LCR DATA BRIDGES
LCR Data bridge Model 6401. 0.25% nominal accuracy. 100/120Hz - 1kHz

6401 LCR Bridge is an economically priced, automatic and microprocessor controlled LCR bridge with a basic accuracy of 0.25%.


The model 6401 is an economically priced, automatic and microprocessor controlled LCR bridge with a basic accuracy of 0.25%.

 It’s useful features and realistic price make it ideally suitable in many applications.

For price and info please email sales@hotektech.com or call 253-927-6186 ext. 1

Related Instruments LCR Meters
Bridges Wheatstone, Kelvin
Component Testers
Ohmmeters
Insulation Testers
Digital Multimeters
Measures R, L, C and Q automatically.
0.25% of reading, ±1 digit basic accuracy.
Eight decades of LCR measurement.
2 selectable measurement frequencies: 100/120* Hz or 1 kHz (0.025% accuracy)
4-digit LED display with automatic decimal point. 
User prompt facility for best accuracy.
Selectable series or parallel measure
Integral 4-terminal test fixture.
External bias voltage up to 50 V possible.
Input protection against charged capacitors up to 10mF and 50V. 
Model 6403 is equipped with an interface for the limits comparator model 6422. Test limits can be defined either of two ways. In the first of these, the upper and lower acceptable values are set in the two banks of thumbwheel switches. Alternatively the PASS band can be determined using a nominal value and upper and lower percentage tolerance. Three indicators are available to show where the measured value lies with respect to the upper and lower pass band limit. 
The same information is also available at 3 floating relay contacts of the 6422 comparator. 
Automatic indication for overrange or underrange. 
Easy operation by means of only 6 keys. 
Automatic selection of measurement range (lockable if many components in the same range are to be measured) and automatic distinction between L and C. 
Validity of measurement after 1 second maximum. 
2 measurements per second. 
An adaptor for axial lead components is provided with the unit. 
Selectable internal 2 V bias voltage for measurements of electrolytic capacitors. 
 Detailed operation manual. 

 

*120 Hz measurement frequency available as factory-fitted option.

Specifications

Basic Accuracy 0.25% of reading, ±1 digit
R Measurement Range  0.001Ω — 100MΩ
Ultimate Resolution  0.001Ω
Conditions for Basic Accuracy 100 Hz:- 1Ω - 2 MΩ, 1 kHz:- 1Ω - 2MΩ
L Measurement Range  0.1mH - 9900 H
Ultimate Resolution 0.1mH
Conditions for Basic Accuracy 100Hz:- 2mH - 2000H, 1 kHz:- 200mH - 200H
C Measurement Range  0.1pF —9900mF
Ultimate Resolution 0.1pF
Conditions for Basic Accuracy 100 Hz:- 2nF - 2000mF, 1 kHz:- 200pF - 200mF
Q Measurement Range  0 — 99
Ultimate Resolution 0.001
Conditions for Basic Accuracy 0.25 - 4
Maximal Voltage on Component  0.3 V rms
DC-Bias Voltage  2 V internal (selectable) or up to 50 V external
Measurement Modes  Series or parallel equivalent circuit
Measurement Update Rate  2 per second
Temperature Range 0°C - 50°C
Power 95 V to 125 V or 195 V to 255 V, 48—63 Hz
Dimensions 440 x 250 x 42 mm
Weight 5.6 kg
Model 6401 with 6404 Interface and 
6422 Comparator 

Ordering Information

6401  Automatic 0.25% LCR Databridge
6401/6403 Complete with Interface for Comparator 
6401/6403/6422  Complete with Interface and Comparator 
Accessories  Models 6601, 6602, 6603, 6604, 6613 (for use with 6604), 6615 to 6623 


Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.

[Home] [Top] [Distribution] [Contact Us]