SHORT
HAUL SUBMARINE CABLE TEST SET
|
|
|
5910R Short Haul
Submarine Cable Test Set
The 5910R is a
rack mount version of the 5910 portable Short Haul Submarine Cable
Test Set.
It brings the latest technology to the tried and tested DC
methods of cable fault location.
Common test procedures
which are associated with the location of cable fault such as
“open circuits” and “shunt faults” are pre-programmed into
the instrument. This allows the procedure to be carried out
automatically therefore eliminating operator errors and reducing the
uncertainty of fault position prediction.
|
Submarine Cable Test Set |
| Our
Price: |
Contact us for
pricing |
| Phone:
|
253-927-6186
866-318-6186 |
| Fax:
|
253-927-6189 |
| email:
|
sales@hotektech.com |
|
|
|
|
|
|
|
The
instrument may be set for a range of voltages
and currents making it
suitable for use with
most un-repeatered submarine communication cable systems.
The RS232
interface is designed for full duplex communication with the
instrument.
All functions may be controlled via a computer and obtained data may be
stored for future analysis.
This interface may also be used to enable two
5910’s to communicate
with each other via a
computer when carrying out double ended tests
such
as “no loss of current”.
|
A basic electroding
generator function is included which can assist the cable repair
vessel in locating the cable on the sea bed prior to retrieval and
repair.
|
Specifications
Ohmmeter – Conductor resistance measurement
|
|
Resistance
Range |
Resolution |
Accuracy
(within compliance of 1V to 140V) |
| 0.2Ω
to 100.00Ω |
0.01Ω |
±0.2Ω |
| 100.00Ω
to 1000.00Ω |
0.1Ω |
±0.2Ω
up to 200Ω
±0.1% above 200Ω |
| 1KΩ
to 10kΩ |
1Ω |
±0.1% at 1KΩ
reducing to
±0.2% at 10KΩ |
| 10KΩ
to 100KΩ |
10Ω |
±0.2% at 10KΩ
reducing to
±2% at 100KΩ |
|
|
|
Current Control
|
|
|
| Range |
Resolution |
Accuracy |
| 0.5 to 10.0mA |
0.1mA |
±2µA |
| 10.1mA to 100mA |
0.1mA |
±20µA |
| Polarity |
Positive “+” or
negative “-” selectable by thumbwheel switch |
|
|
|
Megohmmeter
– Insulation resistance measurement
Voltage settings +50V, 100V, 250V, 500V, or 1000V
Polarity selectable positive “+” or negative “-”
|
|
|
Range
|
Applicable
Voltage
|
Accuracy
|
| 10 5Ω
to 109Ω |
50V, 100V |
±1dB (10 5 to 108Ω
) using Standard Resistors |
| 10 6Ω to 1010Ω |
250V |
±1dB (106 to 109Ω
) using Standard Resistors |
| 107Ω to 1011Ω |
500V, 1000V |
±1dB (107 to 1010Ω
) using Standard Resistors |
|
|
|
Capacitance
Direct reading capacitance measurements between 0.05µF
and 2000µF.
|
|
|
Range
|
Resolution
|
Accuracy
(using standard caps)
|
Measurement
Cycle
|
| 0.05 to 2µF |
0.001µF |
±0.005µF |
2 secs @ 1µF |
| 2 to 20µF |
0.001µF |
±0.2% |
2 secs @ 10µF |
| 20 to 200µF |
0.01µF |
±0.2% |
20 secs @ 100µF |
| 200 to 2000µF |
0.1µF |
±0.2% |
200 secs @ 1000µF |
|
|
|
Electroding
Generator (GEN)
Output polarity either positive to LINE or negative to
LINE
|
|
|
| Control |
Constant current – compliance 0.1
to 50V |
| Frequency |
25Hz or 16.7Hz |
| Output |
50mA + 50mA sin wt
(i.e. minimum 0mA, maximum 100mA, synthesized, sinusoidal current |
|
|
|
|