCE FAULT LOCATOR


5762N High Resistance Fault Locator.
The 5762N Resistance Fault Locator is a portable cable fault locator for determining the position of a single shorting fault between two conductors, or between one conductor and the cable sheath or ground.
The fault locator can be powered from a mains electricity supply or from the internal battery that is automatically recharged when the instrument is connected to the mains.

 High Resistance Fault locator

High Resistance Fault Locator
Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com

A high input-resistance differential amplifier circuit is used to drive the detector allowing
faults of up to 200MΩ to be accurately located.
A maximum potential of 6V (limited to 1A, or 10A if 10A BOOST button is held) is applied
across the bridge circuit and therefore the cable under test.

The 5762N High Resistance Fault Locator is housed in a rugged highly luminous
case manufactured from a high impact copolymer and featuring an internal metal sub-frame
for added strength. When closed, the case is rated to IP67.

The 5762N High Resistance Fault Locator is equally suitable for use in the factory, to find faults in stock cable, or in the field to locate faults in an installed system. The fault is located by manually adjusting a potentiometer to find the null point as indicated on a front panel meter. The instrument uses a bridge circuit with the two sections of the faulty conductor (either side of the fault) acting as one half of the bridge network and the internal pushbutton potentiometer forming the other. The potentiometer is adjusted until the null is found, at which point the distance to fault can be read from the instrument as a percentage of cable length.

5762N High Resistance Fault Locator Specification

Measurement
Fault location Shorting faults of up to 200MΩ
Resolution 0.1% of length of cable on test
Accuracy 0.5%, subject to cable uniformity
Technical data
Battery Yuasa NP2.8-6, 6V 2.8Ah sealed lead-acidIP67.
Operating temp 0ºC to+40ºC
Storage temp +85ºC
Mains connection 85 – 264V AC, 47 – 63Hz
Power consumption 9W (when fast charging)
Dimensions 270 x 246 x 124 mm
Mass TBC

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.