THREE PHASE PRECISION POWER ANALYZER
LMG310 3 Phase Precision Power Analyzer

LMG310 Three Phase Precision Power Analyzer
Measurements on 3 phase frequency inverters, AC/DC motors, electronic and conventional transformers, switching power supplies, lighting, efficiency of MW-transformers
The LMG310 Three Phase Precision Power Analyzer is suitable for all measurements of 1-, 2- and 3-phase systems.
The measuring channels can be used independently.
The LMG310 Three Phase Precision Power Analyzer is used in the fields of development, manufacturing and quality control, where the simultaneous three phase measurement of active power loss in reactive and nonlinear components like transformers, chokes, motors, capacitors is required; the power consumption and efficiency of power supplies, electrical lamp ballasts and inverters have to be specified.



LMG310 Three Phase Precision Power Analyzer
 
  • Measures voltage, current, power,
    power factor, energy etc.

  • Isolated measurement inputs with extremely
    low capacitance against earth to assure high
    dynamics CMR and interference immunity

  • Dynamic measuring ranges 1 to 1000V and 3mA to 30A directly, only one socket pair per U and I input

  • Harmonics and flicker measurement acc. to
    EN 61000-3-2 and EN 61000-3-3

For price and info please email sales@hotektech.com or call 253-927-6186 ext. 1

Related Instruments Power Analyzers
Power Converters

Digital Multimeters
Clamp-On Meters
  • Inputs for torque and number of revolution

  • External shunt voltage input

  • Analog and digital in/outputs

  • PC interfaces, memory-card slot

  • Graphical 6,4'' TFT color display to show 4 to 24 measured values, wave form, diagram of calculated values, bar charts and vector diagrams (fresnel diagrams)

  • Ergonomic user friendly interface

  • Formula editor

  • LabVIEW driver

  • Also available as measuring box without display  (LMG310-B)

Analysis of harmonics including the limit check of the harmonics according
IEC61000-3-2 (classes A, B, C and D), the flicker measuring according IEC61000-3-3 and transient recording and monitoring are available options of the LMG310 Three Phase Precision Power Analyzer. These three options are used to evaluate the power line disturbances which may be emitted by electrical devices, further to analyze the quality of supplied power in the mains.

TECHNICAL DATA

Voltage measuring
ranges

trms values/V
Permissible peak value /V

 
  1
2
3
6
10
20
30
60
100
200
300
600
1000
2000
         
Overload strength
Input resistance

Crest factor

1500V permanently 2000V for 3s  8000V for 1.2/50ms
>1.5MΩ II 50pF
8...2 for Utrms=25...100% of the selected measuring range
20...8 for Utrms=10...25% of the selected measuring range
Current measuring ranges
trms values /A
Permissible peak /A
Overload strength/A
Overload strength for 3s/A
Input resistance/Ω
 

3m
6m
1.5
4

5


10m
20m
1.5
4

5


30m
60m
1.5
4

5


0.1
0.2
1.5
4

0.3


0.3
0.6
1.5
4

0.3


1
2
1.5
4

0.3


3
6
35
45

0.01


10
20
35
45

0.01


30
60
35
45

0.01
     
Crest factor 8...2 for Itrms=25...100% of the selected measuring range
20...8 for Itrms=10...25% of the selected measuring range
External
shunt voltage ranges

trms values/mV
Permissible peak /mV

Overload strength/V
 

15
30

3


50
100

3


150
300

3
         
Input resistance/Ω
Crest factor
>1.5MΩ
8...2 for Utrms=25...100% of the selected measuring range
20...8 for Utrms=10...25% of the selected measuring range
Range selection Auto, manual or remote-controlled, separately for each current and voltage channel, input of factors for transformers (U and I), level control display for each channel in the status line
Isolation Current and voltage path may float against each other and against earth up to 1500V
Measuring method Simultaneous sampling of the current and voltage inputs and A/D conversion of the instantaneous values
Measuring cycle, synchronization,
averaging
For the measurement of the trms values for current, voltage and active power, the measuring cycle time is adjustable in the range of 100ms to 60s. The synchronization can be performed on the measuring signal, the fundamental harmonic of the measuring signal, the mains or an external signal. A single measurement with automatic stop after 1 measuring cycle is possible. Averaging over 1 to 16 measuring cycles.
Measuring accuracy
Measuring
Accuracy

±(% of measuring value + % measuring range)

DC 0.1...15Hz 15..500Hz 0.5...50kHz 50...150kHz 150...250kHz 250...400kHz 400kHz...1MHz
Voltage 0.1+0.05 0.05+0.05 0.1+0.05 0.2+0.1 0.3+0.2 0.5+0.5 1.0+0.1*(f-400kHz)/kHz
Current 0.01+0.05 0.05+0.05 0.1+0.05 0.2+0.1 0.3+0.2 0.5+0.5 1.0+0.1*(f-400kHz)/kHz
Active Power 0.015+0.1 0.07+0.08 0.15+0.1 0.3+0.3 0.5+0.5 0.7+1.0 1.5+0.15*(f-400kHz)/kHz
Accuracies based on 1. sinusoidal voltages and currents
2. ambient temperature 20...25°C
3. warm up time 15 minutes
4. definition of power range as the product of current and voltage measuring range, 0 ≤λ≤ 1
5. calibration interval 12 month
Other values All other values are derived from the values for current, voltage and active power. Accuracies for the derived values depends on the functional relation (e.g.  S = I * U,  ΔS/S = ΔI/I + ΔU/U)
Frequency measurement 0.01Hz...250kHz ±0.01% of measuring value, measuring channel selectable.
Measuring wirings a) phase current with phase voltage (star/star)
b) phase current with linked voltage (star/delta)
c) linked current with phase voltage (delta/star)
d) linked current with linked voltage (delta/delta)
e) Aron-wiring (2-watt-meter-method, third channel free)
With the optional star/delta conversion the displaying of values from the respectively other wiring is possible (only wiring a) to d))
Display of measured and computed values
Representation


Voltage/current

Power
Impedance
With standard abbreviation of electrical quantity, phase and dimension, 5-digits (0...99999), with sign, decimal point and unit prefix after the digits (e.g. I2trms/A 0.7385m). 1 to 30 values can be displayed simultaneously, selectable via default or user defined menus (max. 10)
Trms value, peak values (min, max, pp), rectified value (rect), mean value (dc), rms value of ac-component, form factor, crest factor
Active power (P), reactive power (Q), apparent power (S), phase anglel (r), power factor (l)
Amount (Z), real and imaginary part of parallel or serial equivalent circuit
Integrated values depending on
the measuring time

Energy, Charge
Date and time, measuring time

Adjustable
parameters
Values of
harmonic analysis
The integration can be controlled manually, automatically using start and stop times, via external trigger or remote on controlled via computer interface

Active energy (Ep), reactive energy (Eq), apparent energy (Es), charge (q)
Current date (day, month, year) with time (hours, minutes, seconds), accu-buffered real time clock,
start time for measurement, running measuring time, on-time, all measuring times with day, hours, minutes, seconds
Scaling factors for external shunts, current and voltage transformers, basic load (stand-by power)

Amount and phase of harmonics for current, voltage and power, angle between phase voltages and phase currents, phasor (Fresnel) diagram, THD (Total Harmonic Distortion), THF (Telephone Harmonic Factor), HVF (Harmonic Voltage Factor), HDF (Harmonic Distortion Factor)
Computer interfaces

Remote control
Output data
Data rates
Isolation
Plug-in unit for 2 interfaces: RS232 and IEEE488.2, only one interface can be used, selectable by user in the setup (High-speed RS232 with 115kBd for sample values available)
All functions can be remote controlled, keyboard-lock possible
Output of all displayable data possible, data format identical for both interfaces
RS232: max. 38400 Baud, IEEE488.2: max. 1MByte/sec
Interfaces isolated against other interface and against other electronics, isolation voltage 500V
Printer interface Parallel PC printer interface with 25-pin SUB-D-connector for printing of measurement values, tables and graphics on pin, ink jet or laser printers
Memory Card For PC cards (PCMCIA), datalogging of measuring and sampling values, storing and restoring of setups
Monitoring and storing
of transients
Storing and graphical displaying of transients with a resolution of 25µs. Storing depth is 20,000 measuring values per channel, selectable recording duration from 0.5 to 60 seconds. Adjustable pre-trigger, different possibilities of triggering, logically combinable between channels
Processing signal interface 4 analogue inputs for registration of auxiliary quantities (13 bit, ±10V)
8 analogue outputs for output of any measured or computed values (16 bit, ±10V)
6 digital inputs for registration of status signals and 2 inputs for registration of frequencies (0.1Hz...2MHz)
16 digital outputs to signal states and alarms (at exceeded limits)
All inputs and outputs are isolated against each other and against other electronics, isolation voltage 500V
Low pass filter,
fundamental harmonic determination
Low pass filter of 8th degree, manually adjustable in the range from 0.1Hz to 50kHz, automatically on fundamental in the range from 0.1Hz to 5kHz
Analysis of harmonic Analysis of the frequency spectrum of voltage, current and power of 50 harmonics in the range from 0.1Hz to 32kHz in accordance with IEC61000-3-2, representation as table or graphically
Flicker measurement Flicker meter by IEC868 in full compliance with IEC61000-3-3
Scope and plot function Graphical representation of sampled or computed values as a function of time
External monitor VGA connector for use with an external monitor
Other data
External synchronization/trigger

Design, dimensions



Weight
Protection class
Electromagnetic compatibility
Protection system
Operation/storage temperature
Climatic class
Power supply


Isolated interfaces for external control of measurement cycle and integration times, outputs for status signals about the actual measuring, isolation voltage 500V
• standard desktop case, 230 mm x 440 mm x 475 mm (H x W x D)
• 19"- mounting kit, 6HU/84PU
• Measuring box, 19"chassis, 63PU (front panel expandable to 84PU), 4HU, 179 mm x 376 mm x 350mm
• Measuring box, bench case, 230 mm x 483 mm x 350 mm (H x W x D)
Depends on design and equipment, 12kg or above
EN61010 (IEC61010, VDE0411), protection class I
EN61000 (IEC61000), EN50081, EN50082

IP20 in accordance with DIN40050

0...40°C / -20...50°C
KYG in accordance with DIN40040
230V/115V (selectable) ±15%, 45...400Hz, about 100VA (70W)
 