TECHNICAL DATA |
Voltage
measuring
ranges
trms values/V
Permissible peak value /V |
| |
1
2 |
3
6 |
10
20 |
30
60 |
100
200 |
300
600 |
1000
2000 |
|
|
|
|
|
|
Overload
strength
Input resistance
Crest factor |
1500V permanently 2000V for 3s 8000V for 1.2/50ms
>1.5MΩ II 50pF
8...2 for Utrms=25...100% of the selected measuring range
20...8 for Utrms=10...25% of the selected measuring range |
Current measuring ranges
trms values /A
Permissible peak /A
Overload strength/A
Overload strength for 3s/A
Input resistance/Ω |
| |
3m
6m
1.5
4
5 |
10m
20m
1.5
4
5 |
30m
60m
1.5
4
5 |
0.1
0.2
1.5
4
0.3 |
0.3
0.6
1.5
4
0.3 |
1
2
1.5
4
0.3 |
3
6
35
45
0.01 |
10
20
35
45
0.01 |
30
60
35
45
0.01 |
|
|
|
|
| Crest factor |
8...2 for Itrms=25...100% of the selected
measuring range
20...8 for Itrms=10...25% of the selected measuring range |
External
shunt voltage ranges
trms values/mV
Permissible peak /mV
Overload strength/V |
| |
15
30
3 |
50
100
3 |
150
300
3 |
|
|
|
|
|
|
Input resistance/Ω
Crest factor |
>1.5MΩ
8...2 for Utrms=25...100% of the selected measuring range
20...8 for Utrms=10...25% of the selected measuring range |
| Range selection |
Auto, manual or
remote-controlled, separately for each current and voltage
channel, input of factors for transformers (U and I), level
control display for each channel in the status line |
| Isolation |
Current and voltage path may
float against each other and against earth up to 1500V |
| Measuring method |
Simultaneous sampling of the
current and voltage inputs and A/D conversion of the
instantaneous values |
Measuring cycle,
synchronization,
averaging |
For the measurement of the
trms values for current, voltage and active power, the
measuring cycle time is adjustable in the range of 100ms to
60s. The synchronization can be performed on the measuring
signal, the fundamental harmonic of the measuring signal,
the mains or an external signal. A single measurement with
automatic stop after 1 measuring cycle is possible.
Averaging over 1 to 16 measuring cycles. |
| Measuring accuracy |
Measuring
Accuracy |
±(% of
measuring value + % measuring range) |
| DC 0.1...15Hz |
15..500Hz |
0.5...50kHz |
50...150kHz |
150...250kHz |
250...400kHz |
400kHz...1MHz |
| Voltage |
0.1+0.05 |
0.05+0.05 |
0.1+0.05 |
0.2+0.1 |
0.3+0.2 |
0.5+0.5 |
1.0+0.1*(f-400kHz)/kHz |
| Current |
0.01+0.05 |
0.05+0.05 |
0.1+0.05 |
0.2+0.1 |
0.3+0.2 |
0.5+0.5 |
1.0+0.1*(f-400kHz)/kHz |
| Active Power |
0.015+0.1 |
0.07+0.08 |
0.15+0.1 |
0.3+0.3 |
0.5+0.5 |
0.7+1.0 |
1.5+0.15*(f-400kHz)/kHz |
|
| Accuracies based
on |
1. sinusoidal voltages and currents
2. ambient temperature 20...25°C
3. warm up time 15 minutes
4. definition of power range as the product of current and
voltage measuring range, 0 ≤λ≤
1
5. calibration interval 12 month |
| Other values |
All other values
are derived from the values for current, voltage and active
power. Accuracies for the derived values depends on the
functional relation (e.g. S = I * U,
ΔS/S =
ΔI/I +
ΔU/U) |
| Frequency
measurement |
0.01Hz...250kHz
±0.01% of measuring value, measuring channel selectable. |
| Measuring wirings |
a) phase current with phase
voltage (star/star)
b) phase current with linked voltage (star/delta)
c) linked current with phase voltage (delta/star)
d) linked current with linked voltage (delta/delta)
e) Aron-wiring (2-watt-meter-method, third channel free)
With the optional star/delta conversion the displaying of
values from the respectively other wiring is possible
(only wiring a) to d)) |
| Display of
measured and computed values |
Representation
Voltage/current
Power
Impedance |
With standard abbreviation
of electrical quantity, phase and dimension, 5-digits
(0...99999), with sign, decimal point and unit prefix after
the digits (e.g. I2trms/A 0.7385m). 1 to 30 values can be
displayed simultaneously, selectable via default or user
defined menus (max. 10)
Trms value, peak values (min, max, pp), rectified value (rect),
mean value (dc), rms value of ac-component, form factor,
crest factor
Active power (P), reactive power (Q), apparent
power (S), phase anglel (r), power factor (l)
Amount (Z), real and imaginary part of parallel or serial
equivalent circuit |
Integrated
values depending on
the measuring time
Energy, Charge
Date and time, measuring time
Adjustable
parameters
Values of
harmonic analysis |
The integration can be controlled manually, automatically
using start and stop times, via external trigger or remote
on controlled via computer interface
Active energy (Ep), reactive energy (Eq), apparent energy
(Es), charge (q)
Current date (day, month, year) with time (hours, minutes,
seconds), accu-buffered real time clock,
start time for measurement, running measuring time, on-time,
all measuring times with day, hours, minutes, seconds
Scaling factors for external shunts, current and voltage
transformers, basic load (stand-by power)
Amount and phase of harmonics for current, voltage and
power, angle between phase voltages and phase currents,
phasor (Fresnel) diagram, THD (Total Harmonic Distortion),
THF (Telephone Harmonic Factor), HVF (Harmonic Voltage
Factor), HDF (Harmonic Distortion Factor) |
Computer
interfaces
Remote control
Output data
Data rates
Isolation |
Plug-in unit for 2
interfaces: RS232 and IEEE488.2, only one interface can be
used, selectable by user in the setup (High-speed RS232 with
115kBd for sample values available)
All functions can be remote controlled, keyboard-lock
possible
Output of all displayable data possible, data format
identical for both interfaces
RS232: max. 38400 Baud, IEEE488.2: max. 1MByte/sec
Interfaces isolated against other interface and against
other electronics, isolation voltage 500V |
| Printer
interface |
Parallel PC printer
interface with 25-pin SUB-D-connector for printing of
measurement values, tables and graphics on pin, ink jet or
laser printers |
| Memory Card |
For PC cards (PCMCIA),
datalogging of measuring and sampling values, storing and
restoring of setups |
Monitoring
and storing
of transients |
Storing and graphical
displaying of transients with a resolution of 25µs. Storing
depth is 20,000 measuring values per channel, selectable
recording duration from 0.5 to 60 seconds. Adjustable
pre-trigger, different possibilities of triggering,
logically combinable between channels |
| Processing
signal interface |
4 analogue inputs for
registration of auxiliary quantities (13 bit, ±10V)
8 analogue outputs for output of any measured or computed
values (16 bit, ±10V)
6 digital inputs for registration of status signals and 2
inputs for registration of frequencies (0.1Hz...2MHz)
16 digital outputs to signal states and alarms (at exceeded
limits)
All inputs and outputs are isolated against each other and
against other electronics, isolation voltage 500V |
Low pass
filter,
fundamental harmonic determination |
Low pass filter of 8th
degree, manually adjustable in the range from 0.1Hz to
50kHz, automatically on fundamental in the range from 0.1Hz
to 5kHz |
| Analysis of
harmonic |
Analysis of the frequency
spectrum of voltage, current and power of 50 harmonics in
the range from 0.1Hz to 32kHz in accordance with
IEC61000-3-2, representation as table or graphically |
| Flicker
measurement |
Flicker meter by IEC868 in
full compliance with IEC61000-3-3 |
| Scope and
plot function |
Graphical representation of
sampled or computed values as a function of time |
| External
monitor |
VGA connector for use with
an external monitor |
Other data
External synchronization/trigger
Design, dimensions
Weight
Protection class
Electromagnetic compatibility
Protection system
Operation/storage temperature
Climatic class
Power supply |
Isolated interfaces for external control of measurement
cycle and integration times, outputs for status signals
about the actual measuring, isolation voltage 500V
• standard desktop case, 230 mm x 440 mm x 475 mm (H x W x
D)
• 19"- mounting kit, 6HU/84PU
• Measuring box, 19"chassis, 63PU (front panel expandable to
84PU), 4HU, 179 mm x 376 mm x 350mm
• Measuring box, bench case, 230 mm x 483 mm x 350 mm (H x W
x D)
Depends on design and equipment, 12kg or above
EN61010 (IEC61010, VDE0411), protection class I
EN61000 (IEC61000), EN50081, EN50082
IP20 in accordance with DIN40050
0...40°C / -20...50°C
KYG in accordance with DIN40040
230V/115V (selectable) ±15%, 45...400Hz, about 100VA (70W) |