AUTOMATIC THERMOMETER BRIDGE RESISTANCE THERMOMETRY
Senator Multi-Range Automatic Thermometry Resistance Bridge Type 5840D

5840D Thermometry Resistance Bridge
The SENATOR Thermometry Resistance Bridge is a multi-range Automatic Resistance Thermometer Bridge. Using the well proven inductive division principle with unique circuitry, the SENATOR provides accurate AC measurement of resistance up to 1100 Ohms for high precision temperature applications to 1 millidegree. No initial setting-up is required and calibration may be readily verified using suitable certified resistance standards such as the 5685 range.

3 Ranges
Choice of currents
Direct reading in Ohms
Readout in temperature 
K or °C (via computer)
IEEE-488 interface
1mΩ resolution
Low frequency operation 75Hz
Recorder output
Self balancing
Ratio Mode
5840D  Automatic Thermometer Bridge Technical Data

Resistance - Temperature Readout

Precise true resistance measurement is clearly displayed on the high-stability digital readout with a resolution of 1mOhm on the 10 Ohms range. The reading can also be transferred over the IEEE-488 interface to a computer controller for temperature conversion. The resultant temperature calculation is then sent back for display on the SENATOR front panel in units of K or C. Suitable software is available from Tinsley to convert the AC resistance measurement into temperature and to calculate the constants of 25 Ohms Platinum Thermometers from ITS-90 Resistance-Temperature data. 

3 Resistance Ranges - Applications

The built-in range selection of 10 Ohms, 100 Ohms and 1 kOhm is very desirable and renders the SENATOR ideal for use in a wide range of applications for high precision temperature measurements.

1. The 10 Ohms range is ideal for use with Rhodium Iron Thermometers, Tinsley type 5187U and 5187W and for capsule 25 Ohms Platinum Thermometers, Tinsley type 5187L.

2. The 100 Ohms range is the most suitable for long stem 25 Ohms Platinum Thermometers such as Tinsley type 5187SA.

3. The 1000 Ohms range is selected for 100 Ohms Platinum Thermometers above 0°C.

Specifications

RANGE

CURRENT
ma +/- 1%
ACCURACY
relative to Calibration Standard
Resolution
Selected Measured
10Ω 0 to 16Ω 0.2
0.5
1.0
2.0
5.0
10.0
5ppm+/-  5mΩ
3ppm+/-  3mΩ
2ppm+/-  2mΩ
2ppm+/-  2mΩ
2ppm+/-  2mΩ
2ppm+/-  2mΩ
1 mΩ
100ΩΩ 0 to 160Ω 0.2
0.5
1.0
2.0
2ppm+/-  20mΩ
1ppm+/-  10mΩ
1ppm+/-  10mΩ
1ppm+/-  10mΩ
10 mΩ
1000Ω 0 to 1600Ω 0.2
0.5
1.0
5ppm+/-  100mΩ
5ppm+/-  100mΩ
5ppm+/-  100mΩ
100 mΩ
'Self Heating Check' Current  - 1.414 X Selected current Internal Standards - 10Ω, 100Ω. and 1000Ω with trim adjusters for calibration to a certified standard resistor.
Zero Stability  -  Negligible drift with time and temperature External Standards - The bridge will read ration Rx to 10Ω, 100Ωand 1000Ω standard resistors.
Calibration Drift  - 1ppm / year -  Internal Standards,
0.25ppm / year - External Standards
Quadrature Balance - Fully automatic balance of thermometer reactance with visual indication if overloaded.
Self-Balance Speed - 10 seconds from unknown resistance to 1ppm accuracy. (100Ω range) Power Requirements - 115/230 Volts; 50/60 Hz, 250 VA.
Accuracy - Better than 0.5 millidegree at 0°C (for 25W PRT's) Net weight -  16Kgs; Dimensions  - 480 X 160 X 400mm.