  Test measurement instrumens. Analytical instruments, ph meters, dissolved oxygen meter, hygrometer, salinity meter, weather instruments. Power analyzers, Gas combustion analyzer, Electronic load, Harmonic analyzer. Pressure calibrator, deadweight tester, temperature calibrator, current loop calibrator. Temperature datalogger, humidity datalogger, barometric pressure datalogger, wireless datalogger. Hygrometer, humidity datalogger, HVAC instrument, humidity controllers transmitters. Candela meter, Luminance meter, Radiometer, Albedometer, Luminance transmitter. Cable detectors, coil tester, insulation resistance, resistance bridge, tachometers. Digital multimeter, clamp on meter, amperometer, current meter, high line voltmeter, phasing sequence meter.  
  Resistance bridge, ohmmeter, insulation resistance meter, multimeters, milli ohmmeter, micro ohmmeter, megohmmeter. Manometer, Pressure gauge, deadweight tester, pressure pump, vacuum pump, pressure calibrator. Simulator, Calibrator, resistance standard, resistor decade, inductance standard, thermometry standards. Thermometry bridge, precision thermometer, infrared thermometer, temperature calibrator, dry block calibrator. Current transmitter, Luminance transmitter, energy transmitter, pressure transmitter, temperature transmitter.  
 

Search  

 
 
 

THREE PHASE PRECISION POWER ANALYZER


LMG500 Precision Power Meter
The One to Eight Channel Precision Power Meter LMG500, modular for 1 to 8 power channels (phases), designed with most advanced technologies will give best measuring results with all applications, conventional
devices as motors and transformers, as well as with equipment designed in sophisticated power electronics as inverters, lighting and automotive components.
Power Analyzer measuring bandwidth from DC to above 5 MHz. Pulsed signals with less than 200ns duration will be captured and measured most precisely.
Vector display of 3φ systems immediately checks the phase sequence and shows phase interchanges.



LMG500 One to Eight Channel Precision
Power Meter
 
Our Price: Contact us for pricing
Phone: 253-927-6186
866-318-6186
Fax: 253-927-6189
email: sales@hotektech.com
  • Analog and digital in/outputs

  • Additional inputs for current sensors

  • Auto-scaling current sensors compensated in
    phase and amplitude

  • PC interfaces

  • Floppy drive as standard

  • Graphical color display to show measured values, wave form, diagrams of calculated values, bar charts and vector diagrams (fresnel diagrams)

  • Ergonomic user friendly interface

  • Formula editor

  • LabVIEW driver

High sampling rate of 3 Msamples per second, synchronous and simultaneous at all channels
non-gapping, so all steady state as well as dynamic events can be measured and analyzed.
For the computation of rms and average values all data of the continuous 3 MHz-sampling will be used, thus preventing any gaps between those values. This non-gapping sampling and
computing technology guarantees highest precision also at fast signal changes.
The graphics display allows scope and plot functions for waveform and timing diagrams, as well as xy diagrams or bar charts for the harmonics. The status bar at the top of each display menu shows
the input level of the 4 (8) voltage and the 4 (8) current inputs – an important item of information for
the quality of the measurement.
The display also indicates in what groups, A and B (C and D), the input channels are wired and which signals the groups are synchronized to.

TECHNICAL DATA

Voltage measuring
ranges U*

Nominal value /V
Maximum trms value /V
Maximum peak value for full scale /V
Input impedance
 

3
3.6
6.25

>4.5MΩ


6
7.2
12.5

<3pF


12.5
14.4
25


25
30
50


60
60
100


130
130
200


250
270
400


400
560
800


600
1000
1600


1000
1000
3200
   
Current measuring ranges I*
Nominal value /A
Maximum trms value /A
Maximum peak value for full scale /A
Input impedance
 

20m
37m
59m


40m
75m
117m


80m
150m
234m


150m
300m
469m


300m
600m
938m


600m
1.25
1.875


1.2
2.5
3.75


2.5
5
7.5


5
10
15


10
20
30


20
32
60


32
32
120
 
  =============
560mΩ
=============
68mΩ
=============
8.2mΩ
=============
1mΩ
 
Current measuring ranges IHF*
Nominal value /A
Maximum trms value /A
Maximum peak value for full scale /A
Input impedance
 

150m
225m
325m

0.1Ω


300m
450m
650m
 


600m
900m
1.3


1.2
1.8
2.6
       
Sensor inputs USensor,ISensor
Nominal value /V
Maximum trms value /V
Maximum peak value for full scale /V
Input impedance


 


30m
37m
63m

100kΩ


60m
75m
125m

34pF


120m
150m
250m


250m
300m
500m


500m
600m
1


1
1.2
2


2
2.5
4


4
5
8
 
Measuring accuracy
Measuring
Accuracy

±(% of measuring value + % measuring range)

DC 0.05Hz…45Hz 45Hz…65Hz 65Hz…3KHz 3kHz...15kHz 15kHz...100kHz 100kHz...1MHz 1MHz...10MHz
USensor 30mV...4V 0.02+0.06 0.015+0.03 0.01+0.02 0.15+0.03 0.03+0.06 0.2+0.4 0.4+0.8 f/MHz*0.7+f/1MHz*1.5
ISensor 30mV...4V 0.02+0.06 0.015+0.03 0.01+0.02 0.15+0.03 0.03+0.06 0.2+0.4 0.4+0.8 f/MHz*0.7+f/1MHz*1.5
Accuracies based on:
1. sinusoidal voltage and current 4. definition of power range as the product of
2. ambient temperature 23 °C current and voltage range, 0 ≤ IλI ≤ 1, (λ=Power factor=P/S)
3. warm up time 1h 5. calibration interval 12 month
4. definition of power range as the product of
current and voltage range, 0 ≤ IλI ≤ 1, (λ=Power factor=P/S)
5. calibration interval 12 month
All other values are derived from the current, voltage and active power values. Accuracies for derived values depend on the
functional relationship (e.g. S = I * U, ∆S/S = ∆I/I + ∆U/U)
Isolation All current and voltage inputs isolated against each other, against remaining electronic and against earth (1000V/CATIII)
Synchronization The measurement is synchronized on the signals period. There is a choice to determine the period from „line“, „extern“,
u(t), i(t) as well as their envelopes, combined with settable filters . By this very stable readings are achieved, even at signals of
pulse width modulated frequency inverters and amplitude modulated electronic ballasts.
Harmonic analysis (option CE Harm) Measuring of current and voltage with evaluation in full compliance with EN61000-3-2
Harmonic analysis (option Harm100) Analysis of current, voltage and power up to 99th harmonics, in total 100 harmonics including DC component.
Fundamental in the range from 0.1Hz to 150kHz. By integer divider (1...128) a new reference fundamental can be created as to
detect interharmonics. Externally on PC up to 1.5Mhz with software TERM-L5
Flicker measuring (option) Flicker meter according to EN61000-4-15 with evaluation in full compliance with EN61000-3-3
Transients (option) Detecting and recording of transients >330ns
Scope function (standard) Graphical representation of sampled values versus time
Plot function (standard) Time diagram of max. 4 readings, minimal resolution 50ms, respectively 10ms in 50Hz half-wave (flicker) mode
Star delta conversion (option) Sums and differences between channels on sample basis
Computer interfaces
Remote control
Output data
Transfer rate
RS232 (standard) and IEEE488.2 (option), only one interface can be used at the same time,
additional USB (option) and Ethernet (option) available.
All functions can be remote-controlled, keyboard lock for measuring parameters
Output of all readings, data formats BIN/ASCII, SCPI command set
RS232: max.115200 Baud, IEEE488.2: max. 1MByte/s
Floppy drive (standard) 3.5’’, 1.44MB
Printer interface (standard) Parallel PC-Printer interface with 25-pin SUB-D socket, printing measuring values, tables and graphics
to matrix, inkjet or laser printers
Processing signal interface (option) 2 x 25 pin SUB-D socket with:
• 8 analog inputs for process magnitudes (16Bit, ±10V)
• 8 analog outputs (16Bit, ±10V)
• 8 digital inputs
• 8 digital outputs
• 2 input for frequency (0.1Hz...500kHz) and rotation direction
• in- and outputs are isolated against other electronics (test voltage 500V)
Other data
Dimensions

Weight
Protection class
Electromagnetic compatibility
Protection system
Operating/storage temperature
Climatic class
Power supply
• Bench case,
W 433mm x H 147mm x D 400mm
• Accessories: brackets for 19'' rack, 84PU, 3HU, D 400mm
about 12kg
EN61010 (IEC61010, VDE0411), protection class I, overvoltage category III
EN50081, EN50082

IP20 in accordance to DIN40050
0...40°C, -20...50°C

KYG in accordance to DIN40040
85...264V, 47...440Hz, about 100W
LMG500 application software
TERM-L5

PQA

SYS61K-3-SOFT
(Name of software is equal with Part No. Please request detailed data sheets)

Individual configuration of measurement, using all features of the LMG500, spectral analysis, remote of LMG500
Software especially designed for power quality analysis (e.g. EN50160), easy configuring of measurement in a few steps
Control and evaluation software for test systems of harmonics and flicker according to EN61000-3-2/-3

Please email sales@hotektech.com or call 253-927-6186 if you are interested in a purchase, or
  need additional information.