Phone: 253-927-6186
 Email: sales@hotektech.com
HOME
CONTACT US
Home > Instruments > Light Meters > MAVOLUX 5032C Luxmeter Meter

MAVOLUX 5032C Luxmeter Footcandle Meter:
Even very strong light (e.g. from car headlights) can be measured without any additional accessories.
MAVOLUX 5032B Luxmeter Footcandle Meter:
Guaranteed precision and high sensitivity, resolution 0.001, suitable for test and certification purposes
according to the existing regulations. Even extremely low light can be accurately measured, e.g. emergency lighting.

Mavolux 5032 Light Meter

Model  Description 
M502G  Mavolux 5032 C USB : transport case, software, USB cable.
M503G  Mavolux 5032 B USB : transport case, software, USB cable.
H997B  Calibration Certificate.
5908V0120  Luminance attachment for measuring in cd/m2 and fL
M499G  Adapter disk for contact measurement

Wide range of applications: for light technicians, for the control of light sources, street lights, lighting of work places, public buildings, sports facilities, for quality control and quality assurance in the manufacture of lamps and light sources; for light designers and architects; for measurements in agriculture and horticulture.
Both MAVOLUX Luxmeter and Footcandle meters types allow measuring very high light intensities (brightest daylight, head lights) without any additional accessories.
Especially the MAVOLUX 5032 B Luxmeter and Footcandle meters having an initial sensitivity of 0.01 lx allows measuring extremely low light intensities, such as emergency lighting.
Most important:
The MAVOLUX 5032B Luxmeter and Footcandle meters are optimally suited for certification and official inspection procedures due to its high precision acc. to Class B.

Features:
Illuminance can be measured in Lux or Footcandle
Luminance can be measured in candles per square meter (cd/m2) or Footlambert (fL) with optional
luminance attachments
Wide measuring ranges
High resolution Meets Standards DIN 5032-7 and CIE no. 69
Silicon photo diode, color corrected, i.e. its spectral responsitivity is matched to the spectral
photonic vision of the human eye V(l).
The accuracy of that V(l) matching is the main difference between the two MAVOLUX types C and B.
Cosine correction for oblique inciding light
USB 1.1 Interface
CD Rom with software for processing the values measured and regulating the meter
Rugged instrument design, easy operation
Ever-ready case, USB-cable included

MAVOLUX 5032 C USB Technical Data

Meas
Quantity 
Measuring Range  Reso-
lution
in lx 
Reso-
lution
in fc 
  in Lux  in footcandle 
Illumi-
nation 
    0.1 ... 199.9    0.01 ...... 19.99     0.1    0.01
II      1 .......1 999    0.1 ...... 199.9     1   0.1
III     10  ... 19 990    1 ........1 999   10   1
IV     100 ... 199 000    10 ... 19 990 100 10
    in candela/m2  in foot-lambert  cd/m2 in fL
Luminance
with
attachment 
1 ... 1 999     0.1 .......... 199.9  1 0.1
II  10 ... 19 990      1 .......... 1 999  10 1
III  100 ... 199 900     10 ....... 19 990  100 10
IV  1000 ... 1 999 000    100 ... 199 900  1000 100


MAVOLUX 5032 B USB Specification

Meas
Quantity 
Measuring Range  Reso-
lution
in lx 
Reso-
lution
in fc 
  in Lux  in footcandle 
Illumi-
nation 
0.01 ....... 19.99 0.001 ... 1.999     0.01    0.001
II     0.1 ......... 199.9    0.01 ... 19.99     0.1   0.01
III     1  ....... 1 990       0.1 ... 199.9   1   0.1
IV     10 ... ..19 900      1 .... 1 999 10 1
   100 ... 199 000    10 ... 19 990   100   10
    in candela/m2 in foot-lambert cd/m2 fL
Luminance
with
attachment 
I 0.1 .... 199.9   0.01 ... 19.99 0.1 0.01
II 1 ........ 1 999   0.1 ..... 199.9 1 0.1
III 10 ...... 19 999   1 ........ 1 999 10 1
IV 100 .... 199 900   10 ...... 19 990 100 10
V 1000 ...1999 000   100 ... 199 900 1000 100